Trilogy Delivery Catheter TRILOGY-DC-LRecalls & reported problems

Jenavalve Technology, Inc. · Transfemoral Delivery Catheter, Size 27MM · Class 3

All models of Trilogy Delivery Catheter (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00850000898158), received between Aug 7, 2026 and Aug 7, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Human-Device Interface Problem” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 31, 2026Open, Classified

    Delivery Catheter may have sleeve liner separation that results in difficulty or inability to deploy heart valve device.

    Z-0136-2027 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Human-Device Interface Problem1

Patient problems

  • Perforation of Vessels1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 7, 2026Human-Device Interface Problem

    It was reported that during a transcatheter aortic valve (THV) procedure, a 27mm Trilogy Heart Valve was successfully deployed. Following deployment, the physician encountered difficulty withdrawing the delivery catheter into the sheath during delivery system removal. Deformation at the sheath tip was observed after retrieval.  This resulted in injury to the femoral artery, and vascular surgery was performed to repair femoral artery. The delivery catheter is not available for return.A full device history record review of the system was conducted, and all system components met acceptance criteria. Instructions for use indicate on 7.5.3 step include the following: "CAUTION Do not retract the Delivery Catheter if resistance is encountered at the introducer Sheath tip ad this can damage the Sleeve and/or introducer Sheath and lead to vascular damage during the system withdrawal. Instead, reposition the introducer Sheath to ensure proper alignment with the Sleeve under fluoroscopy and slowly retract the Sleeve into the Introducer Sheath"

    Patient problems: Perforation of Vessels

    MAUDE report 50184032 ↗

Other Aortic Valve, Prosthesis, Percutaneously Delivered devices

All Aortic Valve, Prosthesis, Percutaneously Delivered devices →