Ovation iX Iliac Stent Graft TV-IL1412120-JRecalls & reported problems
Endologix LLC · The Ovation iX™ Iliac Limbs are components of the Ovation Prime, Ovation iX, and Alto™ Abdominal Stent Graft Systems. The Ovation iX™ Iliac Limbs are comprised of a nitinol stent encapsulated in low-permeability PTFE. The iliac limbs are deployed into the leg sections of the aortic body. Additional iliac limbs may be used to extend the length of the device, where indicated. · Class 3 · Implant
All models of Ovation iX Iliac Stent Graft (66)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00850007370176), received between Aug 30, 2021 and Jul 22, 2022 — classified as 2 injuries.
The most frequently reported specific device problem is “Complete Blockage” (1 report).
3 other recalls cover devices cleared under the same FDA submission (P120006); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Complete Blockage1
- Partial Blockage1
Patient problems
- Obstruction/Occlusion1
- Stenosis1
- Unspecified Infection1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJul 22, 2022Complete Blockage; Complete Blockage
THE PATIENT WAS BEING TREATED FOR AN ABDOMINAL AORTIC ANEURYSM (AAA) WITH THE ALTO STENT GRAFT SYSTEM (ONE (1) ALTO MAIN BODY AND THREE (3) OVATION IX ILIAC LIMBS). POST-DEPLOYMENT, WHILE ATTEMPTING TO ADVANCE THE Q50 MERIT MEDICAL BALLOON TO MOLD THE RINGS, THE NOSE CONE OF THE BALLOON BECAME CAUGHT ON THE DISTAL SUPRARENAL PORTION OF THE ALTO MAIN BODY, PUSHING THE ALTO MAIN BODY PROXIMALLY ABOUT 1 CM AND PARTIALLY COVERING LEFT RENAL. ARM ACCESS WAS OBTAINED, AND THE LEFT RENAL WAS SUCCESSFULLY STENTED. THERE SEEMED TO BE SEAL IN THE NECK AT THE SECOND RING HOWEVER A TYPE 2 ENDOLEAK CONFIRMED TO STILL BE PRESENT. THIS EVENT HAS BEEN REPORTED UNDER MRN 3008011247-2022-00047. SEVERAL HOURS POST-PROCEDURE THE PATIENT DEVELOPED A COLD LEG AND WAS BROUGHT BACK INTO THE OPERATING ROOM. THE RIGHT LIMB WAS NOT GETTING FLOW DO TO THE ALTO MAIN BODY LEGS BECOMING A LITTLE TWISTED AND OCCLUDING ON THE RIGHT SIDE. TWO (2) NON-ENDOLOGIX BALLOON-EXPANDABLE STENTS WERE IMPLANTED AT THE LEVEL OF THE SECOND SEALING RING AND INTO THE LEGS OF THE MAIN BODY. NORMAL FLOW WAS ACHIEVED. THE FINAL ANGIOGRAM LOOKED PERFECT. THE PATIENT WAS REPORTED TO BE DOING WELL. THIS EVENT HAS BEEN REPORTED UNDER MR
Patient problems: Obstruction/Occlusion; Unspecified Infection
MAUDE report 15089208 ↗InjuryAug 30, 2021Partial Blockage; Partial Blockage
THE PATIENT WAS INITIALLY IMPLANTED WITH THE ALTO STENT GRAFT SYSTEM TO TREAT AN ABDOMINAL AORTIC ANEURYSM (AAA). APPROXIMATELY THREE (3) WEEKS POST INITIAL PROCEDURE THERE WAS STENOSIS WITHIN THE LEFT STENT GRAFT LIMB AND THE ILIAC LIMB WAS FOUND TO BE OCCLUDED. THE PATIENT UNDERWENT A LYSIS PROCEDURE FOLLOWING BY IMPLANT OF LUMINEXX (NON-ENDOLOGIX) BARE METAL STENTS TO OPEN THE LIMBS AT THE BIFURCATION OF THE GRAFT. THE PATIENT WAS REPORTED AS DISCHARGED POST THIS SECONDARY PROCEDURE.
Patient problems: Stenosis
MAUDE report 12388891 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P120006. One submission often covers many models, so check the product and lot details in each record.
- Z-2263-2020 · May 6, 2020 · TerminatedOvation iX Abdominal Stent Graft Systems, REF: TV-AB2080-I, TV-AB2080-J, TV-AB2380-I, TV-AB2380-J, TV-AB2680-I, TV-AB2680-J, TV-AB2980-I, TV-AB2980-J, TV-AB3480-I, TV-AB3480-J
- Z-0125-2015 · Sep 20, 2014 · TerminatedOvation Prime Abdominal Stent Graft System - 29mm Aortic Body stent grafts. The TriVascular Ovation Prime Abdominal Stent Graft System is a low-profile endovascular device deliver
- Z-0288-2013 · Oct 19, 2012 · TerminatedTriVascular brand Fill Polymer Kit, Endovascular Graft, Aneurysm Treatment System, European Model Number: TVFP14-B, (in European distributor inventory only); PRODUCT Usage: The
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page