Cool Line® Intravascular Heat Exchange Catheter Applause Custom Luer CL-2295CORecalls & reported problems
Zoll Circulation, Inc. · The Cool Line® Heat Exchange Catheter is a 9.3 Fr, 22 cm, triple infusion lumen catheter designed for placement in the central venous circulation from an insertion site in the jugular, subclavian, or femoral veins. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00849111075510), received between Nov 1, 2017 and May 6, 2018 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Leak/Splash2
- Device Issue1
Patient problems
- No Known Impact Or Consequence To Patient3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 6, 2018Leak/Splash; Leak/Splash
ON (B)(6) 2018 AT 1 PM, THE THERMOGARD IVTM SYSTEM WAS USED FOR PATIENT TREATMENT. AFTER 11 HOURS OF RUNNING THE THERMOGARD IVTM SYSTEM, IT WAS OBSERVED THAT THE SALINE FLUID LEVEL HAD DROPPED QUICKLY. THE DOCTOR REMOVED THE 1ST CATHETER AND FOUND IT HAD A LEAK. THE DOCTOR REPLACED THE 1ST CATHETER WITH A NEW CATHETER (2ND) AND THE SYSTEM FUNCTIONED PROPERLY. AROUND 1 AM, THE THERMOGARD IVTM SYSTEM ALARMED AND INDICATED THAT AN AIR TRAP LEAK WAS AT FAULT. THE DOCTOR AND BME TURNED OFF THE SYSTEM AND CLEANED UP ALL THE MOISTURE ON THE AIRTRAP. AFTER CLEANING, THE SYSTEM WAS RESTARTED AGAIN, BUT IT DID NOT FUNCTION. HOWEVER, THE USER STATED THAT THE SYSTEM FUNCTION PROPERLY AT A LATER UNSPECIFIED TIME. THE USER REPORTED THAT THE TWO CATHETERS HAD A LEAK. ONE CATHETER LEAKED IN ONE PLACE AND THE OTHER CATHETER LEAKED IN TWO PLACES. NO FURTHER PATIENT INFORMATION WAS PROVIDED. NO KNOWN IMPACT OR CONSEQUENCE TO PATIENT. REFERENCE MFR REPORTS: 3010617000-2018-00521 (2ND ICY CATHETER) AND 3010617000-2018-00522 (THERMOGARD XP CONSOLE).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7489574 ↗MalfunctionMay 6, 2018Leak/Splash; Leak/Splash
ON (B)(6) 2018 AT 1 PM, THE THERMOGARD IVTM SYSTEM WAS USED FOR PATIENT TREATMENT. AFTER 11 HOURS OF RUNNING THE THERMOGARD IVTM SYSTEM, IT WAS OBSERVED THAT THE SALINE FLUID LEVEL HAD DROPPED QUICKLY. THE DOCTOR REMOVED THE 1ST CATHETER AND FOUND IT HAD A LEAK. THE DOCTOR REPLACED THE 1ST CATHETER WITH A NEW CATHETER (2ND) AND THE SYSTEM FUNCTIONED PROPERLY. AROUND 1 AM, THE THERMOGARD IVTM SYSTEM ALARMED AND INDICATED THAT AN AIR TRAP LEAK WAS AT FAULT. THE DOCTOR AND BME TURNED FUNCTION. HOWEVER, THE USER STATED THAT THE SYSTEM FUNCTION PROPERLY AT A LATER UNSPECIFIED TIME. THE USER REPORTED THAT THE TWO CATHETERS HAD A LEAK. ONE CATHETER LEAKED IN ONE PLACE AND THE OTHER CATHETER LEAKED IN TWO PLACES. NO FURTHER PATIENT INFORMATION WAS PROVIDED. NO KNOWN IMPACT OR CONSEQUENCE TO PATIENT. REFERENCE MFR REPORTS: 3010617000-2018-00520 (1ST ICY CATHETER) AND 3010617000-2018-00522 (THERMOGARD XP CONSOLE).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7489573 ↗MalfunctionNov 1, 2017Device Issue; Device Issue
AS REPORTED, PRIOR TO INSERTION THE ATTENDING PHYSICIAN OBSERVED THE BALLOONS ON THE SOLEX CATHETER HAD SEPARATED FROM THE CATHETER. ACCORDING TO THE REPORTER, THE BALLOONS WERE NOT INFLATED PRIOR TO THE PLACEMENT. ANOTHER SOLEX CATHETER KIT WAS OBTAINED. THERE WAS NO SIGNIFICANT DELAY IN THE REPLACEMENT NOR WAS THERE ANY REPORT OF PATIENT CONSEQUENCE AS A RESULT OF THE ISSUE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6994039 ↗
Other System, Hypothermia, Intravenous, Cooling devices
- Thermogard XP® Thermogard XP, Lang set A, Non-RohsZoll Circulation, Inc. · System, Hypothermia, Intravenous, Cooling3 deaths2,522 reports
- ICY® Intravascular Heat Exchange Catheter Kit EU Applause Custom Luer (Heparin) IC-3893AEZoll Circulation, Inc. · System, Hypothermia, Intravenous, Cooling5 deaths198 reports
- QUATTRO® Intravascular Heat Exchange Catheter Kit Custom Luer Applause (Heparin) IC-4593Zoll Circulation, Inc. · System, Hypothermia, Intravenous, Cooling13 deaths165 reports
- SOLEX® Intravascular Heat Exchange Catheter Primum Kit Custom Luer Applause Hep SL-2593Zoll Circulation, Inc. · System, Hypothermia, Intravenous, Cooling2 deaths154 reports
- ICY® Intravascular Heat Exchange Catheter Kit Applause (Heparin) IC-3893Zoll Circulation, Inc. · System, Hypothermia, Intravenous, Cooling1 death154 reports
- SOLEX® Intravascular Heat Exchange Catheter Kit Custom Luer Applause Heparin EU SL-2593AEZoll Circulation, Inc. · System, Hypothermia, Intravenous, Cooling2 deaths117 reports
- QUATTRO® Intravascular Heat Exchange Catheter Kit Custom Luer Applause Heparin E IC-4593AEZoll Circulation, Inc. · System, Hypothermia, Intravenous, Cooling2 deaths98 reports
- QUATTRO® Intravascular Heat Exchange Catheter Kit Applause (Heparin) IC-4593Zoll Circulation, Inc. · System, Hypothermia, Intravenous, Cooling1 death87 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗