Skip to content

Ovation iX Iliac Stent Graft TV-IL1422160-JRecalls & reported problems

Endologix LLC · The Ovation iX™ Iliac Limbs are components of the Ovation Prime, Ovation iX, and Alto™ Abdominal Stent Graft Systems. The Ovation iX™ Iliac Limbs are comprised of a nitinol stent encapsulated in low-permeability PTFE. The iliac limbs are deployed into the leg sections of the aortic body. Additional iliac limbs may be used to extend the length of the device, where indicated. · Class 3 · Implant

All models of Ovation iX Iliac Stent Graft (66)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00850007370350), received between Mar 8, 2023 and Nov 21, 2025 — classified as 3 injuries.

The most frequently reported specific device problem is “Lack of Effect” (2 reports).

3 other recalls cover devices cleared under the same FDA submission (P120006); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Lack of Effect2
  • Appropriate Device Problem Term/Code Not Available1
  • Migration1
  • Off-Label Use1

Patient problems

  • Clinical Endoleak3
  • Abdominal Pain1
  • Aneurysm1
  • Pseudoaneurysm1
  • Ruptured Aneurysm1
  • Unspecified Infection1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 21, 2025Migration; Lack of Effect; Migration; Lack of Effect

    THE PATIENT WAS INITIALLY TREATED FOR AN ABDOMINAL AORTIC ANEURYSM (AAA) ON (B)(6) 2025 WITH THE IMPLANT OF AN ALTO ABDOMINAL STENT GRAFT SYSTEM AND AN OVATION IX EXTENDER. THE PATIENT PRESENTED EMERGENTLY ON (B)(6) 2025 WITH ABDOMINAL PAIN AND A RUPTURED ANEURYSM. REPORTEDLY THE OVATION IX LIMB PULLED BACK INTO THE AORTA CAUSING A TYPE IB ENDOLEAK IN THE RIGHT COMMON ILIAC ARTERY. REINTERVENTION WAS COMPLETED ON (B)(6) 2025 WITH THE IMPLANT OF ANOTHER OVATION IX LIMB TO EXTEND TO THE RIGHT HYPOGASTRIC. THE TYPE IB ENDOLEAK ON THE RIGHT SIDE WAS SEALED. THE PATIENTS ABDOMINAL PAIN SUBSIDED AND WAS SYMPTOM FREE. ONE WEEK LATER ON (B)(6) 2025 THE PATIENT WAS BROUGHT BACK IN TO COMPLETE AN EXTENSION ON THE LEFT COMMON ILIAC ARTERY (LCIA). THERE WAS NO ENDOLEAK PRESENT IN THE LCIA BEFORE THE PROCEDURE. THE PHYSICIAN ELECTED TO IMPLANT AN OVATION IX LIMB TO EXTEND THE LCIA TO THE LEFT HYPOGASTRIC AS A PREVENTATIVE MEASURE. THE FINAL PATIENT STATUS WAS REPORTED TO BE WELL POST-SECONDARY PROCEDURES.

    Patient problems: Abdominal Pain; Clinical Endoleak; Ruptured Aneurysm

    MAUDE report 23618811 ↗
  • InjuryMar 5, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    THE PATIENT WAS TREATED FOR AN ABDOMINAL AORTIC ANEURYSM (AAA) WITH THE IMPLANT OF THE ALTO ABDOMINAL STENT GRAFT SYSTEM ON (B)(6) 2024. IT WAS REPORTED ON (B)(6) 2025, DURING A ROUTINE FOLLOW UP, THE PATIENT PRESENTED WITH A TYPE IB ENDOLEAK. AN INTERVENTION IS PLANNED FOR (B)(6) 2025. THE PATIENT STATUS IS STABLE.

    Patient problems: Clinical Endoleak

    MAUDE report 21528170 ↗
  • InjuryMar 8, 2023Off-Label Use; Lack of Effect; Off-Label Use; Lack of Effect

    THE PATIENT WAS INITIALLY TREATED FOR AN ABDOMINAL AORTIC ANEURYSM (AAA) WITH IMPLANT OF THE ALTO ABDOMINAL AORTIC ANEURYSM STENT SYSTEM. APPROXIMATELY THREE (3) MONTHS POST INITIAL PROCEDURE, THE PATIENT PRESENTED WITH A TYPE IB ENDOLEAK, AN INCREASED ANEURYSM SAC (UNKNOWN DIAMETER) OF THE LEFT EXTERNAL ILIAC ARTERY (LEIA), A CONTAINED PSEUDOANEURYSM AND GRAFT INFECTION. THE PATIENT HAS BEEN ADMINISTERED ANTIBIOTICS. THE PHYSICIAN ELECTED TO TREAT THE PATIENT BY COILING THE INTERNAL ILIAC ARTERY AND EXTENDING THE ILIAC STENT INTO THE LEIA WITH TWO (2) ADDITIONAL OVATION IX ILIAC LIMBS ON 21 FEBRUARY 2023 (IN ORDER TO EXCLUDE THE PERSISTENT SCIATIC ARTERY). BLOOD CULTURES WERE ALSO REPEATED, AND PATIENT IS REPORTEDLY IN STABLE CONDITION. THE PHYSICIANS PLANS TO FURTHER TREAT BY CONVERTING PATIENT TO OPEN REPAIR WITH EXPLANTATION OF THE GRAFT, BUT THIS SURGERY HAS NOT BEEN SCHEDULED.

    Patient problems: Aneurysm; Clinical Endoleak; Pseudoaneurysm; Unspecified Infection

    MAUDE report 16503518 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P120006. One submission often covers many models, so check the product and lot details in each record.

  • Z-2263-2020 · May 6, 2020 · Terminated
    Ovation iX Abdominal Stent Graft Systems, REF: TV-AB2080-I, TV-AB2080-J, TV-AB2380-I, TV-AB2380-J, TV-AB2680-I, TV-AB2680-J, TV-AB2980-I, TV-AB2980-J, TV-AB3480-I, TV-AB3480-J
  • Z-0125-2015 · Sep 20, 2014 · Terminated
    Ovation Prime Abdominal Stent Graft System - 29mm Aortic Body stent grafts. The TriVascular Ovation Prime Abdominal Stent Graft System is a low-profile endovascular device deliver
  • Z-0288-2013 · Oct 19, 2012 · Terminated
    TriVascular brand Fill Polymer Kit, Endovascular Graft, Aneurysm Treatment System, European Model Number: TVFP14-B, (in European distributor inventory only); PRODUCT Usage: The

Other System, Endovascular Graft, Aortic Aneurysm Treatment devices

All System, Endovascular Graft, Aortic Aneurysm Treatment devices →

Sources and updates