Skip to content

Ovation iX Iliac Stent Graft TV-IL1428140-JRecalls & reported problems

Endologix LLC · The Ovation iX™ Iliac Limbs are components of the Ovation Prime, Ovation iX, and Alto™ Abdominal Stent Graft Systems. The Ovation iX™ Iliac Limbs are comprised of a nitinol stent encapsulated in low-permeability PTFE. The iliac limbs are deployed into the leg sections of the aortic body. Additional iliac limbs may be used to extend the length of the device, where indicated. · Class 3 · Implant

All models of Ovation iX Iliac Stent Graft (66)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00850007370091), received between Dec 14, 2022 and May 13, 2024 — classified as 3 injuries.

The most frequently reported specific device problem is “Collapse” (1 report).

3 other recalls cover devices cleared under the same FDA submission (P120006); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available1
  • Collapse1
  • Complete Blockage1
  • Obstruction of Flow1

Patient problems

  • Obstruction/Occlusion2
  • Failure of Implant1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 13, 2024Complete Blockage; Complete Blockage

    THE PATIENT WAS INITIALLY TREATED FOR AN ABDOMINAL AORTIC ANEURYSM (AAA) ON (B)(6) 2024 WITH THE IMPLANT OF AN ALTO ABDOMINAL STENT GRAFT SYSTEM. ROUTINE FOLLOW-UP CONDUCTED ON (B)(6) 2024 IDENTIFIED OCCLUSION OF THE LEFT OVATION IX ILIAC LIMB. THE PHYSICIAN ELECTED TO RESOLVE THE OCCLUSION BY RELINING WITH A GORE (NON-ENDOLOGIX) VIABAHN STENT DURING REINTERVENTION CONDUCTED ON (B)(6) 2023. THE PATIENT WAS REPORTED AS DOING WELL POST-SECONDARY PROCEDURE.

    Patient problems: Obstruction/Occlusion; Thrombosis/Thrombus

    MAUDE report 19306250 ↗
  • InjuryApr 3, 2024Collapse; Obstruction of Flow; Collapse; Obstruction of Flow

    THE PATIENT WAS IMPLANTED WITH THE ALTO STENT GRAFT SYSTEM TO TREAT AN ABDOMINAL AORTIC ANEURYSM (AAA) ON (B)(6) 2024. APPROXIMATELY TWO (2) MONTHS AFTER THE INITIAL PROCEDURE, THE PATIENT PRESENTED WITH A THROMBOSED LEFT ILIAC LIMB. THE PHYSICIAN ELECTED TO PERFORM A PROCEDURE TO ADDRESS THE THROMBOSIS, INCLUDING MECHANICAL CLOT REMOVAL AND RELINING OF THE LIMB WITH A GORE (NON-ENDOLOGIX) VBX STENT GRAFT AND AN OVATION IX ILIAC LIMB. ADDITIONALLY, THE OTHER LIMB RECEIVED A GORE (NON-ENDOLOGIX) VBX STENT GRAFT (KISSING STENTS) TO MAINTAIN PATENCY. THE PATIENT'S STATUS WAS REPORTED AS STABLE FOLLOWING THE SECONDARY PROCEDURE.

    Patient problems: Obstruction/Occlusion

    MAUDE report 19030446 ↗
  • InjuryDec 14, 2022Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    THE PATIENT WAS INITIALLY TREATED FOR AN ABDOMINAL AORTIC ANEURYSM (AAA) WITH THE ENDOVASCULAR IMPLANT THE ALTO STENT GRAFT SYSTEM. APPROXIMATELY ONE (1) MONTH POST INITIAL PROCEDURE, A TYPE IIIA ENDOLEAK WAS IDENTIFIED AT THE ROUTINE FOLLOW-UP. REINTERVENTION INFORMATION HAS NOT BEEN PROVIDED. THE PATIENT STATUS HAS NOT BEEN PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 15980402 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P120006. One submission often covers many models, so check the product and lot details in each record.

  • Z-2263-2020 · May 6, 2020 · Terminated
    Ovation iX Abdominal Stent Graft Systems, REF: TV-AB2080-I, TV-AB2080-J, TV-AB2380-I, TV-AB2380-J, TV-AB2680-I, TV-AB2680-J, TV-AB2980-I, TV-AB2980-J, TV-AB3480-I, TV-AB3480-J
  • Z-0125-2015 · Sep 20, 2014 · Terminated
    Ovation Prime Abdominal Stent Graft System - 29mm Aortic Body stent grafts. The TriVascular Ovation Prime Abdominal Stent Graft System is a low-profile endovascular device deliver
  • Z-0288-2013 · Oct 19, 2012 · Terminated
    TriVascular brand Fill Polymer Kit, Endovascular Graft, Aneurysm Treatment System, European Model Number: TVFP14-B, (in European distributor inventory only); PRODUCT Usage: The

Other System, Endovascular Graft, Aortic Aneurysm Treatment devices

All System, Endovascular Graft, Aortic Aneurysm Treatment devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page