AutoPulse®Resuscitation System 100 (REF 8700-0740-24)Recalls & reported problems
Zoll Circulation, Inc. · The AutoPulse System consists of 4 primary components, a reusable Platform, a single use chest compression assembly (LifeBand), a rechargeable battery, and a battery charger. · Class 2
All models of AutoPulseResuscitation System (46)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00849111001151), received between Jul 30, 2020 and May 23, 2025 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Visual Prompts will not Clear” (6 reports).
1 other recall cover devices cleared under the same FDA submission (K112998); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Visual Prompts will not Clear6
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Insufficient Information1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 23, 2025Use of Device Problem; Use of Device Problem
BUSINESS PARTNER REPORTED THAT THE CUSTOMER'S AUTOPULSE PLATFORM (SN (B)(6)) DOES NOT ADJUST THE LIFEBAND. THE CUSTOMER TESTED THE PLATFORM WITH THE NEW LIFEBAND, BUT THE ISSUE PERSISTED. NO FURTHER INFORMATION WAS PROVIDED. IT IS UNKNOWN WHEN THE PROBLEM OCCURRED. HOWEVER, PATIENT USE INFORMATION WAS REQUESTED, BUT NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Insufficient Information
MAUDE report 22081050 ↗MalfunctionApr 28, 2023Visual Prompts will not Clear; Visual Prompts will not Clear
DURING THE SHIFT CHECK, THE AUTOPULSE PLATFORM (SN (B)(4)) DISPLAYED USER ADVISORY (UA) 28 (LOSS OF CLUTCH CONNECTIVITY) AND FAULT CODE 29 (LOSS OF BRAKE CONNECTIVITY) ERROR MESSAGES, AND THE PLATFORM WILL NOT PERFORM CHEST COMPRESSIONS. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16839985 ↗MalfunctionMar 2, 2022Visual Prompts will not Clear; Visual Prompts will not Clear
DURING PREVENTIVE MAINTENANCE (PM) ON (B)(6) 2022, THE AUTOPULSE PLATFORM (SN (B)(4)) FAILED FUNCTIONAL TESTING DUE TO UA11 (MAX PATIENT TEMPERATURE EXCEEDED) AND UA41 (PATIENT TEMPERATURE SENSOR FAILURE) ERROR MESSAGES. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13645811 ↗MalfunctionAug 15, 2021Visual Prompts will not Clear; Visual Prompts will not Clear
AFTER PATIENT USE, THE AUTOPULSE PLATFORM (SERIAL #(B)(4)) DISPLAYED USER ADVISORY "(UA)02" (COMPRESSION TRACKING ERROR) ERROR MESSAGE. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12322305 ↗MalfunctionJul 31, 2021Visual Prompts will not Clear; Visual Prompts will not Clear
DURING THE PATIENT CALL, THE CUSTOMER REPORTING THE CONTAMINATION OF THE BODILY FLUIDS INSIDE OF THE AUTOPULSE PLATFORM. IN ADDITION, THE AUTOPULSE PLATFORM (SN (B)(4)) DISPLAYED A FAULT CODE 08 (MOTOR CONTROLLER FAULT DETECTED). NO FURTHER INFORMATION WAS PROVIDED. NO CONSEQUENCES OR IMPACT TO PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12257233 ↗MalfunctionSep 11, 2020Visual Prompts will not Clear; Visual Prompts will not Clear
CUSTOMER REPORTED THAT THE AUTOPULSE PLATFORM (SERIAL #(B)(4) DISPLAYED FAULT CODE 16 (TIMEOUT MOVING TO TAKE-UP POSITION) ERROR MESSAGE. NO FURTHER INFORMATION WAS PROVIDED. PATIENT USE INFORMATION WAS REQUESTED BUT NO ADDITIONAL INFORMATION WAS PROVIDED, THEREFORE PATIENT USE IS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10525657 ↗MalfunctionJul 30, 2020Visual Prompts will not Clear; Visual Prompts will not Clear
DURING DEPLOYMENT FOR PATIENT USE, CUSTOMER REPORTED THAT THE AUTOPULSE PLATFORM (SERIAL #(B)(4)) DISPLAYED FAULT CODE 16 (TIMEOUT MOVING TO TAKE-UP POSITION) ERROR MESSAGE UPON POWER UP. USER WAS UNABLE TO CLEAR ERROR MESSAGE. ANOTHER AUTOPULSE PLATFORM WAS USE TO CONTINUE WITH COMPRESSION. NO CONSEQUENCES OR IMPACT TO PATIENT WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10348330 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K112998. One submission often covers many models, so check the product and lot details in each record.
- Z-0118-2023 · Sep 22, 2022 · Open, ClassifiedAutoPulse Li-ion Battery used on the AutoPulse Resuscitation System Model 100. used as an adjunct to manual CPR
Other Compressor, Cardiac, External devices
- AutoPulse®Resuscitation System 100Zoll Circulation, Inc. · Compressor, Cardiac, External3,208 reports
- AutoPulse Li-ion Battery™ 100Zoll Circulation, Inc. · Compressor, Cardiac, External1,151 reports
- AutoPulse® Resuscitation System 100Zoll Circulation, Inc. · Compressor, Cardiac, External1,136 reports
- AutoPulse®Resuscitation System 100Zoll Circulation, Inc. · Compressor, Cardiac, External317 reports
- AutoPulse®Resuscitation System 100Zoll Circulation, Inc. · Compressor, Cardiac, External276 reports
- AutoPulse®Resuscitation System 100Zoll Circulation, Inc. · Compressor, Cardiac, External250 reports
- AutoPulse®Resuscitation System 100Zoll Circulation, Inc. · Compressor, Cardiac, External241 reports
- LUCAS 99576Jolife AB · Compressor, cardiac, external2 recalls9 deaths99 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page