Sunset Compressor Nebulizer w/ Disposable Adult Neb Kit NEB100Recalls & reported problems

Sunset Healthcare Solutions, Inc. · Sunset Compressor Nebulizer w/ Disposable Adult Neb Kit · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00848530087326), received between Jan 29, 2019 and Jan 2, 2020 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Failure to Power Up” (1 report).

1 other recall cover devices cleared under the same FDA submission (K091272); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Failure to Power Up1
  • Improper or Incorrect Procedure or Method1

Patient problems

  • Asthma1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 2, 2020Failure to Power Up; Failure to Power Up

    THE END USER REPORTED THE COMPRESSOR NEBULIZER (NEB100) WOULD NOT POWER ON RIGHT OUT OF THE BOX DURING AN ASTHMA ATTACK. END USER REPORTS GOING TO THE EMERGENCY ROOM FOR TREATMENT.

    Patient problems: Asthma

    MAUDE report 9541238 ↗
  • MalfunctionJan 29, 2019Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    THE SUNSET COMPRESSOR NEBULIZER, NEB100, CLEARED FOR USE IN THE U.S. WAS SOLD BY A THIRD-PARTY DISTRIBUTOR IN (B)(4) WHERE THE VOLTAGE IS 220V. THIS DEVICE NEB100 DEVICE IS DESIGNED FOR THE U.S. MARKET, WITH A VOLTAGE REQUIREMENT OF 120V. WHEN THE DEVICE WAS USED OUTSIDE OF ITS SPECIFIED VOLTAGE, THE MOTOR OVERHEATED AND STOPPED FUNCTIONING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8286974 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K091272. One submission often covers many models, so check the product and lot details in each record.

  • Z-1058-2019 · Aug 29, 2018 · Terminated