Cool Line® Intravascular Heat Exchange Catheter Kit Japan, Final Pkged Kit Ne CL-2295Recalls & reported problems
Zoll Circulation, Inc. · The Cool Line® Heat Exchange Catheter is a 9.3 Fr, 22 cm, triple infusion lumen catheter designed for placement in the central venous circulation from an insertion site in the jugular, subclavian, or femoral veins. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00849111075176), received between Oct 21, 2024 and Jul 1, 2026 — classified as 1 injury, 8 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak8
- Material Integrity Problem8
- Adverse Event Without Identified Device or Use Problem1
- Difficult to Remove1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- Thrombosis/Thrombus1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 1, 2026Fluid/Blood Leak; Material Integrity Problem
ON (B)(6) 2026, DURING IVTM THERAPY USING THE COOL LINE CATHETER (LOT 214575), THERE WAS BACKFLOW OF BLOOD. THEREFORE, THE CUSTOMER DETERMINED THAT THE CATHETER BALLOON HAD DAMAGED, CAUSING A SALINE LEAK. THE CATHETER HAD BEEN SMOOTHLY INSERTED ON THE FIRST ATTEMPT. THE DWELL TIME OF THE CATHETER WAS 10-20 MINUTES. SINCE THERE WERE NO PRODUCTS IN STOCK AT THE HOSPITAL, THE TREATMENT WAS SWITCHED TO MEDICATION. NO DEVICE MALFUNCTION WAS REPORTED FOR THE SUK. NO BLOOD WAS OBSERVED IN THE TUBING OF THE SUK. NO PATIENT INJURIES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25685223 ↗InjuryApr 17, 2026Adverse Event Without Identified Device or Use Problem
A 50-YEAR-OLD FEMALE PATIENT UNDERWENT IVTM THERAPY ON (B)(6) 2026, USING THE THERMOGARD SYSTEM FOLLOWING CARDIAC ARREST. A COOL LINE CATHETER (LOT # UNKNOWN) WAS INSERTED FROM A FEMORAL VEIN AND PLACED IN THE INFERIOR VENA CAVA. NO ADDITIONAL ADJUNCT PROCEDURES WERE PERFORMED. PRE-EXISTING BLOOD COAGULOPATHY HAD BEEN ASSESSED BEFORE HOSPITALIZATION, AND THE RESULTS WERE AVAILABLE BEFORE INITIATION OF THERAPY. THE PATIENT DID NOT RECEIVE DVT PROPHYLAXIS BEFORE DETECTION OF THROMBOSIS AND WAS NOT SYSTEMICALLY ANTICOAGULATED AT THAT TIME. THE CATHETER PLACEMENT WAS NOT ASSOCIATED WITH ANY VESSEL INJURY. THE PATIENT WAS ALSO NOTED TO HAVE A SUBARACHNOID HEMORRHAGE. ON (B)(6) 2026, AN ULTRASOUND CONFIRMED THE PRESENCE OF A THROMBUS IN THE INFERIOR VENA CAVA, WHICH WAS SUBSEQUENTLY SHOWN ON CT IMAGING TO EXTEND INTO THE RIGHT COMMON ILIAC VEIN. ANTICOAGULANT THERAPY WITH HEPARIN WAS THEN INITIATED. ON (B)(6) 2026, THE CATHETER WAS REMOVED, AND THE PATIENT WAS TRANSFERRED TO ANOTHER HOSPITAL. NO FURTHER INFORMATION REGARDING THE PATIENT'S CONDITION FOLLOWING TRANSFER WAS AVAILABLE.
Patient problems: Thrombosis/Thrombus
MAUDE report 24914592 ↗MalfunctionApr 8, 2026Fluid/Blood Leak; Material Integrity Problem
ON 3/17/2026, SIX DAYS AFTER INITIATING IVTM THERAPY USING THE COOL LINE CATHETER (LOT 212939) FOR A PATIENT WITH A HEAD INJURY, DURING COOLING, THE AIR TRAP ALARM SOUNDED, AND THE SALINE IN THE SALINE BAG WAS EMPTY. TEMPERATURE MANAGEMENT WAS TERMINATED AT THAT POINT. THE ALARM ON THE CONSOLE WAS CLEARED, AND THE CONSOLE IS BACK IN SERVICE. NO INJURIES TO THE PATIENT HAVE BEEN REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24820635 ↗MalfunctionDec 9, 2025Fluid/Blood Leak; Material Integrity Problem; Fluid/Blood Leak; Material Integrity Problem
ON (B)(6) 2025, IVTM THERAPY USING A COOL LINE CATHETER (LOT 204870) WAS INITIATED. AT AROUND 4:00 AM ON (B)(6) 2025, A NURSE REPORTED TO THE CLINICAL ENGINEER THAT THE THERMOGARD HQ CONSOLE GENERATED AN AIR TRAP ALARM. THERAPY WAS DISCONTINUED. THE ALARM ON THE CONSOLE WAS CLEARED AND THE CONSOLE IS FUNCTIONING AS INTENDED. THE DWELL TIME OF THE CATHETER WAS ABOUT 6 DAYS. ON (B)(6) 2025, AZM SALES REP. CONTACTED THE PHYSICIAN AND CONFIRMED THAT A COOL LINE CATHETER HAD BEEN USED AND THAT THE BALLOON WAS SUSPECTED TO HAVE BEEN DAMAGED. NO PATIENT INJURIES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23763911 ↗MalfunctionSep 18, 2025Fluid/Blood Leak; Difficult to Remove; Material Integrity Problem; Fluid/Blood Leak; Difficult to Remove; Material Integrity Problem
AT 11:00 A.M. ON (B)(6) 2025, IVTM THERAPY USING COOL LINE CATHETER (LOT 199403) BEGAN. THE CATHETER WAS INSERTED INTO THE LEFT INTERNAL JUGULAR VEIN. AT 9:00 A.M. ON 8/24/2025, THE AMOUNT OF SALINE IN THE SALINE BAG WAS DECREASING, AND DAMAGE TO THE CATHETER BALLOON WAS SUSPECTED, SO THE MEDICAL DOCTOR CONSIDERED REPLACING THE CATHETER. AT 5:00 P.M. ON THE SAME DAY, THE DOCTOR CONSIDERED SURGICAL REMOVAL OF THE CATHETER, AS HE CONSIDERED IT DIFFICULT TO REMOVE. HOWEVER, THE CATHETER WAS ABLE TO BE REMOVED WITHOUT ANY PROBLEM. ADDITIONAL SURGERY/ SURGICAL REMOVAL WAS NOT NEEDED. NO INJURY OR HARM TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23092461 ↗MalfunctionAug 28, 2025Fluid/Blood Leak; Material Integrity Problem; Fluid/Blood Leak; Material Integrity Problem
ON (B)(6) 2025, IVTM THERAPY USING COOL LINE CATHETER (LOT 206444) FOR A HEAD TRAUMA PATIENT BEGAN. ON (B)(6) 2025, IT WAS REPORTED THAT THE SALINE IN THE SALINE BAG HAD EMPTIED. THIS WAS CONFIRMED BY THE MEDICAL DOCTOR. IN ADDITION, AS IT WAS TIME TO END TEMPERATURE MANAGEMENT, THE CATHETER WAS REMOVED, AND TEMPERATURE MANAGEMENT WAS TERMINATED. SALINE INFUSION OF ABOUT 200-300 ML INTO THE PATIENT WAS SUSPECTED. THE DWELL TIME OF THE CATHETER WAS ABOUT 5 DAYS. NO CONSEQUENCES OR IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22918760 ↗MalfunctionMar 12, 2025Fluid/Blood Leak; Material Integrity Problem; Fluid/Blood Leak; Material Integrity Problem
IVTM THERAPY USING THE COOL LINE CATHETER (LOT 196904) WAS STARTED. APPROXIMATELY TWO HOURS AFTER THE START, THE AIR TRAP ALARM ON THE THERMOGARD XP IVTM SYSTEM SOUNDED, AND BLOOD WAS OBSERVED FLOWING INTO THE SALINE BAG. SINCE THE PATIENT'S CONDITION WAS STABLE, THE USE OF THERMOGARD WAS DISCONTINUED. THE ALARM ON THE CONSOLE WAS CLEARED. NO DEVICE MALFUNCTION IS SUSPECTED FOR THE START-UP KIT (SUK). NO PATIENT INJURIES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21590099 ↗MalfunctionNov 16, 2024Fluid/Blood Leak; Material Integrity Problem; Fluid/Blood Leak; Material Integrity Problem
FOUR DAYS AFTER THE IVTM THERAPY STARTED, THE CUSTOMER NOTICED THE SOLUTION IN THE SALINE BAG HAD DECREASED. THE PHYSICIAN REPLACED THE START-UP KIT (SUK) (LOT # 200924) AND THE 500-ML SALINE BAG TO CONTINUE THE THERAPY. THE CUSTOMER PERFORMED THE CATHETER AND THE SUK LEAK CHECK PER IFU BEFORE REPLACING THE SUK, AS MEDICAL STAFF SUSPECTED A LEAK FROM THE SUK. HOWEVER, THE SALINE LEVEL DECREASED AGAIN, PROMPTING THE PHYSICIAN TO REPLACE THE COOL LINE CATHETER (LOT # 199403) AND THE SECOND SUK (LOT # UNKNOWN) TO CONTINUE THE TREATMENT. NO FURTHER INFORMATION WAS PROVIDED. NO CONSEQUENCES OR IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20701224 ↗MalfunctionOct 21, 2024Fluid/Blood Leak; Material Integrity Problem; Fluid/Blood Leak; Material Integrity Problem
AS REPORTED, DURING THE IVTM TREATMENT FOR THE TRAUMA PATIENT, THE CUSTOMER NOTICED THE SOLUTION IN THE SALINE BAG HAD DECREASED. THE PHYSICIAN PERFORMED THE COOL LINE CATHETER (LOT # 195818) CHECK AND THE CATHETER LEAK WAS SUSPECTED. NO FURTHER DETAILS WERE PROVIDED. NO CONSEQUENCES OR IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20501260 ↗
Other System, Hypothermia, Intravenous, Cooling devices
- Thermogard XP® Thermogard XP, Lang set A, Non-RohsZoll Circulation, Inc. · System, Hypothermia, Intravenous, Cooling3 deaths2,522 reports
- ICY® Intravascular Heat Exchange Catheter Kit EU Applause Custom Luer (Heparin) IC-3893AEZoll Circulation, Inc. · System, Hypothermia, Intravenous, Cooling5 deaths198 reports
- QUATTRO® Intravascular Heat Exchange Catheter Kit Custom Luer Applause (Heparin) IC-4593Zoll Circulation, Inc. · System, Hypothermia, Intravenous, Cooling13 deaths165 reports
- SOLEX® Intravascular Heat Exchange Catheter Primum Kit Custom Luer Applause Hep SL-2593Zoll Circulation, Inc. · System, Hypothermia, Intravenous, Cooling2 deaths154 reports
- ICY® Intravascular Heat Exchange Catheter Kit Applause (Heparin) IC-3893Zoll Circulation, Inc. · System, Hypothermia, Intravenous, Cooling1 death154 reports
- SOLEX® Intravascular Heat Exchange Catheter Kit Custom Luer Applause Heparin EU SL-2593AEZoll Circulation, Inc. · System, Hypothermia, Intravenous, Cooling2 deaths117 reports
- QUATTRO® Intravascular Heat Exchange Catheter Kit Custom Luer Applause Heparin E IC-4593AEZoll Circulation, Inc. · System, Hypothermia, Intravenous, Cooling2 deaths98 reports
- QUATTRO® Intravascular Heat Exchange Catheter Kit Applause (Heparin) IC-4593Zoll Circulation, Inc. · System, Hypothermia, Intravenous, Cooling1 death87 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page