AutoPulse®Resuscitation System 100 (REF 8700-0740-29)Recalls & reported problems

Zoll Circulation, Inc. · The AutoPulse System consists of 4 primary components, a reusable Platform, a single use chest compression assembly (LifeBand), a rechargeable battery, and a battery charger. · Class 2

All models of AutoPulseResuscitation System (46)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00849111001250), received between Oct 20, 2016 and Jan 15, 2024 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Visual Prompts will not Clear” (5 reports).

1 other recall cover devices cleared under the same FDA submission (K112998); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Visual Prompts will not Clear5
  • No Display/Image1

Patient problems

  • No Patient Involvement4
  • Insufficient Information2

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 15, 2024Visual Prompts will not Clear; Visual Prompts will not Clear

    THE CUSTOMER REPORTED THE AUTOPULSE PLATFORM SN (B)(6) DISPLAYS FAULT 41 (PATIENT TEMPERATURE SENSOR FAILURE) MESSAGE THAN CANNOT BE CLEARED. IT IS UNKNOWN WHEN THE PROBLEM OCCURRED. HOWEVER, PATIENT USE INFORMATION WAS REQUESTED BUT NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 18511124 ↗
  • MalfunctionJan 12, 2024Visual Prompts will not Clear; Visual Prompts will not Clear

    AS REPORTED, THE AUTOPULSE PLATFORM (SN (B)(4), DISPLAYED A 'SYSTEM ERROR, OUT OF SERVICE, REVERT TO MANUAL CPR' ERROR MESSAGE". IT IS UNKNOWN WHEN THE PROBLEM OCCURRED. HOWEVER, PATIENT USE INFORMATION WAS REQUESTED BUT NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 18509199 ↗
  • MalfunctionJan 30, 2020Visual Prompts will not Clear; Visual Prompts will not Clear

    DURING PREVENTIVE MAINTENANCE ON JANUARY 9, 2020, THE RETURNED AUTOPULSE PLATFORM (SERIAL #(B)(4)) DISPLAYED "SYSTEM ERROR, OUT OF SERVICE, REVERT TO MANUAL CPR " DURING FUNCTIONAL TEST.

    Patient problems: No Patient Involvement

    MAUDE report 9648345 ↗
  • MalfunctionJan 29, 2020Visual Prompts will not Clear; Visual Prompts will not Clear

    DURING PREVENTIVE MAINTENANCE ON JANUARY 8, 2020, THE RETURNED AUTOPULSE PLATFORM (SERIAL #(B)(4)) DISPLAYED "SYSTEM ERROR, OUT OF SERVICE, REVERT TO MANUAL CPR " DURING FUNCTIONAL TEST.

    Patient problems: No Patient Involvement

    MAUDE report 9645155 ↗
  • MalfunctionJan 29, 2020Visual Prompts will not Clear; Visual Prompts will not Clear

    DURING PREVENTIVE MAINTENANCE ON (B)(6) 2020, THE RETURNED AUTOPULSE PLATFORM (SERIAL #(B)(4)) DISPLAYED "SYSTEM ERROR, OUT OF SERVICE, REVERT TO MANUAL CPR " DURING FUNCTIONAL TEST.

    Patient problems: No Patient Involvement

    MAUDE report 9645154 ↗
  • MalfunctionOct 20, 2016No Display/Image; No Display/Image

    IT WAS REPORTED THAT DURING AN EVALUATION OF THE AUTOPULSE PLATFORM (S/N (B)(4) PRIOR TO USE THE DEVICE SHOWED MISSING PIXELS IN THE PLATFORM DISPLAY SCREEN. NO PATIENT INVOLVED DURING THIS EVENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Patient Involvement

    MAUDE report 6045612 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K112998. One submission often covers many models, so check the product and lot details in each record.

  • Z-0118-2023 · Sep 22, 2022 · Open, Classified
    AutoPulse Li-ion Battery used on the AutoPulse Resuscitation System Model 100. used as an adjunct to manual CPR

Other Compressor, Cardiac, External devices

All Compressor, Cardiac, External devices →