Mako 180737-1Recalls & reported problems
Mako Surgical Corp. · Tibial Bearing Insert · Class 2 · Implant
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00848486016647), received between Aug 14, 2018 and Aug 9, 2022 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Insufficient Device Problem Information2
- Degraded1
- Device Difficult to Setup or Prepare1
- Naturally Worn1
Patient problems
- Ambulation or Postural Difficulties1
- Bacterial Infection1
- Insufficient Information1
- No Consequences Or Impact To Patient1
- Unspecified Infection1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryAug 9, 2022Degraded; Naturally Worn; Degraded; Naturally Worn
IT WAS REPORTED THAT THE PATIENT'S RIGHT MEDIAL PKA WAS CONVERTED TO A TKA DUE TO A FRACTURED BASEPLATE WHICH LED TO BOWING OF THE INSERT. PATIENT'S BMI WAS CITED AS A CAUSE OR CONTRIBUTING FACTOR. INTRA-OPERATIVELY, THE BASEPLATE WAS FOUND TO BE IN 5 PIECES. ALL COMPONENTS WERE REVISED.
Patient problems: Ambulation or Postural Difficulties; Insufficient Information
MAUDE report 15190751 ↗InjuryMar 23, 2021Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT A STAGE 1 REVISION WAS PERFORMED ON THE PATIENT'S RIGHT MAKO PKA DUE TO INFECTION. ALL COMPONENTS WERE REMOVED AND A SPACER WAS PLACED. REP PROVIDED THE PRIMARY IMLANT SHEET AND CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Unspecified Infection
MAUDE report 11551855 ↗InjuryOct 12, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
THE DOCTOR PERFORMED AN I&D INSERT SWAP ON A PATIENT DUE TO INFECTION OF A MAKO RIGHT MEDIAL UNI. ORIGINAL DOS WAS (B)(6) 2020.
Patient problems: Bacterial Infection
MAUDE report 10665864 ↗MalfunctionAug 14, 2018Device Difficult to Setup or Prepare; Device Difficult to Setup or Prepare
AS REPORTED: "DURING THE CASE THERE WAS AN ISSUE WITH THE FIRST 7X8MM TIBIAL POLY USED AS THE POLY WOULD NOT LOCK INTO PLACE PROPERLY. A SECOND 7X8 POLY WAS OPENED AND LOCKED INTO PLACE USING SAME TECHNIQUE WITH NO ISSUES ON FIRST TRY". REP CONFIRMED A SURGICAL DELAY OF 1 MINUTE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7780539 ↗
Other Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive devices
- iTotal® PS TPS111111102Conformis, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer + Additive/Metal/Polymer + Additive327 reports
- ATTUNE 1506-00-006Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive287 reports
- ATTUNE 1506-00-008Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive233 reports
- iTotal® Cruciate Retaining (CR) Knee Replacement System TCR121111102Conformis, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer + Additive/Metal/Polymer + Additive150 reports
- ATTUNE 1506-70-007Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive1 recall2 deaths74 reports
- ATTUNE 1516-40-506Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive121 reports
- ATTUNE 1516-40-605Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive120 reports
- ATTUNE 1506-70-006Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive1 recall70 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.