DermaSensor 10101Recalls & reported problems

Dermasensor, Inc. · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00850017786004), received between Dec 29, 2025 and Feb 9, 2026 — classified as 11 malfunctions.

The most frequently reported specific device problem is “Manufacturing, Packaging or Shipping Problem” (11 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 7 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Manufacturing, Packaging or Shipping Problem11
  • Nonstandard Device11
  • Product Quality Problem11

Patient problems

  • No Clinical Signs, Symptoms or Conditions11

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 9, 2026Nonstandard Device; Product Quality Problem; Manufacturing, Packaging or Shipping Problem

    THE COMPLAINT IS FOR DEVICE S/N (B)(6) WHERE THE CORRECT, UNIQUE SPECTROMETER IDS AND THEIR ASSOCIATED COEFFICIENTS WERE NOT CORRECTLY CONFIGURED. THE INSTALLED SPECTROMETER DID NOT CORRECTLY ANALYZE THE WAVELENGTHS OF LIGHT WHICH RESULTED IN LOWER DEVICE RESULT SCORES, WHICH MAY IMPACT THE CLINICIAN'S CARE AND RESULT IN DELAY OF REFERRAL TO A DERMATOLOGIST. THERE IS NO ASSERTION OF DISCRETE PATIENT HARM. THE DEVICE IS AN ADJUNCTIVE DIAGNOSTIC DEVICE THAT IS INTENDED TO PROVIDE ADDITIONAL INFORMATION TO CLINICIANS, AIDING IN MANAGEMENT BUT NOT USED INDEPENDENTLY TO DETERMINE MANAGEMENT, DIAGNOSIS NOR THERAPY. DERMASENSOR DEVICE SHOULD BE USED IN CONJUNCTION WITH THE TOTALITY OF CLINICALLY RELEVANT INFORMATION FROM THE CLINICAL ASSESSMENT IN ACCORDANCE WITH THE DEVICE INSTRUCTIONS FOR USE, INCLUDING VISUAL ANALYSIS OF THE LESION, BY PHYSICIANS WHO ARE NOT DERMATOLOGISTS. THE DEVICE SHOULD BE USED ON LESIONS ALREADY ASSESSED AS SUSPICIOUS FOR SKIN CANCER AND NOT AS A SCREENING TOOL. THE DEVICE SHOULD NOT BE USED AS THE SOLE DIAGNOSTIC CRITERION NOR TO CONFIRM CLINICAL DIAGNOSIS OF SKIN CANCER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24298721 ↗
  • MalfunctionFeb 2, 2026Nonstandard Device; Product Quality Problem; Manufacturing, Packaging or Shipping Problem

    THE COMPLAINT IS FOR DEVICE S/N (B)(6) WHERE THE CORRECT, UNIQUE SPECTROMETER IDS AND THEIR ASSOCIATED COEFFICIENTS WERE NOT CORRECTLY CONFIGURED. THE INSTALLED SPECTROMETER DID NOT CORRECTLY ANALYZE THE WAVELENGTHS OF LIGHT WHICH RESULTED IN LOWER DEVICE RESULT SCORES, WHICH MAY IMPACT THE CLINICIAN'S CARE AND RESULT IN DELAY OF REFERRAL TO A DERMATOLOGIST. THERE IS NO ASSERTION OF DISCRETE PATIENT HARM. THE DEVICE IS AN ADJUNCTIVE DIAGNOSTIC DEVICE THAT IS INTENDED TO PROVIDE ADDITIONAL INFORMATION TO CLINICIANS, AIDING IN MANAGEMENT BUT NOT USED INDEPENDENTLY TO DETERMINE MANAGEMENT, DIAGNOSIS NOR THERAPY. DERMASENSOR DEVICE SHOULD BE USED IN CONJUNCTION WITH THE TOTALITY OF CLINICALLY RELEVANT INFORMATION FROM THE CLINICAL ASSESSMENT IN ACCORDANCE WITH THE DEVICE INSTRUCTIONS FOR USE, INCLUDING VISUAL ANALYSIS OF THE LESION, BY PHYSICIANS WHO ARE NOT DERMATOLOGISTS. THE DEVICE SHOULD BE USED ON LESIONS ALREADY ASSESSED AS SUSPICIOUS FOR SKIN CANCER AND NOT AS A SCREENING TOOL. THE DEVICE SHOULD NOT BE USED AS THE SOLE DIAGNOSTIC CRITERION NOR TO CONFIRM CLINICAL DIAGNOSIS OF SKIN CANCER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24239823 ↗
  • MalfunctionJan 27, 2026Nonstandard Device; Product Quality Problem; Manufacturing, Packaging or Shipping Problem

    THE COMPLAINT IS FOR DEVICE S/N (B)(6) WHERE THE CORRECT, UNIQUE SPECTROMETER IDS AND THEIR ASSOCIATED COEFFICIENTS WERE NOT CORRECTLY CONFIGURED. THE INSTALLED SPECTROMETER DID NOT CORRECTLY ANALYZE THE WAVELENGTHS OF LIGHT WHICH RESULTED IN LOWER DEVICE RESULT SCORES, WHICH MAY IMPACT THE CLINICIAN'S CARE AND RESULT IN DELAY OF REFERRAL TO A DERMATOLOGIST. THERE IS NO ASSERTION OF DISCRETE PATIENT HARM. THE DEVICE IS AN ADJUNCTIVE DIAGNOSTIC DEVICE THAT IS INTENDED TO PROVIDE ADDITIONAL INFORMATION TO CLINICIANS, AIDING IN MANAGEMENT BUT NOT USED INDEPENDENTLY TO DETERMINE MANAGEMENT, DIAGNOSIS NOR THERAPY. DERMASENSOR DEVICE SHOULD BE USED IN CONJUNCTION WITH THE TOTALITY OF CLINICALLY RELEVANT INFORMATION FROM THE CLINICAL ASSESSMENT IN ACCORDANCE WITH THE DEVICE INSTRUCTIONS FOR USE, INCLUDING VISUAL ANALYSIS OF THE LESION, BY PHYSICIANS WHO ARE NOT DERMATOLOGISTS. THE DEVICE SHOULD BE USED ON LESIONS ALREADY ASSESSED AS SUSPICIOUS FOR SKIN CANCER AND NOT AS A SCREENING TOOL. THE DEVICE SHOULD NOT BE USED AS THE SOLE DIAGNOSTIC CRITERION NOR TO CONFIRM CLINICAL DIAGNOSIS OF SKIN CANCER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24196388 ↗
  • MalfunctionJan 27, 2026Nonstandard Device; Product Quality Problem; Manufacturing, Packaging or Shipping Problem

    THE COMPLAINT IS FOR DEVICE S/N (B)(6) WHERE THE CORRECT, UNIQUE SPECTROMETER IDS AND THEIR ASSOCIATED COEFFICIENTS WERE NOT CORRECTLY CONFIGURED. THE INSTALLED SPECTROMETER DID NOT CORRECTLY ANALYZE THE WAVELENGTHS OF LIGHT WHICH RESULTED IN LOWER DEVICE RESULT SCORES, WHICH MAY IMPACT THE CLINICIAN'S CARE AND RESULT IN DELAY OF REFERRAL TO A DERMATOLOGIST. THERE IS NO ASSERTION OF DISCRETE PATIENT HARM. THE DEVICE IS AN ADJUNCTIVE DIAGNOSTIC DEVICE THAT IS INTENDED TO PROVIDE ADDITIONAL INFORMATION TO CLINICIANS, AIDING IN MANAGEMENT BUT NOT USED INDEPENDENTLY TO DETERMINE MANAGEMENT, DIAGNOSIS NOR THERAPY. DERMASENSOR DEVICE SHOULD BE USED IN CONJUNCTION WITH THE TOTALITY OF CLINICALLY RELEVANT INFORMATION FROM THE CLINICAL ASSESSMENT IN ACCORDANCE WITH THE DEVICE INSTRUCTIONS FOR USE, INCLUDING VISUAL ANALYSIS OF THE LESION, BY PHYSICIANS WHO ARE NOT DERMATOLOGISTS. THE DEVICE SHOULD BE USED ON LESIONS ALREADY ASSESSED AS SUSPICIOUS FOR SKIN CANCER AND NOT AS A SCREENING TOOL. THE DEVICE SHOULD NOT BE USED AS THE SOLE DIAGNOSTIC CRITERION NOR TO CONFIRM CLINICAL DIAGNOSIS OF SKIN CANCER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24196308 ↗
  • MalfunctionJan 27, 2026Nonstandard Device; Product Quality Problem; Manufacturing, Packaging or Shipping Problem

    THE COMPLAINT IS FOR DEVICE S/N (B)(6) WHERE THE CORRECT, UNIQUE SPECTROMETER IDS AND THEIR ASSOCIATED COEFFICIENTS WERE NOT CORRECTLY CONFIGURED. THE INSTALLED SPECTROMETER DID NOT CORRECTLY ANALYZE THE WAVELENGTHS OF LIGHT WHICH RESULTED IN LOWER DEVICE RESULT SCORES, WHICH MAY IMPACT THE CLINICIAN'S CARE AND RESULT IN DELAY OF REFERRAL TO A DERMATOLOGIST. THERE IS NO ASSERTION OF DISCRETE PATIENT HARM. THE DEVICE IS AN ADJUNCTIVE DIAGNOSTIC DEVICE THAT IS INTENDED TO PROVIDE ADDITIONAL INFORMATION TO CLINICIANS, AIDING IN MANAGEMENT BUT NOT USED INDEPENDENTLY TO DETERMINE MANAGEMENT, DIAGNOSIS NOR THERAPY. DERMASENSOR DEVICE SHOULD BE USED IN CONJUNCTION WITH THE TOTALITY OF CLINICALLY RELEVANT INFORMATION FROM THE CLINICAL ASSESSMENT IN ACCORDANCE WITH THE DEVICE INSTRUCTIONS FOR USE, INCLUDING VISUAL ANALYSIS OF THE LESION, BY PHYSICIANS WHO ARE NOT DERMATOLOGISTS. THE DEVICE SHOULD BE USED ON LESIONS ALREADY ASSESSED AS SUSPICIOUS FOR SKIN CANCER AND NOT AS A SCREENING TOOL. THE DEVICE SHOULD NOT BE USED AS THE SOLE DIAGNOSTIC CRITERION NOR TO CONFIRM CLINICAL DIAGNOSIS OF SKIN CANCER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24196236 ↗
  • MalfunctionJan 27, 2026Nonstandard Device; Product Quality Problem; Manufacturing, Packaging or Shipping Problem

    THE COMPLAINT IS FOR DEVICE S/N (B)(6) WHERE THE CORRECT, UNIQUE SPECTROMETER IDS AND THEIR ASSOCIATED COEFFICIENTS WERE NOT CORRECTLY CONFIGURED. THE INSTALLED SPECTROMETER DID NOT CORRECTLY ANALYZE THE WAVELENGTHS OF LIGHT WHICH RESULTED IN LOWER DEVICE RESULT SCORES, WHICH MAY IMPACT THE CLINICIAN'S CARE AND RESULT IN DELAY OF REFERRAL TO A DERMATOLOGIST. THERE IS NO ASSERTION OF DISCRETE PATIENT HARM. THE DEVICE IS AN ADJUNCTIVE DIAGNOSTIC DEVICE THAT IS INTENDED TO PROVIDE ADDITIONAL INFORMATION TO CLINICIANS, AIDING IN MANAGEMENT BUT NOT USED INDEPENDENTLY TO DETERMINE MANAGEMENT, DIAGNOSIS NOR THERAPY. DERMASENSOR DEVICE SHOULD BE USED IN CONJUNCTION WITH THE TOTALITY OF CLINICALLY RELEVANT INFORMATION FROM THE CLINICAL ASSESSMENT IN ACCORDANCE WITH THE DEVICE INSTRUCTIONS FOR USE, INCLUDING VISUAL ANALYSIS OF THE LESION, BY PHYSICIANS WHO ARE NOT DERMATOLOGISTS. THE DEVICE SHOULD BE USED ON LESIONS ALREADY ASSESSED AS SUSPICIOUS FOR SKIN CANCER AND NOT AS A SCREENING TOOL. THE DEVICE SHOULD NOT BE USED AS THE SOLE DIAGNOSTIC CRITERION NOR TO CONFIRM CLINICAL DIAGNOSIS OF SKIN CANCER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24196069 ↗
  • MalfunctionJan 27, 2026Nonstandard Device; Product Quality Problem; Manufacturing, Packaging or Shipping Problem

    THE COMPLAINT IS FOR DEVICE S/N (B)(6) WHERE THE CORRECT, UNIQUE SPECTROMETER IDS AND THEIR ASSOCIATED COEFFICIENTS WERE NOT CORRECTLY CONFIGURED. THE INSTALLED SPECTROMETER DID NOT CORRECTLY ANALYZE THE WAVELENGTHS OF LIGHT WHICH RESULTED IN LOWER DEVICE RESULT SCORES, WHICH MAY IMPACT THE CLINICIAN'S CARE AND RESULT IN DELAY OF REFERRAL TO A DERMATOLOGIST. THERE IS NO ASSERTION OF DISCRETE PATIENT HARM. THE DEVICE IS AN ADJUNCTIVE DIAGNOSTIC DEVICE THAT IS INTENDED TO PROVIDE ADDITIONAL INFORMATION TO CLINICIANS, AIDING IN MANAGEMENT BUT NOT USED INDEPENDENTLY TO DETERMINE MANAGEMENT, DIAGNOSIS NOR THERAPY. DERMASENSOR DEVICE SHOULD BE USED IN CONJUNCTION WITH THE TOTALITY OF CLINICALLY RELEVANT INFORMATION FROM THE CLINICAL ASSESSMENT IN ACCORDANCE WITH THE DEVICE INSTRUCTIONS FOR USE, INCLUDING VISUAL ANALYSIS OF THE LESION, BY PHYSICIANS WHO ARE NOT DERMATOLOGISTS. THE DEVICE SHOULD BE USED ON LESIONS ALREADY ASSESSED AS SUSPICIOUS FOR SKIN CANCER AND NOT AS A SCREENING TOOL. THE DEVICE SHOULD NOT BE USED AS THE SOLE DIAGNOSTIC CRITERION NOR TO CONFIRM CLINICAL DIAGNOSIS OF SKIN CANCER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24195965 ↗
  • MalfunctionJan 27, 2026Nonstandard Device; Product Quality Problem; Manufacturing, Packaging or Shipping Problem

    THE COMPLAINT IS FOR DEVICE S/N (B)(6) WHERE THE CORRECT, UNIQUE SPECTROMETER IDS AND THEIR ASSOCIATED COEFFICIENTS WERE NOT CORRECTLY CONFIGURED. THE INSTALLED SPECTROMETER DID NOT CORRECTLY ANALYZE THE WAVELENGTHS OF LIGHT WHICH RESULTED IN LOWER DEVICE RESULT SCORES, WHICH MAY IMPACT THE CLINICIAN'S CARE AND RESULT IN DELAY OF REFERRAL TO A DERMATOLOGIST. THERE IS NO ASSERTION OF DISCRETE PATIENT HARM. THE DEVICE IS AN ADJUNCTIVE DIAGNOSTIC DEVICE THAT IS INTENDED TO PROVIDE ADDITIONAL INFORMATION TO CLINICIANS, AIDING IN MANAGEMENT BUT NOT USED INDEPENDENTLY TO DETERMINE MANAGEMENT, DIAGNOSIS NOR THERAPY. DERMASENSOR DEVICE SHOULD BE USED IN CONJUNCTION WITH THE TOTALITY OF CLINICALLY RELEVANT INFORMATION FROM THE CLINICAL ASSESSMENT IN ACCORDANCE WITH THE DEVICE INSTRUCTIONS FOR USE, INCLUDING VISUAL ANALYSIS OF THE LESION, BY PHYSICIANS WHO ARE NOT DERMATOLOGISTS. THE DEVICE SHOULD BE USED ON LESIONS ALREADY ASSESSED AS SUSPICIOUS FOR SKIN CANCER AND NOT AS A SCREENING TOOL. THE DEVICE SHOULD NOT BE USED AS THE SOLE DIAGNOSTIC CRITERION NOR TO CONFIRM CLINICAL DIAGNOSIS OF SKIN CANCER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24195841 ↗
  • MalfunctionJan 27, 2026Nonstandard Device; Product Quality Problem; Manufacturing, Packaging or Shipping Problem

    THE COMPLAINT IS FOR DEVICE S/N (B)(6) WHERE THE CORRECT, UNIQUE SPECTROMETER IDS AND THEIR ASSOCIATED COEFFICIENTS WERE NOT CORRECTLY CONFIGURED. THE INSTALLED SPECTROMETER DID NOT CORRECTLY ANALYZE THE WAVELENGTHS OF LIGHT WHICH RESULTED IN LOWER DEVICE RESULT SCORES, WHICH MAY IMPACT THE CLINICIAN'S CARE AND RESULT IN DELAY OF REFERRAL TO A DERMATOLOGIST. THERE IS NO ASSERTION OF DISCRETE PATIENT HARM. THE DEVICE IS AN ADJUNCTIVE DIAGNOSTIC DEVICE THAT IS INTENDED TO PROVIDE ADDITIONAL INFORMATION TO CLINICIANS, AIDING IN MANAGEMENT BUT NOT USED INDEPENDENTLY TO DETERMINE MANAGEMENT, DIAGNOSIS NOR THERAPY. DERMASENSOR DEVICE SHOULD BE USED IN CONJUNCTION WITH THE TOTALITY OF CLINICALLY RELEVANT INFORMATION FROM THE CLINICAL ASSESSMENT IN ACCORDANCE WITH THE DEVICE INSTRUCTIONS FOR USE, INCLUDING VISUAL ANALYSIS OF THE LESION, BY PHYSICIANS WHO ARE NOT DERMATOLOGISTS. THE DEVICE SHOULD BE USED ON LESIONS ALREADY ASSESSED AS SUSPICIOUS FOR SKIN CANCER AND NOT AS A SCREENING TOOL. THE DEVICE SHOULD NOT BE USED AS THE SOLE DIAGNOSTIC CRITERION NOR TO CONFIRM CLINICAL DIAGNOSIS OF SKIN CANCER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24194358 ↗
  • MalfunctionDec 30, 2025Nonstandard Device; Product Quality Problem; Manufacturing, Packaging or Shipping Problem; Nonstandard Device; Product Quality Problem; Manufacturing, Packaging or Shipping Problem

    THE COMPLAINT IS FOR A DEVICE WHERE THE CORRECT, UNIQUE SPECTROMETER IDS AND THEIR ASSOCIATED COEFFICIENTS WERE NOT CORRECTLY CONFIGURED. THE INSTALLED SPECTROMETER DID NOT CORRECTLY ANALYZE THE WAVELENGTHS OF LIGHT WHICH RESULTED IN LOWER DEVICE RESULT SCORES, WHICH MAY IMPACT THE CLINICIAN'S CARE AND RESULT IN DELAY OF REFERRAL TO A DERMATOLOGIST. THERE IS NO ASSERTION OF DISCRETE PATIENT HARM. THE DEVICE IS AN ADJUNCTIVE DIAGNOSTIC DEVICE THAT IS INTENDED TO PROVIDE ADDITIONAL INFORMATION TO CLINICIANS, AIDING IN MANAGEMENT BUT NOT USED INDEPENDENTLY TO DETERMINE MANAGEMENT, DIAGNOSIS NOR THERAPY. DERMASENSOR DEVICE SHOULD BE USED IN CONJUNCTION WITH THE TOTALITY OF CLINICALLY RELEVANT INFORMATION FROM THE CLINICAL ASSESSMENT IN ACCORDANCE WITH THE DEVICE INSTRUCTIONS FOR USE, INCLUDING VISUAL ANALYSIS OF THE LESION, BY PHYSICIANS WHO ARE NOT DERMATOLOGISTS. THE DEVICE SHOULD BE USED ON LESIONS ALREADY ASSESSED AS SUSPICIOUS FOR SKIN CANCER AND NOT AS A SCREENING TOOL. THE DEVICE SHOULD NOT BE USED AS THE SOLE DIAGNOSTIC CRITERION NOR TO CONFIRM CLINICAL DIAGNOSIS OF SKIN CANCER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23929898 ↗
  • MalfunctionDec 29, 2025Nonstandard Device; Product Quality Problem; Manufacturing, Packaging or Shipping Problem; Nonstandard Device; Product Quality Problem; Manufacturing, Packaging or Shipping Problem

    THE COMPLAINT IS FOR A DEVICE WHERE THE CORRECT, UNIQUE SPECTROMETER IDS AND THEIR ASSOCIATED COEFFICIENTS WERE NOT CORRECTLY CONFIGURED. THE INSTALLED SPECTROMETER DID NOT CORRECTLY ANALYZE THE WAVELENGTHS OF LIGHT WHICH RESULTED IN LOWER DEVICE RESULT SCORES, WHICH MAY IMPACT THE CLINICIAN'S CARE AND RESULT IN DELAY OF REFERRAL TO A DERMATOLOGIST. THERE IS NO ASSERTION OF DISCRETE PATIENT HARM. THE DEVICE IS AN ADJUNCTIVE DIAGNOSTIC DEVICE THAT IS INTENDED TO PROVIDE ADDITIONAL INFORMATION TO CLINICIANS, AIDING IN MANAGEMENT BUT NOT USED INDEPENDENTLY TO DETERMINE MANAGEMENT, DIAGNOSIS NOR THERAPY. DERMASENSOR DEVICE SHOULD BE USED IN CONJUNCTION WITH THE TOTALITY OF CLINICALLY RELEVANT INFORMATION FROM THE CLINICAL ASSESSMENT IN ACCORDANCE WITH THE DEVICE INSTRUCTIONS FOR USE, INCLUDING VISUAL ANALYSIS OF THE LESION, BY PHYSICIANS WHO ARE NOT DERMATOLOGISTS. THE DEVICE SHOULD BE USED ON LESIONS ALREADY ASSESSED AS SUSPICIOUS FOR SKIN CANCER AND NOT AS A SCREENING TOOL. THE DEVICE SHOULD NOT BE USED AS THE SOLE DIAGNOSTIC CRITERION NOR TO CONFIRM CLINICAL DIAGNOSIS OF SKIN CANCER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23918516 ↗