Endologix LLC · The Ovation iX™ Iliac Extensions are components of the Ovation Prime, Ovation iX, and Alto™ Abdominal Stent Graft Systems. The Ovation iX™ Iliac Extensions are comprised of a nitinol stent encapsulated in low-permeability PTFE. Iliac extensions may be used to extend the length of the device, where indicated. · Class 3 · Implant

All models of Ovation iX Iliac Stent Graft (66)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00850007370053), received between Jul 2, 2024 and Jul 2, 2024 — classified as 1 death.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

3 other recalls cover devices cleared under the same FDA submission (P120006); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1
  • Malposition of Device1
  • Off-Label Use1

Patient problems

  • Hypovolemic Shock1
  • Ischemia1
  • Obstruction/Occlusion1
  • Pulmonary Embolism1
  • Renal Failure1
  • Respiratory Failure1
  • Vascular Dissection1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • DeathJul 2, 2024Off-Label Use; Malposition of Device; Device Dislodged or Dislocated; Off-Label Use; Malposition of Device; Device Dislodged or Dislocated

    THE PATIENT WAS BEING TREATED FOR OCCLUSIVE WORK ON (B)(6) 2024. REPORTEDLY, THE PATIENT¿S DISTAL AORTA WAS FULL OF CLOTS ALL THE WAY UP TO THE RIGHT RENAL. THE RIGHT COMMON FEMORAL ACCESS WAS ACHIEVED, AND PERCLOSE (ABBOTT VASCULAR) DEVICES WERE DEPLOYED. A DISSECTION OCCURRED IN THE LEFT EXTERNAL ILIAC. THE PHYSICIAN WAS ABLE TO GO UP AND OVER FROM THE RIGHT COMMON FEMORAL AND BALLOON THE DISSECTION AND MANUAL PRESSURE WAS HELD ON THE ACCESS SITE. DUE TO THE DISSECTION IN THE LEFT COMMON ILIAC, TWO OVATION IX LIMB EXTENSIONS 18X45MM WERE IMPLANTED, OVERLAPPING FROM JUST ABOVE THE BIFURCATION UP TO THE LEFT RENAL. THESE DEVICES WERE DEPLOYED WITHOUT ANY DIFFICULTY. THE DISTAL AORTA WAS DILATED USING A Q50 (MERIT MEDICAL) BALLOON. POST ANGIOGRAM SHOWED THAT THE STENTS HAD EXPANDED, BUT THEY SQUEEZED THE CLOT OUT INTO THE DISTAL AORTA. DUE TO THIS THE PHYSICIAN ELECTED TO PERFORM A CUT DOWN LEFT GROIN AND GET ACCESS SO AN AFX2 BIFURCATED STENT GRAFT (BSG) COULD BE PLACED IN THE DISTAL AORTA AND COVER THE CLOT THAT HAD EMBOLIZED. EVEN WITH THE DIRECT CUT DOWN TO THE LEFT COMMON FEMORAL ARTERY (CFA) GETTING INTO THE TRUE LUMEN IN THE LEFT EXTERNAL ILIAC ARTERY WAS DIFFICULT. REQUIRING

    Patient problems: Hypovolemic Shock; Ischemia; Obstruction/Occlusion; Pulmonary Embolism; Renal Failure; Respiratory Failure; Vascular Dissection

    MAUDE report 19658797 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P120006. One submission often covers many models, so check the product and lot details in each record.

  • Z-2263-2020 · May 6, 2020 · Terminated
    Ovation iX Abdominal Stent Graft Systems, REF: TV-AB2080-I, TV-AB2080-J, TV-AB2380-I, TV-AB2380-J, TV-AB2680-I, TV-AB2680-J, TV-AB2980-I, TV-AB2980-J, TV-AB3480-I, TV-AB3480-J
  • Z-0125-2015 · Sep 20, 2014 · Terminated
    Ovation Prime Abdominal Stent Graft System - 29mm Aortic Body stent grafts. The TriVascular Ovation Prime Abdominal Stent Graft System is a low-profile endovascular device deliver
  • Z-0288-2013 · Oct 19, 2012 · Terminated
    TriVascular brand Fill Polymer Kit, Endovascular Graft, Aneurysm Treatment System, European Model Number: TVFP14-B, (in European distributor inventory only); PRODUCT Usage: The

Other System, Endovascular Graft, Aortic Aneurysm Treatment devices

All System, Endovascular Graft, Aortic Aneurysm Treatment devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page