WOLF Thrombectomy System, 7F FD0427-04Recalls & reported problems
Devoro Medical Inc. · WOLF Thrombectomy System, 7F - 25cm x 140cm · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00850029546238), received between Mar 8, 2023 and Jun 9, 2023 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Defective Device” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Defective Device1
- Detachment of Device or Device Component1
- Entrapment of Device1
- Material Deformation1
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Embolism/Embolus1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 9, 2023Material Deformation; Material Deformation
REPORTABLE UPON ANALYSIS COMPLETED ON 22MAY2023. IT WAS REPORTED THAT THE CATHETER HAD KINKED. A 7FR WOLF ARTERIAL THROMBECTOMY CATHETER WAS SELECTED FOR USE. LEFT FEMORAL ACCESS WAS OBTAINED AND A GUIDE SHEATH WAS INSERTED OVER THE BIFURCATION INTO THE RIGHT COMMON FEMORAL ARTERY. A 7FR WOLF ARTERIAL THROMBECTOMY CATHETER WAS ADVANCED. UPON INGESTING THE CLOT USING A PIN AND PULL METHOD, RESISTANCE WAS OBSERVED. AFTER THE SLEEVE HAD CAPTURED THE CLOT, THE INNER CATHETER WAS ATTEMPTED TO BE REMOVED. HOWEVER, DURING REMOVAL THE SHAFT SNAPPED IN HALF. THE ENTIRE DEVICE WAS SUCCESSFULLY REMOVED. A SECOND DEVICE, SAME MODEL, WAS SELECTED FOR USE. THE REMAINING CLOT WAS SUCCESSFULLY REMOVED USING THIS CATHETER. UPON REMOVAL, IT WAS OBSERVED THE CATHETER WAS KINKED. THE PROCEDURE WAS COMPLETED. THERE WERE NO PATIENT COMPLICATIONS. HOWEVER, DEVICE ANALYSIS REVEALED A FRACTURE AT THE KINKED AREA ON THE CATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17099415 ↗MalfunctionApr 19, 2023Mechanical Problem; Defective Device; Detachment of Device or Device Component; Mechanical Problem; Defective Device; Detachment of Device or Device Component
IT WAS REPORTED THAT THE SHAFT DETACHED. LEFT FEMORAL ACCESS WAS OBTAINED USING A NON-BSC 7FR SHEATH THAT WENT UP AND OVER THE BIFURCATION, AND WAS PARKED IN THE RIGHT COMMON FEMORAL ARTERY. AFTER CROSSING FROM ABOVE AND BELOW, DR. HAD A NON-BSC 300CM .014 GUIDEWIRE THAT EXTENDED FROM THE NON-BSC SHEATH, THROUGH THE SUPERFICIAL FEMORAL ARTERY (SFA) LESION, DOWN THE ANTERIOR TIBIAL ARTERY AND EXITING A SMALL SHEATH IN THE PATIENTS RIGHT FOOT, WHERE A HEMOSTAT WAS CLAMPED ON THE WIRE TO ENSURE WIRE ACCESS WOULD NOT BE LOST. A 7FR WOLF ARTERIAL THROMBECTOMY CATHETER WAS ADVANCED TO THE TARGET LOCATION. WHILE PINNING AND PULLING THE DEVICE TO CAPTURE THE CLOT, RESISTANCE WAS NOTED. WHILE FINISHING THE SLEEVE PULL, AND REMOVAL OF THE INNER CATHETER, THE SHAFT SNAPPED IN HALF. THE CATHETER WAS THEN REMOVED FROM THE PATIENT. A NEW DEVICE, SAME MODEL, WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16772559 ↗InjuryMar 8, 2023Entrapment of Device; Entrapment of Device
IT WAS REPORTED THAT THERE WAS LOSS OF BLOOD FLOW TO THE FOOT. THE WOLF THROMBECTOMY SYSTEM WAS SELECTED TREATMENT OF THROMBOSIS OF LEFT LOWER EXTREMITY FEMORAL PERONEAL SYNTHETIC BYPASS GRAFT (PLACED IN 2019). THE WOLF WAS CHOSEN AFTER THE PATIENT PULLED OUT THE EKOS CATHETER PREMATURELY OVERNIGHT, FOLLOWED BY ASPIRATION THROMBECTOMY WITH A NON-BSC DEVICE AND SUBSEQUENT PLACEMENT OF A STANDARD THROMBOLYSIS CATHETER. THE PATIENT WAS BROUGHT BACK TO THE LAB AND ONCE THROMBOLYSIS CATHETER WAS REMOVED, PHYSICIAN CONFIRMED PATIENT HAD BLOOD FLOW IN THE PERONEAL AND INTO THE FOOT, HOWEVER, HAD A 25CM OCCLUSION IN THE PROXIMAL PORTION OF THE BYPASS GRAFT. DURING WOLF PLACEMENT, THE PHYSICIAN HAD COMPLETED THE FIRST SLEEVE PULL AND CONFIRMED WIRE MOVEMENT THROUGH THE INNER SHAFT. WHILE REMOVING, THE CATHETER AND INNER SHAFT/SLEEVE BECAME TIGHT AND UNABLE TO REMOVE WITHOUT LOSING WIRE. THE REMOVAL WAS STOPPED, AND THE CATHETER WAS PUSHED FURTHER DISTAL AND THEN REMOVING THE INNER SHAFT WITH THE NON-BSC WIRE TO RELOAD AND TRY AGAIN. ONLY MINIMAL CLOT WAS ABLE TO BE REMOVED, SO PHYSICIAN OPTED TO RUN TWO OTHER NON-BSC DEVICES WITH MINIMAL CLOT REMOVED. AS A RESULT, IT APPEARED THAT CLOT WAS
Patient problems: Embolism/Embolus
MAUDE report 16503919 ↗
Other Peripheral Mechanical Thrombectomy With Aspiration devices
- Indigo System LITNG12HTORQ115-APenumbra, Inc. · Peripheral Mechanical Thrombectomy With Aspiration4 deaths105 reports
- Indigo System LITNG7XTORQ130-APenumbra, Inc. · Peripheral mechanical thrombectomy with aspiration100 reports
- Indigo System LITNG12XTORQ100-APenumbra, Inc. · Peripheral Mechanical Thrombectomy With Aspiration1 death80 reports
- ClotTriever XL Catheter AS-0235Inari Medical, Inc. · Peripheral Mechanical Thrombectomy With Aspiration1 recall9 deaths20 reports
- Indigo System LITNGBT7TQ130-APenumbra, Inc. · Peripheral Mechanical Thrombectomy With Aspiration46 reports
- Indigo System LITNG12HTORQ100-APenumbra, Inc. · Peripheral Mechanical Thrombectomy With Aspiration1 death46 reports
- Triever 24 22-101Inari Medical, Inc. · Peripheral Mechanical Thrombectomy With Aspiration20 deaths34 reports
- ClotTriever Catheter, 16 mm 40-102Inari Medical, Inc. · Peripheral Mechanical Thrombectomy With Aspiration11 deaths27 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page