Skip to content

X-PAC EI2473LTRecalls & reported problems

Expanding Innovations, Inc. · X-Pac 12mm x 32mm Lordotic Tall · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00850001720250), received between Sep 3, 2025 and Apr 1, 2026 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Collapse” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Collapse2

Patient problems

  • Intervertebral Disc Compression or Protrusion2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 1, 2026Collapse

    THE DEVICE WAS IMPLANTED AT L3-L4 ON (B)(6) 2025. INTRAOPERATIVE IMAGES PROVIDED BY THE SURGEON INDICATE THE CAGE WAS PARTIALLY EXPANDED AT THE TIME OF DEPLOYMENT. FOLLOW-UP IMAGING OBTAINED ON (B)(6) 2025 DEMONSTRATES A REDUCTION IN CAGE HEIGHT. THE DEVICE REMAINS IMPLANTED IN THIS CONDITION. PER THE SURGEON, THE PATIENT HAS DEMONSTRATED CONTINUED CLINICAL IMPROVEMENT AND REPORTED SATISFACTION AT FOLLOW-UP VISITS ON (B)(6) 2025 AND (B)(6) 2026. THE EVENT WAS REPORTED TO EI BY THE SURGEON DURING A ROUTINE FOLLOW-UP COMMUNICATION ON (B)(6) 2026.

    Patient problems: Intervertebral Disc Compression or Protrusion

    MAUDE report 24760408 ↗
  • InjurySep 3, 2025Collapse; Collapse

    EXPANDING INNOVATIONS WAS NOTIFIED THAT A 2 LEVEL TLIF PROCEDURE HAD USED A 12X32 LORDOTIC IMPLANT WHICH EXHIBITED AN INTRA-OP REDUCTION IN HEIGHT AFTER EXPANSION. THE SCRUB TECH REPORTED HAVING DIFFICULTY ASSEMBLING THE CAGE/CANNULA/INSERTER SYSTEM. THE CAGE WAS INSERTED BY THE SURGEON AND EXPANDED. INTRA-OP IMAGES TAKEN DURING THE PROCEDURE SHOWED THE REDUCTION IN EXPANSION AFTER INITIAL EXPANSION.

    Patient problems: Intervertebral Disc Compression or Protrusion

    MAUDE report 22954894 ↗

Other Intervertebral Fusion Device With Bone Graft, Lumbar devices

All Intervertebral Fusion Device With Bone Graft, Lumbar devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page