AutoPulse®Resuscitation System 100 (REF 8700-0740-04)Recalls & reported problems

Zoll Circulation, Inc. · The AutoPulse System consists of 4 primary components, a reusable Platform, a single use chest compression assembly (LifeBand), a rechargeable battery, and a battery charger. · Class 2

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00849111002516), received between Feb 21, 2020 and Dec 13, 2024 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Visual Prompts will not Clear” (5 reports).

1 other recall cover devices cleared under the same FDA submission (K112998); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Visual Prompts will not Clear5
  • Display Difficult to Read1
  • Display or Visual Feedback Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Insufficient Information1
  • No Consequences Or Impact To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 13, 2024Visual Prompts will not Clear; Visual Prompts will not Clear

    THE DISTRIBUTOR REPORTED THAT THE AUTOPULSE PLATFORM (SN (B)(6)) DISPLAYED A "SYSTEM ERROR, OUT OF SERVICE, REVERT TO MANUAL CPR" MESSAGE UPON POWERING ON. NO FURTHER INFORMATION WAS PROVIDED. IT IS UNKNOWN WHEN THE PROBLEM OCCURRED. HOWEVER, PATIENT USE INFORMATION WAS REQUESTED BUT NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 20930036 ↗
  • MalfunctionDec 9, 2024Visual Prompts will not Clear; Visual Prompts will not Clear

    THE DISTRIBUTOR REPORTED THAT DURING THE DEVICE INSPECTION, THE AUTOPULSE PLATFORM (SN: (B)(6) DISPLAYED USER ADVISORY 07 (DISCREPANCY BETWEEN LOAD 1 AND LOAD 2 TOO LARGE) AND UA19 (MAX APPLIED LOAD EXCEEDED) MESSAGES AFTER PERFORMING A FEW COMPRESSIONS. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20880372 ↗
  • MalfunctionSep 27, 2021Display Difficult to Read; Display or Visual Feedback Problem; Display Difficult to Read; Display or Visual Feedback Problem

    DURING SHIFT CHECK, THE BACKLIGHT DOES NOT ILLUMINATE ON THE LCD OF THE AUTOPULSE PLATFORM (SN (B)(4)). PER REPORTER, THE DISPLAY IS UNREADABLE WHEN THE BACKLIGHT IS NOT ILLUMINATING. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12533776 ↗
  • MalfunctionDec 3, 2020Visual Prompts will not Clear; Visual Prompts will not Clear

    DURING SHIFT CHECK, CUSTOMER REPORTED THAT THE AUTOPULSE PLATFORM (SERIAL #(B)(4)) DISPLAYED "SYSTEM ERROR, OUT OF SERVICE, REVERT TO MANUAL CPR " UPON POWERING ON. ERROR MESSAGE COULD NOT BE CLEARED BY THE USER. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10939751 ↗
  • MalfunctionMar 11, 2020Visual Prompts will not Clear; Visual Prompts will not Clear

    THE AUTOPULSE PLATFORM (SN (B)(4) DISPLAYED FAULT CODE 16 (TIMEOUT MOVING TO TAKE-UP POSITION) ERROR MESSAGE. THE PLATFORM POWERED OFF AND POWERED ON BY ITSELF . PATIENT USE INFORMATION WAS REQUESTED BUT NO ADDITIONAL INFORMATION WAS PROVIDED, THEREFORE PATIENT USE IS UNKNOWN.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9821168 ↗
  • MalfunctionFeb 21, 2020Visual Prompts will not Clear; Visual Prompts will not Clear

    DURING SHIFT CHECK, WHEN THE AUTOPULSE PLATFORM (SN: (B)(4)) WAS POWERED ON, THE LIFEBAND (LOT # UNKNOWN) DID NOT RETRACT TO START COMPRESSIONS, AND FAULT CODE 16 (TIMEOUT MOVING TO TAKE-UP POSITION) WAS DISPLAYED ON THE PLATFORM. THE USER RE-ADJUSTED THE LIFEBAND AND PRESSED THE RESTART BUTTON TO CLEAR THE FAULT CODE; HOWEVER, THE ISSUE WAS NOT RESOLVED. NO PATIENT INVOLVEMENT. PLEASE SEE THE FOLLOWING RELATED MFR REPORT: MFR # 3010617000-2020-00257 FOR THE LIFEBAND (LOT # UNKNOWN).

    Patient problems: No Patient Involvement

    MAUDE report 9741996 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K112998. One submission often covers many models, so check the product and lot details in each record.

  • Z-0118-2023 · Sep 22, 2022 · Open, Classified
    AutoPulse Li-ion Battery used on the AutoPulse Resuscitation System Model 100. used as an adjunct to manual CPR

Other Compressor, Cardiac, External devices

All Compressor, Cardiac, External devices →