Skip to content

Prodigy™ Catheter PRODIGY8FRecalls & reported problems

Imperative Care, Inc. · 8F Prodigy™ Catheter · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00850032079143), received between Dec 19, 2025 and Dec 24, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Detachment of Device or Device Component1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Vascular Dissection1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE SUPERFICIAL FEMORAL ARTERY (SFA) WAS TREATED USING A THIRD-PARTY SHEATH, 8F PRODIGY CATHETER, AND 8F PRODIGY TWIST. THE CLOT WAS ORGANIZED. A VESSEL DISSECTION WAS DIAGNOSED, WHICH THE PHYSICIAN WAS UNAWARE OF UNTIL THE INITIAL ANGIOGRAM. THE DISSECTION IMPEDED BLOOD FLOW TO THE FOOT. IT WAS DETERMINED THAT THE DISSECTION WAS THE PRIMARY REASON FOR THE LACK OF FLOW AND WAS UNRELATED TO THE CATHETER'S ABILITY TO REMOVE CLOT. IT IS UNKNOWN WHETHER ANY ADDITIONAL INTERVENTION WAS PERFORMED TO ADDRESS THE DISSECTION. THERE WAS NO MALFUNCTION REPORTED DURING THE PROCEDURE.

    Patient problems: Vascular Dissection

    MAUDE report 23889460 ↗
  • InjuryDec 19, 2025Break; Material Separation; Detachment of Device or Device Component; Break; Material Separation; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN IN-STENT RESTENOSIS (ISR) PROCEDURE, THE PRODIGY 8F CATHETER WAS USED WITH A THIRD-PARTY SHEATH AND GUIDEWIRE. AFTER THE SECOND PASS, THE PHYSICIAN NOTICED THE DISTAL TIP WAS MISSING. UPON FLUOROSCOPY, THE MARKER BAND WAS VISIBLE INSIDE THE STENT IN THE SFA. THE PHYSICIAN SUCCESSFULLY RETRIEVED THE MARKER BAND WITH THE USE OF AN ADDITIONAL THIRD-PARTY DEVICE. THERE WERE NO DEVICE MALFUNCTIONS OR DIFFICULTY TRACKING THE DEVICE DURING THE PROCEDURE. THE PATIENT WAS STABLE POST PROCEDURE, AND NO PATIENT SEQUELAE WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23854634 ↗

FDA clearance and classification

Cleared through 510(k) K214114 on Apr 6, 2022 (decision: Substantially Equivalent). Submitted by Truvic Medical, Inc. as “Prodigy Thrombectomy System”. The FDA review took 97 days. FDA record ↗

The FDA regulates product code QEW (peripheral mechanical thrombectomy with aspiration) under 21 CFR 870.5150 as a Class II device. 21 CFR 870.5150 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Prodigy™ Catheter PRODIGY8F

Has the Prodigy™ Catheter PRODIGY8F been recalled?

No FDA recall record in our data names this exact device identifier (DI 00850032079143), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Prodigy™ Catheter PRODIGY8F?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Dec 19, 2025 and Dec 24, 2025. In those reports, the most common specific device problems are “Break” (1), “Detachment of Device or Device Component” (1) and “Material Separation” (1); the most common patient problem is “Vascular Dissection” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Prodigy™ Catheter PRODIGY8F?

None of the reports are classified as deaths, 2 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Prodigy™ Catheter PRODIGY8F FDA approved?

It was cleared through 510(k) K214114 on Apr 6, 2022. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Prodigy™ Catheter PRODIGY8F, and is it still sold?

The FDA's device database lists Imperative Care, Inc. as the company responsible for it (brand Prodigy™ Catheter, model PRODIGY8F). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Peripheral Mechanical Thrombectomy With Aspiration devices

All Peripheral Mechanical Thrombectomy With Aspiration devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page