Skip to content

FlowTriever Catheter, Large 10-103Recalls & reported problems

Inari Medical, Inc. · FlowTriever Catheter, 18 mm · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00850291007024), received between Jun 26, 2023 and Jun 26, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 26, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    A 51-YEAR-OLD FEMALE PATIENT UNDERWENT BOWEL SURGERY 2 MONTHS AGO AND AFTER BEING DISCHARGED, THE PATIENT WAS VERY INACTIVE AND DEVELOPED A BILATERAL PULMONARY EMBOLISM (PE). A COMPUTED TOMOGRAPHY (CT) SCAN REVEALED THE MAIN CLOT BURDEN WAS IN THE RIGHT PULMONARY ARTERY (INTERLOBAR AND TRUNCUS ANTERIOR) WITH A LITTLE ON THE LEFT SIDE. THE PATIENT UNDERWENT A THROMBECTOMY PROCEDURE WITH THE INARI FLOWTRIEVER SYSTEM ON (B)(6) 2023. NINE ASPIRATIONS WERE PERFORMED WITH THE INARI TRIEVER24 (T24) WHICH CLEARED THE MAJORITY OF CLOT ON THE RIGHT. THE PHYSICIAN THEN REPOSITIONED THE T24 TO THE MAIN PULMONARY ARTERY AND PERFORMED 3 SUBSEQUENT ASPIRATIONS. AN ANGIOGRAM WAS THEN PERFORMED WHICH SHOWED SOME REMAINING CLOT AND THE PHYSICIAN DECIDED TO USE THE INARI FLOWTRIEVER (FT) L DISKS TO ADDRESS THE REMAINING CLOT. THE DISKS WERE ADVANCED WITHOUT ISSUE AND ONCE THE PHYSICIAN LOCATED THE RADIOPAQUE MARKER BEHIND THE CLOT, THE PHYSICIAN BEGAN RETRACTING THE DELIVERY SYSTEM. HOWEVER, THE DISKS DID NOT OPEN AS EXPECTED AND REMAINED RELATIVELY FLAT. A SECOND ATTEMPT WAS MADE TO INSERT THE DISKS INTO THE DELIVERY SYSTEM, BUT THE RADIOPAQUE MARKER INADVERTENTLY DETACHED FROM THE FT L DISK AND WEN

    Patient problems: Foreign Body In Patient

    MAUDE report 17206050 ↗

Other Peripheral Mechanical Thrombectomy With Aspiration devices

All Peripheral Mechanical Thrombectomy With Aspiration devices →