The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00850008449826), received between May 16, 2024 and Jun 27, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 29, 2024Open, Classified

    This product is being recalled due to potentially insufficient rework on a part (E7130-B) on the lumbar section of the frame on 6 distributed tables. There were 4 additional tables with the faulty E7130-B part that were manufactured, but had not been distributed yet.

    Z-2326-2024 · Root cause: Nonconforming Material/Component · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component3

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 27, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    A WELD ON PART E7130 FAILED WHILE THE TABLE WAS IN USE. THERE WAS NO INJURY, HOWEVER THE DOCTOR'S FLOOR WAS DAMAGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19629993 ↗
  • MalfunctionMay 17, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER CONTACTED THE MANUFACTURER TO REPORT THAT AS THEY WERE UNCRATING THE TABLE AND WERE IN THE PRCOESS OF MOVING THE TABLE, THE LUMBAR SECTION OF THE TABLE WAS OBSERVED AS NOT BEING HORIZONTAL AND LEVEL WITH THE OTHER CUSHION SECTIONS, IT WAS TIPPED DOWNARD. PICTURES WERE SENT BY THE CUSTOMER TO THE MANUFACTURER AND IT WAS DETERMINED THAT THERE MAY BE A PARTIAL WELD BREAK AND THE IMMEDIATE DECISION WAS MADE TO HAVE THE TABLE SHIPPED BACK TO THE MANUFACTURER. THERE WAS NO PATIENT CONTACT WITH THE DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19342108 ↗
  • MalfunctionMay 16, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THE WELD ON THE LUMBAR SECTION OF THE TABLE FAILED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19336621 ↗