Ovation iX Iliac Stent Graft TV-IL1428120-JRecalls & reported problems
Endologix LLC · The Ovation iX™ Iliac Limbs are components of the Ovation Prime, Ovation iX, and Alto™ Abdominal Stent Graft Systems. The Ovation iX™ Iliac Limbs are comprised of a nitinol stent encapsulated in low-permeability PTFE. The iliac limbs are deployed into the leg sections of the aortic body. Additional iliac limbs may be used to extend the length of the device, where indicated. · Class 3 · Implant
All models of Ovation iX Iliac Stent Graft (66)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00850007370190), received between May 6, 2024 and Sep 19, 2025 — classified as 3 injuries.
The most frequently reported specific device problem is “Migration” (2 reports).
3 other recalls cover devices cleared under the same FDA submission (P120006); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Migration2
- Off-Label Use1
- Unintended Movement1
Patient problems
- Clinical Endoleak2
- Aneurysm1
- Failure of Implant1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjurySep 19, 2025Off-Label Use; Off-Label Use
THE PATIENT WAS INITIALLY IMPLANTED WITH AN AFX2 BIFURCATED STENT GRAFT AND TWO OVATION IX ILIAC LIMB EXTENSIONS TO TREAT AN ABDOMINAL AORTIC ANEURYSM (AAA). APPROXIMATELY TWO (2) MONTHS POST INITIAL PROCEDURE, DURING A ROUTINE FOLLOW UP, A TYPE 1B ENDOLEAK ON THE LEFT COMMON ILIAC ARTERY WAS IDENTIFIED. THE PHYSICIAN ELECTED TO IMPLANT AN OVATION IX ILIAC LIMB TO SUCCESSFULLY RESOLVE THE TYPE 1B ENDOLEAK. THE PATIENT HAS BEEN DISCHARGED HOME. NOTE: THE INITIAL AND REINTERVENTION PROCEDURES ARE OUTSIDE THE INDICATIONS OF USE (OFF-LABEL) AS THE SAFETY AND EFFECTIVENESS OF OTHER STENT GRAFT DEVICES USED IN CONJUNCTION WITH THE AFX2 SYSTEM HAVE NOT BEEN ESTABLISHED.
Patient problems: Clinical Endoleak
MAUDE report 23096862 ↗InjuryAug 13, 2024Unintended Movement; Migration; Unintended Movement; Migration
THE PATIENT UNDERWENT IMPLANTATION WITH AN ALTO STENT GRAFT SYSTEM TO TREAT AN ABDOMINAL AORTIC ANEURYSM (AAA) ON (B)(6) 2023. DURING A FOLLOW-UP ON (B)(6) 2024, A SIGNIFICANT INCREASE IN THE ANEURYSM SAC SIZE AND A NON-DEVICE-RELATED TYPE 2 ENDOLEAK WERE IDENTIFIED. ADDITIONALLY, A 28MM OVATION LIMB WAS USED FOR THE RIGHT COMMON ILIAC ARTERY (CIA) LIMB, ALTHOUGH THE AVERAGE DIAMETER WAS ONLY 15MM. THIS SIZE MISMATCH CAUSED THE LIMB TO "MELON-SEED" INTO THE DISTAL AAA SAC, RESULTING IN A TYPE 1B ENDOLEAK. A NEW LIMB IS REQUIRED TO ADDRESS THIS ISSUE. ON (B)(6) 2024, THE PATIENT UNDERWENT A RE-INTERVENTION WHERE THE PHYSICIAN ELECTED TO IMPLANT AN OVATION IX EXTENDER AND AN OVATION IX ILIAC LIMB. THE FINAL PATIENT STATUS REMAINS UNKNOWN AND WAS NOT MADE AVAILABLE TO ENDOLOGIX.
Patient problems: Aneurysm; Clinical Endoleak
MAUDE report 19975707 ↗InjuryMay 6, 2024Migration; Migration
THE PATIENT WAS TREATED FOR AN ABDOMINAL AORTIC ANEURYSM (AAA) WITH THE IMPLANTATION OF THE ALTO ABDOMINAL STENT GRAFT SYSTEM ON (B)(6) 2024. DURING A ROUTINE FOLLOW-UP ON (B)(6) 2024, VIA COMPUTED TOMOGRAPHY (CT) SCAN, IT WAS NOTED THAT THE PATIENT'S LEFT ILIAC LIMB HAD MIGRATED DISTALLY FROM THE LEFT INTERNAL ARTERY, NEARING THE SAC. ON (B)(6) 2024, THE PATIENT UNDERWENT RE-INTERVENTION WITH THE IMPLANTATION OF AN ADDITIONAL OVATION IX ILIAC LIMB. THE LIMB WAS PLACED FROM THE MAIN BODY LIMB DOWN TO THE DISTAL COMMON ILIAC. THE FINAL PATIENT STATUS WAS REPORTED AS STABLE.
Patient problems: Failure of Implant
MAUDE report 19256228 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P120006. One submission often covers many models, so check the product and lot details in each record.
- Z-2263-2020 · May 6, 2020 · TerminatedOvation iX Abdominal Stent Graft Systems, REF: TV-AB2080-I, TV-AB2080-J, TV-AB2380-I, TV-AB2380-J, TV-AB2680-I, TV-AB2680-J, TV-AB2980-I, TV-AB2980-J, TV-AB3480-I, TV-AB3480-J
- Z-0125-2015 · Sep 20, 2014 · TerminatedOvation Prime Abdominal Stent Graft System - 29mm Aortic Body stent grafts. The TriVascular Ovation Prime Abdominal Stent Graft System is a low-profile endovascular device deliver
- Z-0288-2013 · Oct 19, 2012 · TerminatedTriVascular brand Fill Polymer Kit, Endovascular Graft, Aneurysm Treatment System, European Model Number: TVFP14-B, (in European distributor inventory only); PRODUCT Usage: The
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page