CANCELLOUS LOCKING SCREW, FULLY THREADED 4.0mm x 40mm 8124-040Recalls & reported problems
Advanced Orthopaedic Solutions, Inc. · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00848665033632), received between Nov 26, 2024 and Jul 24, 2025 — classified as 4 injuries.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Patient Device Interaction Problem1
Patient problems
- Failure of Implant3
- Nonunion/Delayed-union Bone Fracture1
- Pain1
- Unspecified Infection1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryJul 24, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
ON (B)(6) 2025, A PATIENT'S LEGAL REPRESENTATIVE REPORTED VIA EMAIL THAT THE PATIENT UNDERWENT OPEN REDUCTION AND INTERNAL FIXATION OF A COMMINUTED FRACTURE OF THE LEFT HUMERAL NECK ON (B)(6) 2024. FOLLOWING THE PROCEDURE, THE PATIENT EXPERIENCED ONGOING COMPLICATIONS, INCLUDING PERSISTENT PAIN, NONUNION OF THE FRACTURE, AND SIGNIFICANT DISCOMFORT WITH ANY MOVEMENT OF THE LEFT SHOULDER. ON (B)(6) 2024, THE PATIENT WAS INFORMED THAT A TARGETER/DRILL GUIDE HAD BEEN INADVERTENTLY RETAINED FROM THE INITIAL PROCEDURE. A REVISION SURGERY WAS PERFORMED ON (B)(6) 2024 TO REMOVE THE RETAINED INSTRUMENT, DEBRIDE THE INFECTED AREA, AND REPLACE THE SURGICAL HARDWARE. ADDITIONAL INFORMATION RECEIVED ON 7/2/2025: DURING THE REVISION SURGERY ON (B)(6) 2024, THE FOLLOWING COMPONENTS WERE REMOVED: ONE 3022-005 PROXIMAL HUMERAL PLATE; TWO 8110-032 CORTICAL SCREWS; TWO 8110-035 CORTICAL SCREWS; ONE 8110-040 CORTICAL SCREW; ONE 8110-045 CORTICAL SCREW; ONE 8124-040 FULLY THREADED CANCELLOUS LOCKING SCREW; ONE 8124-044 FULLY THREADED CANCELLOUS LOCKING SCREW; THREE 8124-046 FULLY THREADED CANCELLOUS LOCKING SCREWS; TWO 8124-048 FULLY THREADED CANCELLOUS LOCKING SCREWS; AND ONE 8124-050 FULLY THREADED C
Patient problems: Nonunion/Delayed-union Bone Fracture; Pain; Unspecified Infection
MAUDE report 22585829 ↗InjuryDec 12, 2024Break; Break
ON 10/29/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT A 3024-01 11-HOLE LEFT ALPHA PLATE BROKE POSTOPERATIVELY THROUGH THE OBLONG HOLE IN THE FIRST BEND OF THE PLATE. ON (B)(6) 2024, THE PATIENT WAS IMPLANTED WITH THE PLATE, AND REVISION SURGERY IS STILL TO BE DETERMINED. PER THE SALES REPRESENTATIVE, THE PATIENT DID NOT REPORT ANY FALLS OR INJURIES. NO ADDITIONAL INFORMATION HAS BEEN REPORTED. ADDITIONAL INFORMATION RECEIVED ON 11/13/2024: DURING THE INITIAL SURGERY ON (B)(6) 2024, A 3024-01 11-HOLE LEFT ALPHA PLATE, TWO 8110-026 CORTICAL BONE SCREW, AN 8110-028 CORTICAL BONE SCREW, THREE 8110-030 CORTICAL BONE SCREWS, AN 8114-024 DOUBLE LEAD LOCK CORTICAL SCREW, AN 8114-026 DOUBLE LEAD LOCK CORTICAL SCREW, AN 8114-028 DOUBLE LEAD LOCK CORTICAL SCREW, AN 8114-030 DOUBLE LEAD LOCK CORTICAL SCREW, AN 8124-028 FULLY THREADED CANCELLOUS LOCKING SCREW, AN 8124-032 CANCELLOUS LOCKING SCREW, FULLY THREADED, AN 8124-038 FULLY THREADED CANCELLOUS LOCKING SCREW, AN 8124-040 FULLY THREADED CANCELLOUS LOCKING SCREW, AN 8124-046 FULLY THREADED CANCELLOUS LOCKING SCREW, AN 8124-048 FULLY THREADED CANCELLOUS LOCKING SCREW, AN 8124-050 FULLY THREADED CANCELLOUS LOCKING SCREW, AN
Patient problems: Failure of Implant
MAUDE report 20914883 ↗InjuryNov 27, 2024Break; Break
ON 10/29/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT A 3024-016 16-HOLE LEFT ALPHA PLATE BROKE POSTOPERATIVELY THROUGH THE OBLONG HOLE IN THE FIRST BEND OF THE PLATE. ON (B)(6) 2024, THE PATIENT WAS IMPLANTED WITH THE PLATE, AND ON (B)(6) 2024, THE PLATE WAS REMOVED. PER THE SALES REPRESENTATIVE, THE PATIENT DID NOT REPORT ANY FALLS OR INJURIES. NO ADDITIONAL INFORMATION HAS BEEN REPORTED. ADDITIONAL INFORMATION RECEIVED ON 11/13/2024: THE CORRECT ORIGINAL SURGERY DATE IS 2/20/2024. THE PATIENT WAS IMPLANTED WITH THE 3024-016 16-HOLE LEFT ALPHA PLATE, AN 8110-026 CORTICAL BONE SCREW, TWO 8110-028 CORTICAL BONE SCREWS, FOUR 8114-026 DOUBLE LEAD LOCK CORTICAL SCREW, AN 8114-036 DOUBLE LEAD LOCK CORTICAL SCREW, AN 8124-030 FULLY THREADED CANCELLOUS LOCKING SCREW, AN 8124-032 FULLY THREADED CANCELLOUS LOCKING SCREW, TWO 8124-036 FULLY THREADED CANCELLOUS LOCKING SCREW, TWO 8124-040 FULLY THREADED CANCELLOUS LOCKING SCREW, AN 8124-042 FULLY THREADED CANCELLOUS LOCKING SCREW, TWO AR-7268T TIGERTAPE CERCLAGE WITH FIBERLINK SHUTTLE SUTURE, AND AN ABS-2010-10 ALLOSYNC PURE. DURING THE REVISION SURGERY, ALL THE HARDWARE WAS REMOVED. THE REVISION SURGERY WAS COMPLETED USING
Patient problems: Failure of Implant
MAUDE report 20791088 ↗InjuryNov 26, 2024Break; Break
ON 10/29/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT A 3024-016 16-HOLE LEFT ALPHA PLATE BROKE POSTOPERATIVELY THROUGH THE OBLONG HOLE IN THE FIRST BEND OF THE PLATE. ON 2/2/2024, THE PATIENT WAS IMPLANTED WITH THE PLATE, AND ON (B)(6) 2024, THE PLATE WAS REMOVED. PER THE SALES REPRESENTATIVE, THE PATIENT DID NOT REPORT ANY FALLS OR INJURIES. NO ADDITIONAL INFORMATION HAS BEEN REPORTED. ADDITIONAL INFORMATION RECEIVED ON 11/13/2024: THE CORRECT ORIGINAL SURGERY DATE IS (B)(6) 2024. THE PATIENT WAS IMPLANTED WITH THE (B)(6); 16-HOLE LEFT ALPHA PLATE, AN 8110-026 CORTICAL BONE SCREW, TWO 8110-028 CORTICAL BONE SCREWS, FOUR 8114-026 DOUBLE LEAD LOCK CORTICAL SCREW, AN 8114-036 DOUBLE LEAD LOCK CORTICAL SCREW, AN 8124-030 FULLY THREADED CANCELLOUS LOCKING SCREW, AN 8124-032 FULLY THREADED CANCELLOUS LOCKING SCREW, TWO 8124-036 FULLY THREADED CANCELLOUS LOCKING SCREW, TWO 8124-040 FULLY THREADED CANCELLOUS LOCKING SCREW, AN 8124-042 FULLY THREADED CANCELLOUS LOCKING SCREW, TWO AR-7268T TIGERTAPE CERCLAGE WITH FIBERLINK SHUTTLE SUTURE, AND AN ABS-2010-10 ALLOSYNC PURE. DURING THE REVISION SURGERY, ALL THE HARDWARE WAS REMOVED. THE REVISION SURGERY WAS COMPLETED USING A
Patient problems: Failure of Implant
MAUDE report 20788668 ↗
Other Appliance, Fixation, Nail/Blade/Plate Combination, Single Component devices
- LCP 241.901Synthes GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, SINGLE COMPONENT33 reports
- CANCELLOUS LOCKING SCREW, FULLY THREADED 4.0mm x 36mm 8124-036Advanced Orthopaedic Solutions, Inc. · Appliance, Fixation, Nail/Blade/Plate Combination, Single Component9 reports
- CANCELLOUS LOCKING SCREW, FULLY THREADED 4.0mm x 48mm 8124-048Advanced Orthopaedic Solutions, Inc. · Appliance, Fixation, Nail/Blade/Plate Combination, Single Component6 reports
- CANCELLOUS LOCKING SCREW, FULLY THREADED 4.0mm x 46mm 8124-046Advanced Orthopaedic Solutions, Inc. · Appliance, Fixation, Nail/Blade/Plate Combination, Single Component6 reports
- CANCELLOUS LOCKING SCREW, FULLY THREADED 4.0mm x 44mm 8124-044Advanced Orthopaedic Solutions, Inc. · Appliance, Fixation, Nail/Blade/Plate Combination, Single Component5 reports
- CANCELLOUS LOCKING SCREW, FULLY THREADED 4.0mm x 38mm 8124-038Advanced Orthopaedic Solutions, Inc. · Appliance, Fixation, Nail/Blade/Plate Combination, Single Component5 reports
- NA 202.830Synthes GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, SINGLE COMPONENT5 reports
- LCP 241.903Synthes GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, SINGLE COMPONENT5 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page