ICY® Intravascular Heat Exchange Catheter Kit (Heparin) IC-3893AERecalls & reported problems
Zoll Circulation, Inc. · The Icy® Heat Exchange Catheter is a sterile, single use flexible 9.3 F catheter designed for placement in the inferior vena cava from an insertion site in the femoral vein. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00849111075107), received between Aug 29, 2017 and Jun 30, 2020 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Leak/Splash2
- Device Operates Differently Than Expected1
- Fluid/Blood Leak1
- Material Integrity Problem1
Patient problems
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
- Patient Problem/Medical Problem1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 30, 2020Fluid/Blood Leak; Material Integrity Problem; Fluid/Blood Leak; Material Integrity Problem
THE PATIENT WAS HOSPITALIZED AND UNDERWENT IVTM THERAPY. A ZOLL ICY CATHETER WAS INSERTED INTO THE PATIENT'S FEMORAL VEIN AND THE TARGET TEMPERATURE WAS SET TO 36 °C. FOUR HOURS AFTER THE CATHETER PLACEMENT, THE CUSTOMER REPORTED THAT THE 500ML BAG OF SALINE HAD EMPTIED. THERE WAS NO FLUID OBSERVED ON THE PATIENT'S BED OR THE FLOOR. THE CATHETER LEAK AND 500 ML OF SALINE INFUSION WAS SUSPECTED. THE CATHETER WAS NOT REPLACED, HOWEVER, SPECIFICS OF THE ALTERNATIVE THERAPY WERE NOT DISCLOSED. NO CONSEQUENCES OR IMPACT TO PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10215891 ↗MalfunctionJun 14, 2018Leak/Splash; Device Operates Differently Than Expected; Leak/Splash; Device Operates Differently Than Expected
AT AN UNSPECIFIED TIME DURING IVTM THERAPY, THE THERMOGARD CONSOLE DISPLAYED AN AIRTRAP ALARM AND A CHECK SALINE ALARM. THE USER OBSERVED THAT THE SALINE BAG WAS EMPTY. UPON FURTHER INSPECTION NO SIGN OF LEAK ON THE SUK, OR ON THE PATIENT BEDSIDE WAS REPORTED. CATHETER LEAK WAS SUSPECTED. FOLLOWING THIS, THE USER DISCONTINUED IVTM THERAPY AND REMOVED THE CATHETER. NO KNOWN IMPACT OR PATIENT CONSEQUENCE WAS REPORTED. NO FURTHER INFORMATION PROVIDED. MFR 3010617000-2018-00654 FOR THERMOGARD CONSOLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7602583 ↗MalfunctionAug 29, 2017Leak/Splash; Leak/Splash
AS REPORTED, 24 HOURS AFTER THE ICY CATHETER (LOT NUMBER: 73032) WAS INSERTED INTO A PATIENT'S LEFT FEMORAL, BLOOD WAS OBSERVED IN THE LUMEN OF THE START-UP KIT. THE HEALTHCARE TEAM BELIEVES THERE ARE LEAKS IN THE BALLOONS, THUS CAUSING SALINE TO BE LOST IN THE SUK. SUBSEQUENTLY, AN AIR TRAP ALARM OCCURRED. THE CATHETER WAS ULTIMATELY REPLACED. FOLLOWING THE REPLACEMENT, THE PATIENT WAS ABLE TO CONTINUE AND COMPLETE THEIR TEMPERATURE MANAGEMENT THERAPY. PER REPORTER, THERE WAS NO PATIENT INJURY AS A RESULT OF THE REPORTED ISSUE.
Patient problems: Patient Problem/Medical Problem
MAUDE report 6827141 ↗
Other System, Hypothermia, Intravenous, Cooling devices
- Thermogard XP® Thermogard XP, Lang set A, Non-RohsZoll Circulation, Inc. · System, Hypothermia, Intravenous, Cooling3 deaths2,522 reports
- ICY® Intravascular Heat Exchange Catheter Kit EU Applause Custom Luer (Heparin) IC-3893AEZoll Circulation, Inc. · System, Hypothermia, Intravenous, Cooling5 deaths198 reports
- QUATTRO® Intravascular Heat Exchange Catheter Kit Custom Luer Applause (Heparin) IC-4593Zoll Circulation, Inc. · System, Hypothermia, Intravenous, Cooling13 deaths165 reports
- SOLEX® Intravascular Heat Exchange Catheter Primum Kit Custom Luer Applause Hep SL-2593Zoll Circulation, Inc. · System, Hypothermia, Intravenous, Cooling2 deaths154 reports
- ICY® Intravascular Heat Exchange Catheter Kit Applause (Heparin) IC-3893Zoll Circulation, Inc. · System, Hypothermia, Intravenous, Cooling1 death154 reports
- SOLEX® Intravascular Heat Exchange Catheter Kit Custom Luer Applause Heparin EU SL-2593AEZoll Circulation, Inc. · System, Hypothermia, Intravenous, Cooling2 deaths117 reports
- QUATTRO® Intravascular Heat Exchange Catheter Kit Custom Luer Applause Heparin E IC-4593AEZoll Circulation, Inc. · System, Hypothermia, Intravenous, Cooling2 deaths98 reports
- QUATTRO® Intravascular Heat Exchange Catheter Kit Applause (Heparin) IC-4593Zoll Circulation, Inc. · System, Hypothermia, Intravenous, Cooling1 death87 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page