Ovation iX Iliac Stent Graft TV-IL1418100-JRecalls & reported problems
Endologix LLC · The Ovation iX™ Iliac Limbs are components of the Ovation Prime, Ovation iX, and Alto™ Abdominal Stent Graft Systems. The Ovation iX™ Iliac Limbs are comprised of a nitinol stent encapsulated in low-permeability PTFE. The iliac limbs are deployed into the leg sections of the aortic body. Additional iliac limbs may be used to extend the length of the device, where indicated. · Class 3 · Implant
All models of Ovation iX Iliac Stent Graft (66)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00850007370282), received between Apr 17, 2025 and Apr 17, 2025 — classified as 1 injury.
The most frequently reported specific device problem is “Lack of Effect” (1 report).
3 other recalls cover devices cleared under the same FDA submission (P120006); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Lack of Effect1
Patient problems
- Aneurysm1
- Clinical Endoleak1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryApr 17, 2025Lack of Effect; Lack of Effect
THE PATIENT WAS INITIALLY TREATED FOR AN ABDOMINAL AORTIC ANEURYSM (AAA) ON (B)(6) 2022 WITH THE IMPLANT OF AN ALTO STENT GRAFT SYSTEM. AFTER EMBOLIZATION OF THE LUMBAR AND INTERNAL ILIAC ARTERIES, THE ALTO MAIN BODY WAS DEPLOYED ALIGNED WITH THE RIGHT RENAL ARTERY. AFTER POLYMER FILLING, THE CONTRALATERAL LEG WAS DEPLOYED. FOURTEEN MINUTES AFTER DEPLOYMENT, THE INTERNAL BALLOON WAS EXPANDED, AND ANGIOGRAPHY CONFIRMED THE ABSENCE OF A TYPE I ENDOLEAK. THE IPSILATERAL LEG WAS DEPLOYED, AND AFTER TOUCH-UP, ANGIOGRAPHY CONFIRMED THE ABSENCE OF AN ENDOLEAK, AND THE PROCEDURE WAS COMPLETED. NO ANEURYSM DIAMETER ENLARGEMENT WAS OBSERVED UNTIL THE 1-YEAR FOLLOW-UP. ON THE 2-YEAR FOLLOW-UP ON (B)(6) 2024, AN ANEURYSM DIAMETER ENLARGEMENT OF 5 MM OR MORE WAS OBSERVED. THE TYPE OF ENDOLEAK WAS UNKNOWN, UNTIL IDENTIFIED AS A TYPE IB ENDOLEAK DURING REINTERVENTION WHICH WAS COMPLETED ON (B)(6) 2025. THE PHYSICIAN ELECTED TO IMPLANT A GORE (NON-ENDOLOGIX) EXCLUDER ILIAC BRANCH ENDOPROSTHESIS TO RESOLVE THIS EVENT. THE FINAL PATIENT STATUS WAS NOT REPORTED.
Patient problems: Aneurysm; Clinical Endoleak
MAUDE report 21862181 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P120006. One submission often covers many models, so check the product and lot details in each record.
- Z-2263-2020 · May 6, 2020 · TerminatedOvation iX Abdominal Stent Graft Systems, REF: TV-AB2080-I, TV-AB2080-J, TV-AB2380-I, TV-AB2380-J, TV-AB2680-I, TV-AB2680-J, TV-AB2980-I, TV-AB2980-J, TV-AB3480-I, TV-AB3480-J
- Z-0125-2015 · Sep 20, 2014 · TerminatedOvation Prime Abdominal Stent Graft System - 29mm Aortic Body stent grafts. The TriVascular Ovation Prime Abdominal Stent Graft System is a low-profile endovascular device deliver
- Z-0288-2013 · Oct 19, 2012 · TerminatedTriVascular brand Fill Polymer Kit, Endovascular Graft, Aneurysm Treatment System, European Model Number: TVFP14-B, (in European distributor inventory only); PRODUCT Usage: The
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page