AutoPulse®Resuscitation System 100 (REF 8700-0780-16)Recalls & reported problems

Zoll Circulation, Inc. · The AutoPulse System consists of 4 primary components, a reusable Platform, a single use chest compression assembly (LifeBand), a rechargeable battery, and a battery charger. · Class 2

All models of AutoPulseResuscitation System (46)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00849111001533), received between Feb 15, 2019 and Feb 15, 2019 — classified as 1 malfunction.

The most frequently reported specific device problem is “Visual Prompts will not Clear” (1 report).

1 other recall cover devices cleared under the same FDA submission (K112998); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Visual Prompts will not Clear1

Patient problems

  • No Patient Involvement1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 15, 2019Visual Prompts will not Clear; Visual Prompts will not Clear

    DURING SHIFT CHECK, CUSTOMER REPORTED THAT THE AUTOPULSE PLATFORM (SERIAL # (B)(4)) DISPLAYED A USER ADVISORY - UA07 (DISCREPANCY BETWEEN LOAD 1 AND LOAD 2 TOO LARGE) ERROR MESSAGE. THE ERROR MESSAGE COULD NOT BE CLEARED BY THE USER. NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 8345346 ↗

FDA clearance and classification

Cleared through 510(k) K112998 on Mar 15, 2012 (decision: Substantially Equivalent). Submitted by Zoll Circulation as “Autopulse Resuscitation System Model 1000”. The FDA review took 160 days. FDA record ↗

The FDA regulates product code DRM (compressor, cardiac, external) under 21 CFR 870.5200 as a Class II device. 21 CFR 870.5200 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K112998. One submission often covers many models, so check the product and lot details in each record.

  • Z-0118-2023 · Sep 22, 2022 · Open, Classified
    AutoPulse Li-ion Battery used on the AutoPulse Resuscitation System Model 100. used as an adjunct to manual CPR

Common questions about the AutoPulse®Resuscitation System 100

Has the AutoPulse®Resuscitation System 100 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00849111001533), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.

What problems have been reported with the AutoPulse®Resuscitation System 100?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Feb 15, 2019. That report lists device problem “Visual Prompts will not Clear” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the AutoPulse®Resuscitation System 100?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the AutoPulse®Resuscitation System 100 FDA approved?

It was cleared through 510(k) K112998 on Mar 15, 2012. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the AutoPulse®Resuscitation System 100, and is it still sold?

The FDA's device database lists Zoll Circulation, Inc. as the company responsible for it (brand AutoPulse®Resuscitation System, model 100). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Compressor, Cardiac, External devices

All Compressor, Cardiac, External devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page