Skip to content

APRO 70 Swift Catheter FG-02133-132-ACRecalls & reported problems

Alembic, LLC · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00850014025267), received between Aug 7, 2025 and Mar 3, 2026 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Material Deformation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Deformation2
  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Device Damaged by Another Device1
  • Material Integrity Problem1
  • Physical Resistance/Sticking1
  • Stretched1

Patient problems

  • Intracranial Hemorrhage1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 3, 2026Break; Device Damaged by Another Device; Material Deformation; Material Integrity Problem; Adverse Event Without Identified Device or Use Problem

    ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE APRO CATHETER WAS BEING EXTRACTED FROM THE INTRODUCER SHEATH IT STARTED TO UNRAVEL AND HAD PLASTIC COMING WITH IT. THE EVENTS LEADING UP TO THE PROCEDURE INCIDENT WERE AS FOLLOWS: THE BALT BALLAST INTRODUCER SHEATH WAS INTRODUCED INTO THE RIGHT ICA, THE APRO CATHETER WAS INTRODUCED THROUGH THE GUIDE SHEATH, THEN THE SOLITAIRE WAS INTRODUCED INTO THE SHEATH. THE SOLITAIRE WAS UNABLE TO BE RETRIEVED THROUGH THE INTRODUCER SHEATH. ALL ITEMS HAD TO BE REMOVED TOGETHER. UPON REMOVAL OF THE ITEMS IN ONE UNIT IT WAS NOTICED THE APRO WAS FRAYED. THE PATIENT SUFFERED DAMAGE FROM THE FRAYED APRO AND WAS IN CRITICAL CONDITION. THIS CONTRADICTS THE EARLIER REPORT THAT THE APRO APPEARED TO BE INTACT.

    Patient problems: Intracranial Hemorrhage

    MAUDE report 24498406 ↗
  • MalfunctionAug 7, 2025Stretched; Material Deformation; Physical Resistance/Sticking; Stretched; Material Deformation; Physical Resistance/Sticking

    ADDITIONAL INFORMATION RECEIVED REPORTED IT IS ASSUMED THAT THE COPILOT TOUGHY WASN'T PUSHED IN ENOUGH CAUSING THE STRETCHING. POST PROCEDURE TICI WAS 2B. CONTINUOUS SALINE FLUSH ADMINISTERED AND MAINTAINED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22742238 ↗

Other Catheter, Thrombus Retriever devices

All Catheter, Thrombus Retriever devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page