AutoPulse®Resuscitation System 100 (REF 8700-0740-22)Recalls & reported problems
Zoll Circulation, Inc. · The AutoPulse System consists of 4 primary components, a reusable Platform, a single use chest compression assembly (LifeBand), a rechargeable battery, and a battery charger. · Class 2
All models of AutoPulseResuscitation System (46)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00849111001175), received between Nov 11, 2015 and Mar 20, 2020 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Visual Prompts will not Clear” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K112998); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Visual Prompts will not Clear2
- Device Displays Incorrect Message1
- Device Operates Differently Than Expected1
- No Display/Image1
Patient problems
- No Patient Involvement2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 20, 2020Visual Prompts will not Clear; Visual Prompts will not Clear
CUSTOMER REPORTED THAT THE AUTOPULSE PLATFORM (SERIAL #(B)(4)) DISPLAYED USER ADVISORY - "(UA)18" (MAX TAKE-UP REVOLUTION EXCEEDED) UPON POWERING ON. NO CONSEQUENCES OR IMPACT TO PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9861607 ↗MalfunctionMar 2, 2016No Display/Image; Device Operates Differently Than Expected; No Display/Image; Device Operates Differently Than Expected
IT WAS REPORTED THAT THE AUTOPULSE PLATFORM'S (S/N (B)(4)) DISPLAY WAS NOT SHOWING ANY INFORMATION AND WAS "DARK." THERE WAS NO REPORT OF PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Patient Involvement
MAUDE report 5474267 ↗MalfunctionNov 11, 2015Visual Prompts will not Clear; Device Displays Incorrect Message
CUSTOMER REPORTED THEIR AUTOPULSE PLATFORM DISPLAYS USER ADVISORY (UA) 16 (TIMEOUT MOVING TO TAKE-UP POSITION) AND METALLIC SOUNDS WERE HEARD WHEN INITIATING THE DEVICE. THE PLATFORM WAS WORKING INTERMITTENTLY. THE CUSTOMER USED KNOWN GOOD BATTERIES AND LIFEBANDS WITH THE DEVICE. NO PATIENT INVOLVEMENT REPORTED. NO FURTHER INFORMATION PROVIDED.
Patient problems: No Patient Involvement
MAUDE report 5218588 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K112998. One submission often covers many models, so check the product and lot details in each record.
- Z-0118-2023 · Sep 22, 2022 · Open, ClassifiedAutoPulse Li-ion Battery used on the AutoPulse Resuscitation System Model 100. used as an adjunct to manual CPR
Other Compressor, Cardiac, External devices
- AutoPulse®Resuscitation System 100Zoll Circulation, Inc. · Compressor, Cardiac, External3,208 reports
- AutoPulse Li-ion Battery™ 100Zoll Circulation, Inc. · Compressor, Cardiac, External1,151 reports
- AutoPulse® Resuscitation System 100Zoll Circulation, Inc. · Compressor, Cardiac, External1,136 reports
- AutoPulse®Resuscitation System 100Zoll Circulation, Inc. · Compressor, Cardiac, External317 reports
- AutoPulse®Resuscitation System 100Zoll Circulation, Inc. · Compressor, Cardiac, External276 reports
- AutoPulse®Resuscitation System 100Zoll Circulation, Inc. · Compressor, Cardiac, External250 reports
- AutoPulse®Resuscitation System 100Zoll Circulation, Inc. · Compressor, Cardiac, External241 reports
- LUCAS 99576Jolife AB · Compressor, cardiac, external2 recalls9 deaths99 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page