Skip to content

FlowTriever Catheter, XL 10-104Recalls & reported problems

Inari Medical, Inc. · FlowTriever Catheter, 25 mm · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
3
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00850291007055), received between May 21, 2020 and Nov 7, 2024 — classified as 3 deaths, 1 injury.

The most frequently reported specific device problem is “Use of Device Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Use of Device Problem2
  • Detachment of Device or Device Component1
  • Difficult to Remove1

Patient problems

  • Pulmonary Embolism3
  • Cardiac Arrest1
  • Cardiopulmonary Arrest1
  • Discomfort1
  • Foreign Body In Patient1
  • Respiratory Arrest1
  • Sudden Cardiac Death1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • DeathNov 7, 2024Difficult to Remove; Use of Device Problem; Adverse Event Without Identified Device or Use Problem; Difficult to Remove; Use of Device Problem

    A 73-YEAR-OLD FEMALE ARRIVED AT THE HOSPITAL 2 WEEKS PRIOR DUE TO DEEP VEIN THROMBOSIS (DVT). THE PATIENT HAS A HISTORY OF DVT, ATRIAL FIBRILLATION, AND HAD A RECENT HYSTERECTOMY. THE HYSTERECTOMY LED TO BLEEDING COMPLICATIONS WHICH RESULTED IN THE PATIENT DISCONTINUING THEIR BLOOD THINNERS. AFTER THE PATIENT WAS ADMITTED TO THE HOSPITAL, THEY WERE PLACED ON HEPARIN, THIS LED TO HER BLOOD COUNT DROPPING. THE HEPARIN WAS DISCONTINUED, AND AN INFERIOR VENA CAVA (IVC) FILTER WAS PLACED. THE PATIENT RETURNED TO THE HOSPITAL DUE TO CONTINUED SYMPTOMS. A COMPUTED TOMOGRAPHY (CT) SCAN REVEALED MINIMAL PULMONARY EMBOLISM (PE) AND EXTENSIVE CLOT FROM THE LEFT ILIAC INTO THE IVC WHERE THE FILTER WAS PRESENT. THE DECISION WAS MADE TO PERFORM A DVT THROMBECTOMY AFTER REMOVING THE IVC FILTER. THE PATIENT WAS FLIPPED SUPINE, AND ACCESS WAS ACHIEVED. THE IVC FILTER WAS REMOVED THROUGH THE INTERNAL JUGULAR AND THE FLOWTRIEVER CATHETER, XL (FTXL) WITH 2 DISCS DEPLOYED. THE CLOTTRIEVER XL (CTXL) CATHETER WAS USED TO CLEAR THE IVC BUT, THE DISTAL RADIOPAQUE TIP FROM THE FTXL GOT CAUGHT ON THE CTXL. TO ADDRESS THIS, THE FTXL DISCS WERE BRIEFLY DROPPED TO PULL THE CTXL DOWN AND THE PHYSICIAN REDEPLOYED

    Patient problems: Cardiac Arrest; Pulmonary Embolism

    MAUDE report 20635922 ↗
  • DeathMar 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A MALE PATIENT PRESENTED TO THE HOSPITAL WITH BORDERLINE PHLEGMASIA CERULEA DOLENS IN THE LOWER EXTREMITIES. IMAGING REVEALED THAT THE PATIENT HAD BILATERAL ILIOFEMORAL DEEP VEIN THROMBOSIS (DVT), INFERIOR VENA CAVA (IVC) THROMBUS SUPRARENAL, AND LEFT RENAL VEIN THROMBOSIS. MOST OF THE CLOT WAS ACUTE, WITH SOME SUBACUTE/CHRONIC CLOT THAT WAS WALL ADHERENT IN THE IVC AND LEFT RENAL VEIN. A CHEST COMPUTED TOMOGRAPHY (CT) SCAN ALSO REVEALED A DISTAL LEFT-SIDED CLOT IN THE PULMONARY VEIN, BUT THERE WAS NO RIGHT HEART STRAIN AND NO ELEVATED TROPONIN LEVEL. THE PATIENT WAS NEWLY DIAGNOSED WITH KIDNEY CANCER AND HAD A COVID VIRAL INFECTION 10 DAYS PRIOR. THE CLOT AGE WAS ESTIMATED AT 3 DAYS. THE PATIENT WAS SCHEDULED FOR A THROMBECTOMY ON (B)(6) 2024 USING THE INARI FLOWTRIEVER RETRIEVAL/ASPIRATION SYSTEM. THE PROCEDURE WAS CONDUCTED WITH THE PATIENT IN THE SUPINE POSITION. ACCESS WAS OBTAINED VIA THE BILATERAL FEMORAL VEIN; THE CLOTTRIEVER XL CATHETER (CTXL) WAS USED WITH THE CLOTTRIEVER (CT) 16 FR SHEATH AND FLOWTRIEVER XL (FTXL) DISCS CONTRALATERALLY. THE TRIEVER16 (T16) AND TRIEVER16 CURVE (T16 CURVE) WERE ALSO USED TO ASPIRATE ACUTE CLOT IN THE IVC AND LEFT RENAL VEIN WITHOUT INCIDEN

    Patient problems: Discomfort; Pulmonary Embolism; Respiratory Arrest

    MAUDE report 18834141 ↗
  • InjuryJun 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    ADDITIONAL INFORMATION WAS PROVIDED BY THE COMPANY REPRESENTATIVE WHO WAS PRESENT DURING THE SURGERY AND IT WAS LEARNED THAT EXCESSIVE FORCE WAS BRIEFLY APPLIED WHEN ADVANCING THE FT DISKS. THE SURGEON STOPPED ADVANCING WITHIN SECONDS OF FEELING RESISTANCE, BUT BY THAT TIME THE DISTAL TIP HAD ALREADY DISLODGED. ADDITIONAL INFORMATION HAS BEEN REQUESTED ON THE PATIENT'S STATUS, BUT NO NEW INFORMATION HAS BEEN PROVIDED.

    Patient problems: Foreign Body In Patient

    MAUDE report 17218158 ↗
  • DeathMay 21, 2020Use of Device Problem; Use of Device Problem

    A (B)(6)-YEAR-OLD MALE PATIENT WITH AN INFERIOR VENA CAVA (IVC) FILTER AND A HISTORY OF DEEP VEIN THROMBOSIS (DVT), PULMONARY EMBOLISM (PE), AND GASTROINTESTINAL (GI) BLEEDING PRESENTED TO THE HOSPITAL WITH LEG SWELLING. THE PATIENT WAS EXAMINED, PLACED ON ANTICOAGULATION MEDICATION, AND DISCHARGED. THREE DAYS LATER HE RETURNED TO THE HOSPITAL WITH SEVERE LEG PAIN, DISCOLORATION, AND EXTREME SWELLING. WHILE IN THE HOSPITAL, HE TESTED POSITIVE FOR COVID-19. THE PATIENT WAS INELIGIBLE FOR THROMBOLYTIC MEDICAL THERAPY DUE TO THE GI BLEED HISTORY, SO THE PLAN WAS TO TREAT THE BILATERAL DVT SURGICALLY. ON (B)(6) 2020, THE PATIENT UNDERWENT THROMBECTOMY USING THE INARI MEDICAL CLOTTRIEVER SHEATH AND FLOWTRIEVER CATHETER. GENERAL ANESTHESIA WAS ADMINISTERED BECAUSE OF THE COVID-POSITIVE STATUS. WITH THE PATIENT SUPINE, ULTRASOUND WAS USED TO HELP GAIN JUGULAR ACCESS WITH A 5 FR MICROPUNCTURE, WHICH WAS THEN DILATED AND EXCHANGED FOR A 13 FR SHEATH. ULTRASOUND WAS ALSO USED TO HELP GAIN BILATERAL LOWER FEMORAL VEIN ACCESS WITH 5 FR MICROPUNCTURES THAT WERE DILATED AND EXCHANGED FOR 8 FR SHEATHS. VENOGRAMS PERFORMED THROUGH THE 8 FR SHEATHS CONFIRMED SIGNIFICANT CLOT FROM THE IVC FILTER DOW

    Patient problems: Cardiopulmonary Arrest; Pulmonary Embolism; Sudden Cardiac Death

    MAUDE report 10080940 ↗

Other Peripheral Mechanical Thrombectomy With Aspiration devices

All Peripheral Mechanical Thrombectomy With Aspiration devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page