Skip to content

AutoPulse Li-ion Battery™ 100 (REF 8700-0752-15)Recalls & reported problems

Zoll Circulation, Inc. · Box containing a Lithium Ion (Li-Ion) battery for use with the AutoPulse ® Platform. · Class 2

All models of AutoPulse Li-ion Battery (9)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00849111001915), received between Jun 30, 2022 and Jun 3, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Battery Problem” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K112998); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Battery Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 3, 2024Battery Problem; Battery Problem

    THE CUSTOMER REPORTED THAT AFTER 30 MINUTES OF PLACING THE AUTOPULSE LI-ION BATTERY (SN (B)(6)) INTO THE AUTOPULSE MULTI-CHEMISTRY BATTERY CHARGER (MCC) (SN (B)(6)) FOR CHARGING, THE BATTERY WAS DAMAGED WITH TWO PENETRATING HOLES. THE BATTERY'S STATUS LEDS SHOWED NO LIGHTS. SMOKE BLACKENED THE LEFT BAY OF THE MCC. BESIDES SMOKE TRACES, THE MCC HAD NO DAMAGE AND FUNCTIONED AS INTENDED. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19448858 ↗
  • MalfunctionJun 30, 2022Battery Problem; Battery Problem

    DURING THE SHIFT CHECK, THE FULLY CHARGED AUTOPULSE LI-ION BATTERY (SN 10997) WILL NOT POWER ON THE AUTOPULSE PLATFORM. THE STATUS INDICATOR OF THE BATTERY SHOWED FOUR GREEN LIGHTS, INDICATING THAT THE BATTERY IS FULLY CHARGED. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14871062 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K112998. One submission often covers many models, so check the product and lot details in each record.

  • Z-0118-2023 · Sep 22, 2022 · Open, Classified
    AutoPulse Li-ion Battery used on the AutoPulse Resuscitation System Model 100. used as an adjunct to manual CPR

Other Compressor, Cardiac, External devices

All Compressor, Cardiac, External devices →

Sources and updates