Skip to content

Remunity DKPI-11036-001Recalls & reported problems

United Therapeutics Corporation · OVER PACK, DISPOSABLE KIT, UNITY 00850017421073 package contains: 10 packs of 12ea. disposable reservoirs · Class 2

All models of Remunity (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00850017421073), received between Jan 8, 2025 and Aug 27, 2026 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component3
  • Break2
  • Difficult or Delayed Separation1
  • Fluid/Blood Leak1
  • Insufficient Device Problem Information1

Patient problems

  • Insufficient Information3
  • No Clinical Signs, Symptoms or Conditions2
  • Anxiety1
  • Fatigue1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 27, 2026Detachment of Device or Device Component

    An initial event notification was received by United Therapeutics Drug Safety on 28-JUL-2026 from CVS Specialty Pharmacy, and forwarded to Millyard Advanced Medical Products, LLC, on 29-JUL-2026.The patient reported that on two separate occasions, the tubing detached from their Remunity Cassettes while they were in use and the Pumps were infusing. The patient stated this first occurred while they were sleeping and then again while they were at work. The patient confirmed they were able to promptly replace the Cassette each time and they did not have any delays in therapy or experience any symptoms.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50485163 ↗
  • MalfunctionJun 26, 2026Detachment of Device or Device Component

    AN INITIAL EVENT NOTIFICATION WAS RECEIVED BY UNITED THERAPEUTICS DRUG SAFETY ON 03-JUN-2026 FROM ACCREDO SPECIALTY PHARMACY AND FORWARDED TO MILLYARD ADVANCED MEDICAL PRODUCTS, LLC ON 04-JUN-2026. IT WAS REPORTED THAT THE TUBING ON THE PATIENT'S REMUNITY CASSETTE HAD "BROKEN" AND THEY LOST REMODULIN OVERNIGHT. THE PATIENT FURTHER REPORTED THAT THIS ISSUE HAD OCCURRED 2-3 TIMES PRIOR, ALWAYS AT NIGHT WHILE THEY WERE SLEEPING. THE PATIENT CHANGED THE CASSETTES TO RESOLVE THE ISSUE.

    Patient problems: Insufficient Information

    MAUDE report 25641914 ↗
  • MalfunctionJun 10, 2026Detachment of Device or Device Component

    AN INITIAL EVENT NOTIFICATION WAS RECEIVED BY UNITED THERAPEUTICS DRUG SAFETY ON 11-MAY-2026 FROM ACCREDO SPECIALTY PHARMACY, INC. AND FORWARDED TO MILLYARD ADVANCED MEDICAL PRODUCTS, LLC ON 12-MAY-2026. THE PATIENT REPORTED THAT WHILE PERFORMING A SCHEDULED CASSETTE CHANGE, THEY NOTICED THAT THE TUBING HAD DETACHED FROM THE REMUNITY CASSETTE WHILE THE PUMP HAD BEEN IN THEIR POCKET. THE PATIENT OBSERVED THAT THE CASSETTE APPEARED TO BE FULL AND STATED THEY HAD POSSIBLY BEEN WITHOUT REMODULIN FOR TWO DAYS. THE PATIENT EXPERIENCED FATIGUE AT THE TIME OF THE EVENT.

    Patient problems: Fatigue

    MAUDE report 25462918 ↗
  • MalfunctionApr 7, 2026Break

    AN INITIAL EVENT NOTIFICATION WAS RECEIVED BY UNITED THERAPEUTICS DRUG SAFETY ON 09-MAR-2026 FROM CVS SPECIALTY PHARMACY, AND FORWARDED TO MILLYARD ADVANCED MEDICAL PRODUCTS, LLC ON 10-MAR-2026. IT WAS REPORTED THAT WHILE THE PATIENT WAS PREPARING A NEW CASSETTE, IT "FELL APART." THE PATIENT MADE A SECOND ATTEMPT WITH THEIR LAST CASSETTE BUT REPORTED THAT THE SAME ISSUE OCCURRED. INFORMATION REGARDING HOW THE CASSETTES "FELL APART" WAS NOT CONFIRMED BY THE SPECIALTY PHARMACY. IT WAS FURTHER REPORTED THAT THE PATIENT HAD NO ADDITIONAL SUPPLIES ON HAND AND EXPERIENCED AN INTERRUPTION IN THERAPY. THE PATIENT ULTIMATELY PRESENTED TO THE HOSPITAL.

    Patient problems: Insufficient Information

    MAUDE report 24809297 ↗
  • MalfunctionFeb 27, 2026Break

    AN INITIAL EVENT NOTIFICATION WAS RECEIVED BY UNITED THERAPEUTICS DRUG SAFETY ON 29-JAN-2026 FROM CVS SPECIALTY PHARMACY, AND FORWARDED TO MILLYARD ADVANCED MEDICAL PRODUCTS, LLC ON 30-JAN-2026. IT WAS REPORTED THAT THE PATIENT'S CASSETTE BROKE WHEN THEIR REMUNITY PUMP FELL. THE PATIENT DID NOT HAVE ANY MORE CASSETTES AVAILABLE AND EXPERIENCED AN INTERRUPTION IN THERAPY. IT WAS LATER REPORTED THAT THE PATIENT WAS HOSPITALIZED AND RESTARTED ON REMODULIN FROM THE HOSPITAL'S SUPPLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24470735 ↗
  • MalfunctionDec 10, 2025Fluid/Blood Leak; Difficult or Delayed Separation; Fluid/Blood Leak; Difficult or Delayed Separation

    AN INITIAL EVENT NOTIFICATION WAS RECEIVED BY UNITED THERAPEUTICS DRUG SAFETY ON 10-NOV-2025 FROM ACCREDO HEALTH GROUP AND FORWARDED TO MILLYARD ADVANCED MEDICAL PRODUCTS, LLC, ON 11-NOV-2025. THE PATIENT REPORTED BEING OUT OF REMODULIN SINCE 1:00 AM ON (B)(6) 2025. IT WAS COMMUNICATED THAT THE PATIENT'S MEDICATION VIAL WAS LEAKING, AND THAT THEIR LAST REMUNITY CASSETTE HAD LEAKED WHILE FILLING. THE PATIENT ALSO REPORTED THAT THEIR SON HAD TO DISCONNECT THEIR TUBING, BUT THEY HAD NO ADDITIONAL SUPPLIES ON HAND. ACCREDO HEALTH GROUP REPORTED THAT NO REMUNITY PRODUCT IS AVAILABLE FOR RETURN. REPORTABLE INFORMATION RELATED TO THIS EVENT WAS LATER RECEIVED BY UNITED THERAPEUTICS DRUG SAFETY ON 10-NOV-2025 FROM ACCREDO HEALTH GROUP AND FORWARDED TO MILLYARD ADVANCED MEDICAL PRODUCTS, LLC, ON 14-NOV-2025. THIS INFORMATION INDICATED THAT THE PATIENT WAS ADMITTED TO THE HOSPITAL ON (B)(6) 2025. ATTEMPTS MADE TO OBTAIN ADDITIONAL INFORMATION FROM ACCREDO HEALTH GROUP WERE UNSUCCESSFUL.

    Patient problems: Insufficient Information

    MAUDE report 23771648 ↗
  • MalfunctionJan 8, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    AN INITIAL REPORT OF HOSPITALIZATION WAS RECEIVED BY UNITED THERAPEUTICS DRUG SAFETY ON 16-DEC-2024 AND FORWARDED TO MILLYARD ADVANCED MEDICAL PRODUCTS, LLC ON 17-DEC-2024. THE PATIENT REPORTED THAT SHE WAS UNABLE TO FILL HER CASSETTES AND TROUBLESHOOTING WAS UNSUCCESSFUL. THE PATIENT WAS OFF REMODULIN FOR TWO HOURS AND WAS FEELING ANXIOUS BUT HAD NO OTHER SIDE EFFECTS. THE PATIENT WAS ADMITTED TO THE HOSPITAL. THERE WERE NO REPORTED PUMP ISSUES. A REPLACEMENT VIAL OF REMODULIN WAS SENT. THE PATIENT WAS ABLE TO RESUME REMUNITY THERAPY AND WAS SCHEDULED TO BE DISCHARGED. NO COMPONENTS OR FURTHER INFORMATION ASSOCIATED WITH THE REPORTED EVENT HAVE BEEN MADE AVAILABLE TO THE MANUFACTURER FOR FURTHER EVALUATION. DESPITE NO REPORT OF DEATH OR SERIOUS INJURY, THIS ISSUE IS BEING REPORTED OUT OF AN ABUNDANCE OF CAUTION.

    Patient problems: Anxiety

    MAUDE report 21103391 ↗

FDA clearance and classification

The FDA regulates product code QJY (infusion pump, drug specific, pharmacy-filled) under 21 CFR 880.5725 as a Class II device. 21 CFR 880.5725 ↗

Devices of this type normally need 510(k) clearance before sale.

FDA definition of this device type: A Pharmacy-Filled, Drug Specific Infusion System is a prescription device intended for delivery of a specific drug in accordance with the FDA approved labeling and may be filled at a location other than point of care.

Common questions about the Remunity DKPI-11036-001

Has the Remunity DKPI-11036-001 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00850017421073), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Remunity DKPI-11036-001?

The FDA's MAUDE database lists 7 adverse event reports for this device, received between Jan 8, 2025 and Aug 27, 2026. In those reports, the most common specific device problems are “Detachment of Device or Device Component” (3), “Break” (2) and “Difficult or Delayed Separation” (1); the most common patient problems are “Anxiety” (1) and “Fatigue” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Remunity DKPI-11036-001?

None of the reports are classified as deaths, 0 as injuries and 7 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Remunity DKPI-11036-001 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the Remunity DKPI-11036-001, and is it still sold?

The FDA's device database lists United Therapeutics Corporation as the company responsible for it (brand Remunity, model DKPI-11036-001). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Infusion Pump, Drug Specific, Pharmacy-Filled devices

All Infusion Pump, Drug Specific, Pharmacy-Filled devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page