The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00850039803024), received between Oct 25, 2023 and Oct 25, 2023 — classified as 1 injury.
The most frequently reported specific device problem is “Migration” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Migration1
Patient problems
- Dysphagia/ Odynophagia1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryOct 25, 2023Migration; Migration
THE PATIENT HAD A C3-C4 ANTERIOR FUSION PROCEDURE THAT INCLUDED THE USE OF AN 18MM ANGLED CONTINUUM IMPLANT WITH AN 8MM INTERBODY CAGE FILLED WITH AUTOGRAFT. TWO WEEKS POST OP IT WAS IDENTIFIED THAT THE INTERBODY CAGE AND CONTINUUM IMPLANT HAD MIGRATED FORWARD RESULTING IN THE NEED FOR A REOPERATION. ON THE X-RAY IMAGING, THE SURGEON OBSERVED THAT THE CONTINUUM IMPLANT WAS NOT FLUSH AGAINST THE INTERBODY CAGE WHICH COULD HAVE ALLOWED THE MIGRATION OF THE INTERBODY CAGE. THE PATIENT HAD A TERRIBLE COUGH DUE TO COVID POST OPERATIVELY WHICH COULD HAVE CONTRIBUTED TO THE MIGRATION OF THE INTERBODY CAGE.
Patient problems: Dysphagia/ Odynophagia
MAUDE report 18007896 ↗
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page