Skip to content

CANCELLOUS LOCKING SCREW, FULLY THREADED 4.0mm x 34mm 8124-034Recalls & reported problems

Advanced Orthopaedic Solutions, Inc. · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00848665033601), received between Apr 4, 2024 and Sep 9, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Positioning Failure” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Positioning Failure3
  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 9, 2025Break; Positioning Failure; Break; Positioning Failure

    ON (B)(6) 2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN 8124-034 CN LOCK SCREW WOULD NOT LOCK INTO THE PLATE. THIS WAS DISCOVERED A PROXIMAL HUMERUS PROCEDURE WHICH WAS COMPLETED WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23004466 ↗
  • MalfunctionJun 12, 2024Positioning Failure; Positioning Failure

    ON 1/31/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT A 8124-034 LOCKING SCREWS WOULD NOT LOCK INTO THE PLATE. THIS OCCURRED DURING A PROCEDURE ON (B)(6) 2022. THE CASE WAS COMPLETED BY USING A NEW 8124-034 WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19519562 ↗
  • MalfunctionApr 4, 2024Positioning Failure; Positioning Failure

    ON 3/12/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN 8124-034 CN LOCK SCREW, F THREAD 4.0MM X 34MM, CN LOCK SCREW, F THREAD 4.0MM X 36MM, AN 8124-036 CN LOCK SCREW, F THREAD, AND QTY. 2 OF AN 8124-038 CN LOCK SCREW, F THREAD 4.0MM X 38MM FROM LOANER SET RAR-9912S WAS NOT FUNCTIONING AND DID NOT LOCK INTO THE PLATE. THIS WAS DISCOVERED DURING A PROXIMAL HUMERUS ORIF PROCEDURE. THE CASE WAS DELAYED BY APPROXIMATELY 60 MINUTES. THE SCREWS WERE INSERTED ON POWER INITIALLY AND TIGHTENED BY HAND. THE PILOT HOLES WERE DRILLED WITH POWER. THE CASE WAS COMPLETED WITH 4.0 PARTIALLY THREADED SCREWS. THE PATIENT HAS NOT EXPERIENCED ANY ADVERSE EFFECTS DUE TO THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19043610 ↗