AutoPulse®Resuscitation System 100 (REF 8700-0730-11)Recalls & reported problems

Zoll Circulation, Inc. · The AutoPulse System consists of 4 primary components, a reusable Platform, a single use chest compression assembly (LifeBand), a rechargeable battery, and a battery charger. · Class 2

Adverse event reports
14
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00849111000598), received between Feb 13, 2017 and Feb 24, 2026 — classified as 14 malfunctions.

The most frequently reported specific device problem is “Visual Prompts will not Clear” (9 reports).

1 other recall cover devices cleared under the same FDA submission (K112998); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Visual Prompts will not Clear9
  • Use of Device Problem4
  • Device Operates Differently Than Expected1
  • Device Operational Issue1
  • Failure to Power Up1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • Insufficient Information2
  • No Patient Involvement2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 24, 2026Use of Device Problem

    THE CUSTOMER REPORTED THAT THE AUTOPULSE PLATFORM (SN (B)(6)) DOES NOT RECOGNIZE THE LIFEBAND. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS WAS NOTICED DURING A MAINTENANCE CHECK. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24436054 ↗
  • MalfunctionJan 8, 2025Visual Prompts will not Clear; Visual Prompts will not Clear

    THE CUSTOMER REPORTED DURING PATIENT USE THE AUTOPULSE PLATFORM (SN (B)(6) DISPLAYED A USER ADVISORY (UA) 07 (DISCREPANCY BETWEEN LOAD 1 AND LOAD 2 TOO LARGE) MESSAGE. THE OPERATOR TRIED TO RE-ALIGN THE LIFEBAND BUT THE ERROR DID NOT CLEAR. NO FURTHER INFORMATION WAS PROVIDED. PATIENT'S STATUS INFORMATION WAS REQUESTED BUT THE CUSTOMER DID NOT PROVIDE A RESPONSE.

    Patient problems: Insufficient Information

    MAUDE report 21105757 ↗
  • MalfunctionJul 3, 2024Visual Prompts will not Clear; Visual Prompts will not Clear

    DURING SHIFT CHECK, THE AUTOPULSE PLATFORM (SN (B)(6)) DISPLAYED USER ADVISORY (UA) 07 (DISCREPANCY BETWEEN LOAD 1 AND LOAD 2 TOO LARGE) UPON POWERING UP. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19668632 ↗
  • MalfunctionFeb 9, 2024Use of Device Problem; Use of Device Problem

    DURING SHIFT CHECK, THE CUSTOMER NOTICED THAT THE AUTOPULSE PLATFORM (SN 31271) EMITTED A BURNING SMELL UPON POWERING UP. THE PLATFORM FUNCTIONS AS INTENDED, ASIDE FROM THE BURNING SMELL. THE CUSTOMER COULD NOT LOCATE WHERE THE BURNING SMELL WAS COMING FROM. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18681487 ↗
  • MalfunctionAug 25, 2023Visual Prompts will not Clear; Visual Prompts will not Clear

    DURING PATIENT USE, CUSTOMER REPORTED THAT THE AUTOPULSE PLATFORM (SERIAL #31331) DISPLAYED USER ADVISORY "(UA) 45" (DRIVESHAFT NOT AT "HOME" POSITION AFTER POWER-ON/RESTART) ERROR MESSAGE. PATIENT'S STATUS INFORMATION WAS REQUESTED BUT THE CUSTOMER DID NOT PROVIDE A RESPONSE.

    Patient problems: Insufficient Information

    MAUDE report 17621159 ↗
  • MalfunctionDec 27, 2022Visual Prompts will not Clear; Visual Prompts will not Clear

    DURING SHIFT CHECK, THE CUSTOMER REPORTED THE AUTOPULSE PLATFORM (SN (B)(4) DISPLAYED USER ADVISORY (UA) 17 (MAX MOTOR ON-TIME EXCEEDED DURING ACTIVE OPERATION) ERROR MESSAGE. THE CUSTOMER TRIED TO REPLACE THE AUTOPULSE LI-ION BATTERY BUT THE ERROR MESSAGE PERSISTED. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16057050 ↗
  • MalfunctionOct 28, 2022Visual Prompts will not Clear; Visual Prompts will not Clear

    DURING SHIFT CHECK, THE AUTOPULSE PLATFORM SN (B)(4) DISPLAYED USER ADVISORY (UA) 07 (DISCREPANCY BETWEEN LOAD 1 AND LOAD 2 TOO LARGE) UPON POWERING ON WHEN THERE IS NO WEIGHT ON THE PLATFORM. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15692627 ↗
  • MalfunctionOct 5, 2022Failure to Power Up; Failure to Power Up

    THE CUSTOMER REPORTED THAT DURING THE SHIFT CHECK, THE AUTOPULSE PLATFORM (B)(4) DID NOT POWER UP USING THE FULLY CHARGED AUTOPULSE LI-ION BATTERIES (B)(4) AND (B)(4). THE CUSTOMER STATED THAT THE BATTERY WAS FULLY CHARGED BEFORE USING IT WITH THE AUTOPULSE PLATFORM, AND THE CUSTOMER COULD NOT PROVIDE INFORMATION ON THE STATUS LEDS. THE CUSTOMER STATED THAT THEY DID NOT HAVE ANOTHER LI-ION BATTERY TO TEST THE AUTOPULSE PLATFORM. IN ADDITION, THE CUSTOMER REPORTED A BROKEN HEAD RESTRAINT ON THE AUTOPULSE PLATFORM. NO PATIENT INVOLVEMENT. PLEASE SEE THE FOLLOWING RELATED MFR REPORT: MFR 3010617000-2022-01755 FOR THE AUTOPULSE LI-ION BATTERY (B)(4). MFR 3010617000-2022-01856 FOR THE AUTOPULSE LI-ION BATTERY (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15545729 ↗
  • MalfunctionOct 2, 2022Visual Prompts will not Clear; Visual Prompts will not Clear

    DURING THE SHIFT CHECK, THE AUTOPULSE PLATFORM (SN (B)(4) DISPLAYED USER ADVISORY (UA) 20 (POSITION OUT OF RANGE) ERROR MESSAGE. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15526334 ↗
  • MalfunctionFeb 16, 2021Use of Device Problem; Use of Device Problem

    DURING DEVICE EVALUATION OF THE AUTOPULSE PLATFORM (SN: (B)(4)), THE PLATFORM FAILED THE LOAD CELL CHARACTERIZATION CHECK DURING FUNCTIONAL TESTING. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11335614 ↗
  • MalfunctionDec 5, 2020Use of Device Problem; Use of Device Problem

    ON (B)(6) 2020, DURING DEVICE EVALUATION, THE AUTOPULSE PLATFORM (SN (B)(4)) FAILED THE LOAD CELL CHARACTERIZATION CHECK DURING FUNCTIONAL TESTING. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10955601 ↗
  • MalfunctionJul 10, 2020Visual Prompts will not Clear; Visual Prompts will not Clear

    DURING THE SHIFT CHECK, THE AUTOPULSE PLATFORM (SN (B)(4)) DISPLAYED USER ADVISORY (UA) 17 (MOTOR ON FOR TOO LONG DURING ACTIVE OPERATION) AND USER ADVISORY (UA) 18 (MAX TAKE-UP REVOLUTION EXCEEDED) ERROR MESSAGES ON THE USER CONTROL PANEL. NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 10261664 ↗
  • MalfunctionApr 1, 2020Visual Prompts will not Clear; Visual Prompts will not Clear

    DURING SHIFT CHECK, CUSTOMER REPORTED THAT THE AUTOPULSE PLATFORM (SERIAL #(B)(4)) DISPLAYED A USER ADVISORY - "(UA) 41" (PATIENT TEMPERATURE SENSOR FAILURE) ERROR MESSAGE AND CRACKED BLACK COVER. NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 9914435 ↗
  • MalfunctionFeb 13, 2017Visual Prompts will not Clear; Device Operates Differently Than Expected; Device Operational Issue; Visual Prompts will not Clear; Device Operates Differently Than Expected; Device Operational Issue

    IT WAS REPORTED THAT A SYSTEM ERROR / OUT OF SERVICE / PERFORM MANUAL CPR ERROR MESSAGE WAS OBSERVED ON THE AUTOPULSE PLATFORM (SN (B)(4)) DISPLAY SCREEN. IT WAS ALSO REPORTED THAT THE PLATFORM'S METAL HOOKS ON THE UPPER HOUSING WAS BROKEN. THE ISSUE OCCURRED DURING SHIFT CHECK. NO PATIENT WAS INVOLVED WITH THIS EVENT. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6327833 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K112998. One submission often covers many models, so check the product and lot details in each record.

  • Z-0118-2023 · Sep 22, 2022 · Open, Classified
    AutoPulse Li-ion Battery used on the AutoPulse Resuscitation System Model 100. used as an adjunct to manual CPR

Other Compressor, Cardiac, External devices