AutoPulse®Resuscitation System 100 (REF 8700-0730-08)Recalls & reported problems

Zoll Circulation, Inc. · The AutoPulse System consists of 4 primary components, a reusable Platform, a single use chest compression assembly (LifeBand), a rechargeable battery, and a battery charger. · Class 2

All models of AutoPulseResuscitation System (46)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
16
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 00849111000536), received between Mar 7, 2018 and Dec 6, 2024 — classified as 16 malfunctions.

The most frequently reported specific device problem is “Visual Prompts will not Clear” (7 reports).

1 other recall cover devices cleared under the same FDA submission (K112998); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Visual Prompts will not Clear7
  • Device Operational Issue6
  • Device Displays Incorrect Message5
  • Device Operates Differently Than Expected1
  • Display Difficult to Read1
  • Unexpected Shutdown1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Patient Involvement7
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 6, 2024Visual Prompts will not Clear; Visual Prompts will not Clear

    DURING SHIFT CHECK, THE AUTOPULSE PLATFORM (B)(6) DISPLAYED THE FAULT 41 (PATIENT TEMPERATURE SENSOR FAILURE) MESSAGE. PER THE REPORTER, THE DEVICE WAS STORED IN THE AUTOPULSE QUICK CASE WITHIN THE DESIGNATED DRAWER OF THE EMERGENCY MEDICAL VEHICLE. THE FAULT DISPLAYED WHEN THE DEVICE WAS ON THE HARD SURFACE. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20860779 ↗
  • MalfunctionFeb 1, 2024Use of Device Problem; Use of Device Problem

    DURING SERVICING AT ZOLL, THE AUTOPULSE PLATFORM (SN (B)(6)) FAILED THE LOAD CELL CHARACTERIZATION TEST. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18623882 ↗
  • MalfunctionJul 17, 2023Visual Prompts will not Clear; Visual Prompts will not Clear

    DURING SHIFT CHECK, THE AUTOPULSE PLATFORM (SN (B)(6)) DISPLAYED USER ADVISORY "(UA)07" (DISCREPANCY BETWEEN LOAD 1 AND LOAD 2 TOO LARGE) MESSAGE UPON POWERING UP. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17334167 ↗
  • MalfunctionDec 20, 2021Visual Prompts will not Clear; Visual Prompts will not Clear

    DURING THE PREVENTIVE MAINTENANCE (PM) ON (B)(6) 2021, THE AUTOPULSE PLATFORM (SN (B)(4)) FAILED INITIAL FUNCTIONAL TESTING DUE TO FAULT 29 (LOSS OF BRAKE CONNECTIVITY) AND MULTIPLE FAULT 08 (MOTOR CONTROLLER FAULT DETECTED) CODES. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13041264 ↗
  • MalfunctionApr 1, 2021Visual Prompts will not Clear; Visual Prompts will not Clear

    ON (B)(6) 2021, DURING PREVENTIVE MAINTENANCE THE AUTOPULSE PLATFORM (SN (B)(4)) FAILED FUNCTIONAL TESTING AS THE DRIVE TRAIN MOTOR WAS UNABLE TO MAINTAIN COMPRESSIONS AND DISPLAYED "SYSTEM ERROR, OUT OF SERVICE, REVERT TO MANUAL CPR" ERROR MESSAGE. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11605607 ↗
  • MalfunctionJan 26, 2021Unexpected Shutdown; Unexpected Shutdown

    DURING REANIMATION TRAINING ON A MANIKIN, CUSTOMER REPORTED THAT THE AUTOPULSE PLATFORM (SERIAL #(B)(4)) POWER OFF BY ITSELF. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11228303 ↗
  • MalfunctionNov 6, 2020Visual Prompts will not Clear; Visual Prompts will not Clear

    DURING SHIFT CHECK, CUSTOMER REPORTED THAT THE AUTOPULSE PLATFORM (SERIAL#: (B)(4)) DISPLAYED "SYSTEM ERROR, OUT OF SERVICE, REVERT TO MANUAL CPR." NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10799130 ↗
  • MalfunctionJul 10, 2020Visual Prompts will not Clear; Visual Prompts will not Clear

    AS REPORTED, THE AUTOPULSE PLATFORM (SN: (B)(4)) DISPLAYED "SYSTEM ERROR, OUT OF SERVICE, REVERT TO MANUAL CPR" ERROR MESSAGE. NO FURTHER INFORMATION WAS PROVIDED BY THE CUSTOMER. PATIENT USE INFORMATION WAS REQUESTED, BUT NO ADDITIONAL INFORMATION WAS PROVIDED; THEREFORE, PATIENT USE IS UNKNOWN.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10256104 ↗
  • MalfunctionApr 11, 2019Display Difficult to Read; Display Difficult to Read

    DURING TRAINING, THE USER OBSERVED BROKEN LCD DISPLAY ON THE AUTOPULSE PLATFORM (SN (B)(4)) AND THE DISPLAY WAS NOT READABLE. NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 8506335 ↗
  • MalfunctionAug 5, 2018Visual Prompts will not Clear; Visual Prompts will not Clear

    AS REPORTED, THE PATIENT WAS REANIMATED WITH THE AUTOPULSE PLATFORM (SN UNKNOWN), AFTER AN INTERRUPTION OF THE RESUSCITATION DUE TO A RE-POSITIONING OF THE PATIENT BY DOCTORS AND NURSES, THE AUTOPULSE PLATFORM DISPLAYED USER ADVISORY - UA 12 (LIFEBAND NOT PRESENT) ERROR MESSAGE. THE RESUSCITATION WAS INTERRUPTED BY PRESSING THE RED CANCEL BUTTON. NO KNOWN IMPACT OR CONSEQUENCE TO PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7749806 ↗
  • MalfunctionMar 9, 2018Device Displays Incorrect Message; Device Operational Issue; Device Displays Incorrect Message; Device Operational Issue

    AS REPORTED, THE AUTOPULSE PLATFORM (B)(4) WAS DISPLAYING USER ADVISORY (UA) 02 (COMPRESSION TRACKING ERROR) AND USER ADVISORY (UA) 17 (MAX MOTOR ON TIME EXCEEDED DURING ACTIVE OPERATION) ERROR MESSAGE. NO ADDITIONAL INFORMATION WAS PROVIDED. NO KNOWN IMPACT OR CONSEQUENCE TO PATIENT.

    Patient problems: No Patient Involvement

    MAUDE report 7329448 ↗
  • MalfunctionMar 9, 2018Device Displays Incorrect Message; Device Operational Issue; Device Displays Incorrect Message; Device Operational Issue

    DURING PREVENTIVE MAINTENANCE, THE AUTOPULSE PLATFORM (B)(4) WAS DISPLAYING USER ADVISORY (UA) 29 (LOSS OF BRAKE CONNECTIVITY) ERROR MESSAGE.

    Patient problems: No Patient Involvement

    MAUDE report 7329447 ↗
  • MalfunctionMar 9, 2018Device Displays Incorrect Message; Device Operational Issue; Device Displays Incorrect Message; Device Operational Issue

    DURING PREVENTIVE MAINTENANCE, THE AUTOPULSE PLATFORM (B)(4) WAS DISPLAYING ERROR MESSAGE "SYSTEM ERROR, OUT OF SERVICE, REVERT TO MANUAL CPR". NO PATIENT INVOLVEMENT WAS REPORTED.

    Patient problems: No Patient Involvement

    MAUDE report 7329446 ↗
  • MalfunctionMar 9, 2018Device Displays Incorrect Message; Device Operational Issue; Device Displays Incorrect Message; Device Operational Issue

    DURING PREVENTIVE MAINTENANCE, THE AUTOPULSE PLATFORM (SN (B)(4) WAS DISPLAYING ERROR MESSAGE "SYSTEM ERROR, OUT OF SERVICE, REVERT TO MANUAL CPR".

    Patient problems: No Patient Involvement

    MAUDE report 7329445 ↗
  • MalfunctionMar 9, 2018Device Operates Differently Than Expected; Device Operational Issue; Device Operates Differently Than Expected; Device Operational Issue

    DURING PREVENTIVE MAINTENANCE, THE AUTOPULSE PLATFORM (SN (B)(4) WAS OBSERVED WITH NO LCD BACKLIGHT AND BROKEN BATTERY BAY.

    Patient problems: No Patient Involvement

    MAUDE report 7329444 ↗
  • MalfunctionMar 7, 2018Device Displays Incorrect Message; Device Operational Issue; Device Displays Incorrect Message; Device Operational Issue

    DURING PREVENTIVE MAINTENANCE, THE AUTOPULSE PLATFORM (SN (B)(4)) WAS DISPLAYING ERROR MESSAGE "SYSTEM ERROR, OUT OF SERVICE, REVERT TO MANUAL CPR".

    Patient problems: No Patient Involvement

    MAUDE report 7322479 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K112998. One submission often covers many models, so check the product and lot details in each record.

  • Z-0118-2023 · Sep 22, 2022 · Open, Classified
    AutoPulse Li-ion Battery used on the AutoPulse Resuscitation System Model 100. used as an adjunct to manual CPR

Other Compressor, Cardiac, External devices