Skip to content

RESTORIS 186004-08Recalls & reported problems

Mako Surgical Corp. · Tapered Femoral Stem, Ext Offset Neck, 12/14 Taper · Class 2 · Implant

All models of RESTORIS (74)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00848486008109), received between Apr 15, 2024 and Apr 15, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Corroded” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Corroded1

Patient problems

  • Foreign Body Reaction1
  • Metal Related Pathology1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 15, 2024Corroded; Adverse Event Without Identified Device or Use Problem; Corroded; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED BY AN ATTORNEY THAT PLAINTIFF UNDERWENT A REVISION OF A HIP DUE TO ADVERSE LOCAL TISSUE REACTION. THIS PI IS FOR THE PATIENT'S LEFT HIP REVISION.

    Patient problems: Foreign Body Reaction; Metal Related Pathology

    MAUDE report 19104329 ↗

Other Hip Prosthesis, semi-constrained, cemented, metal/polymer, + additive, porous, uncemented devices

All Hip Prosthesis, semi-constrained, cemented, metal/polymer, + additive, porous, uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page