Panosol 6-3DRecalls & reported problems
Daavlin Distributing CO., The · UVB-680 · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00850054882141), received between Dec 13, 2024 and Dec 13, 2024 — classified as 1 injury.
The most frequently reported specific device problem is “Use of Device Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Use of Device Problem1
- Use of Incorrect Control/Treatment Settings1
Patient problems
- Burn(s)1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryDec 13, 2024Use of Incorrect Control/Treatment Settings; Use of Device Problem; Use of Incorrect Control/Treatment Settings; Use of Device Problem
PATIENT HAS BEEN DIAGNOSED WITH PSORIASIS VULGARIS (L40.0). HER DOCTOR PRESCRIBED THE USE OF A PANOSL 63D (ULTRAVIOLET LIGHT THERAPY SYSTEM) TO TREAT HER CONDITION AT HOME BECAUSE OF FREQUENT FLAREUPS OF THE DISEASE. PATIENT WAS ADVISED THAT SHE WOULD NEED 2-3 TREATMENTS PER WEEK AND THAT SHE WOULD BE CONTACTED BY A DOCTOR TO PROVIDE MORE DETAIL REGARDING HER TREATMENT PROTOCOL. UPON RECEIPT OF HER EQUIPMENT, THE PATIENT STILL HAD NOT RECEIVED THE CALL FROM THE PHYSICIAN AND DECIDED TO BEGIN TREATMENT ON HER OWN. USING HER PREVIOUS TANNING EXPERIENCE AS A BASELINE, THE PATIENT DECIDED TO USE AN INITIAL TREATMENT TIME OF SEVEN MINUTES. SHE DID NOT REFER TO THE RECOMMENDED PROTOCOLS INCLUDED IN THE DEVICE OPERATIONS MANUAL NOR CONSULT HER PHYSICIAN OR NATBIO PRIOR TO THE TREATMENT. THIS ARBITRARY TREATMENT TIME DOES NOT CORRESPOND TO THE RECOMMENDED PROTOCOL FOR HER SKIN TYPE (FITZPATRICK SKIN TYPE II) NOR ANY OTHER PHOTOTHERAPY PROTOCOL. AS A RESULT, THE PATIENT RECEIVED FIRST DEGREE BURNS ACROSS HER BODY. SHE SOUGHT TREATMENT FOR HER BURNS AT A LOCAL EMERGENCY ROOM AND WAS PRESCRIBED IBUPROFEN AND PYREGESIC TO TREAT THE INJURIES. HER CURRENT CONDITION IS MUCH IMPROVED WITH SKIN RED
Patient problems: Burn(s)
MAUDE report 20925440 ↗
FDA clearance and classification
Cleared through Special 510(k) K233811 on Dec 13, 2023 (decision: Substantially Equivalent). Submitted by Daavlin Distributing Co. as “DT Controlled Phototherapy Equipment”. The FDA review took 14 days. FDA record ↗
The FDA regulates product code FTC (light, ultraviolet, dermatological) under 21 CFR 878.4630 as a Class II device. 21 CFR 878.4630 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Panosol 6-3D
Has the Panosol 6-3D been recalled?
No FDA recall record in our data names this exact device identifier (DI 00850054882141), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Panosol 6-3D?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Dec 13, 2024. That report lists device problems “Use of Device Problem” (1) and “Use of Incorrect Control/Treatment Settings” (1); patient problem “Burn(s)” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Panosol 6-3D?
None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Panosol 6-3D FDA approved?
It was cleared through Special 510(k) K233811 on Dec 13, 2023. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Panosol 6-3D, and is it still sold?
The FDA's device database lists Daavlin Distributing CO., The as the company responsible for it (brand Panosol 6-3D, model Panosol 6-3D). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other LIGHT, ULTRAVIOLET, DERMATOLOGICAL devices
- Zerigo Handheld System HD-02Zerigo Health, Inc. · Light, Ultraviolet, Dermatological2 reports
- National Biological Corporation UVB-631-9National Biological Corporation · Light, Ultraviolet, Dermatological2 reports
- 308nm Excimer System XECL-308C, XECL-308DChongqing Peninsula Medical Technology Co., Ltd. · Light, Ultraviolet, Dermatological1 report
- DAAVLIN 7 SERIESDaavlin Distributing CO., The · Light, Ultraviolet, Dermatological1 report
- DAAVLIN 7 SERIESDaavlin Distributing CO., The · Light, Ultraviolet, Dermatological1 report
- National Biological Corporation UHU-318National Biological Corporation · Light, Ultraviolet, Dermatological1 report
- DAAVLIN Phothera 600-3DDaavlin Distributing CO., The · LIGHT, ULTRAVIOLET, DERMATOLOGICAL1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page