Protrieve Sheath 60-101Recalls & reported problems
Inari Medical, Inc. · Protrieve™ Sheath (Sheath) · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 1
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00850291007222), received between Sep 15, 2022 and Aug 13, 2025 — classified as 1 death, 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Material Separation” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Material Separation1
- Use of Device Problem1
Patient problems
- Air Embolism1
- Cardiac Arrest1
- Cardiovascular Dysfunction/Insufficiency1
- Hypoxia1
- No Clinical Signs, Symptoms or Conditions1
- Pulmonary Embolism1
- Tachycardia1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 13, 2025Material Separation; Material Separation
ON JULY 15, 2025, INARI MEDICAL RECEIVED A COPY OF A VOLUNTARY MEDWATCH MW5172147 THAT STATED A PROTRIEVE SHEATH TORE APART ON INSERTION INTO THE INSERTION PORT AND IT DID NOT ENTER THE PATIENT. THE DEVICE WAS NO LONGER USED DURING THE PROCEDURE. PER MEDWATCH EVENT/PROBLEM SECTION "INARI SHEATH, ON INSERTION, TORE APART, MD DISCONTINUED USE OF THE PRODUCT. THE DEVICE DID NOT ENTER THE PT JUST THE INSERTION PORT ."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22799034 ↗InjuryMar 7, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A FEMALE PATIENT WHO RECENTLY UNDERWENT A THROMBECTOMY ON (B)(6) 2024 TO ADDRESS HER PULMONARY EMBOLISM (PE) WAS SCHEDULED FOR A THROMBECTOMY ON (B)(6) 2024 WITH THE INARI CLOTTRIEVER THROMBECTOMY SYSTEM TO ADDRESS A LEFT LOWER EXTREMITY DEEP VEIN THROMBOSIS (DVT) (WHICH WAS NOT ADDRESSED DURING HER PREVIOUS PE THROMBECTOMY). THE PRIOR PE THROMBECTOMY (PERFORMED USING INARI DEVICES) WAS UNEVENTFUL AND THE CASE PLAN WAS TO ADDRESS THE PE FIRST AND SCHEDULE THE DVT THROMBECTOMY AT A LATER DATE. THE CLOT AGE WAS ESTIMATED AT 30 DAYS. THE PATIENT WAS PLACED IN A SUPINE POSITION AND THE INTRAJUGULAR VEIN WAS ACCESSED WITH AN INARI PROTRIEVE. THE INARI CLOTTRIEVER XL CATHETER (CTXL) WAS THEN INTRODUCED, AND 3 PASSES WERE PERFORMED WHICH REMOVED A SUBSTANTIAL AMOUNT OF CLOT. THE PROTRIEVE WAS SECURED PRIOR TO THE PASSES BUT DID RETRACT APPROXIMATELY 2 INCHES DURING THE PROCEDURE. THE PROTRIEVE WAS RE-POSITIONED AND ONE MORE PASS WAS PERFORMED WITH THE CTXL. A VENOGRAM WAS PERFORMED WHICH REVEALED THAT FULL PATENCY WAS RESTORED. THE PROCEDURE WAS CONCLUDED, AND THE ACCESS SITE WAS CLOSED. UNFORTUNATELY, THE PATIENT BEGAN TO DECOMPENSATE AND IMAGING REVEALED CLOT IN THE RIGHT PULMONARY ARTE
Patient problems: Cardiovascular Dysfunction/Insufficiency; Hypoxia; Pulmonary Embolism; Tachycardia
MAUDE report 18858388 ↗DeathSep 15, 2022Use of Device Problem; Use of Device Problem
A FEMALE PATIENT IN HER EARLY 60'S PRESENTED TO THE HOSPITAL WITH LEG EDEMA AND PAIN. THE PATIENT WAS DIAGNOSED WITH BILATERAL DEEP VEIN THROMBOSIS (DVT) AND SCHEDULED FOR A THROMBECTOMY WITH THE INARI CLOTTRIEVER THROMBECTOMY SYSTEM ON (B)(6) 2022. THE PATIENT'S MEDICAL HISTORY INCLUDED BREAST CANCER IN 2011; TREATMENT CONSISTED OF A MASTECTOMY AND CHEMOTHERAPY. THE PATIENT WAS POSITIONED SUPINE IN BILATERAL FROG-LEG POSITION AND ACCESS WAS GAINED THROUGH THE INTERNAL JUGULAR (IJ) VEIN. ACCESS WAS THEN GAINED TO THE POPLITEAL VEIN AND THE PHYSICIAN DILATED UP TO THE 13 FR CLOTTRIEVER SHEATH. THE INARI PROTRIEVE SHEATH WAS THEN INSERTED INTO THE IJ OVER A 145 CM AMPLATZ SUPERSTIFF GUIDEWIRE WITH J-TIP. THE PATIENT HAD A SMALL THROMBUS AT THE TOP OF THE INFERIOR VENA CAVA (IVC) AT THE ATRIAL JUNCTION. THE PHYSICIAN WANTED TO ATTEMPT A VACUUM ASPIRATION IN THE RIGHT ATRIUM (RA) THROUGH THE PROTRIEVE SHEATH PRIOR TO DEPLOYING THE FUNNEL. WHEN THE PHYSICIAN REMOVED THE DILATOR FROM THE PROTRIEVE SHEATH, THE AMPLATZ SS GUIDEWIRE WAS INADVERTENTLY REMOVED WITH ALONG WITH IT. THE PHYSICIAN RECOGNIZED IMMEDIATELY THAT THE GUIDEWIRE WAS REMOVED AND WANTED TO RELEASE SUCTION TO PREVENT ASPIR
Patient problems: Air Embolism; Cardiac Arrest
MAUDE report 15433183 ↗
FDA clearance and classification
The FDA regulates product code QEW (peripheral mechanical thrombectomy with aspiration) under 21 CFR 870.5150 as a Class II device. 21 CFR 870.5150 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Protrieve Sheath 60-101
Has the Protrieve Sheath 60-101 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00850291007222), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Protrieve Sheath 60-101?
The FDA's MAUDE database lists 3 adverse event reports for this device, received between Sep 15, 2022 and Aug 13, 2025. In those reports, the most common specific device problems are “Material Separation” (1) and “Use of Device Problem” (1); the most common patient problems are “Air Embolism” (1), “Cardiac Arrest” (1) and “Cardiovascular Dysfunction/Insufficiency” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Protrieve Sheath 60-101?
1 report is classified as a death, 1 as an injury and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Protrieve Sheath 60-101, and is it still sold?
The FDA's device database lists Inari Medical, Inc. as the company responsible for it (brand Protrieve Sheath, model 60-101). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Peripheral Mechanical Thrombectomy With Aspiration devices
- Indigo System LITNG12HTORQ115-APenumbra, Inc. · Peripheral Mechanical Thrombectomy With Aspiration4 deaths105 reports
- Indigo System LITNG7XTORQ130-APenumbra, Inc. · Peripheral mechanical thrombectomy with aspiration100 reports
- Indigo System LITNG12XTORQ100-APenumbra, Inc. · Peripheral Mechanical Thrombectomy With Aspiration1 death80 reports
- ClotTriever XL Catheter AS-0235Inari Medical, Inc. · Peripheral Mechanical Thrombectomy With Aspiration1 recall9 deaths21 reports
- Indigo System LITNGBT7TQ130-APenumbra, Inc. · Peripheral Mechanical Thrombectomy With Aspiration46 reports
- Indigo System LITNG12HTORQ100-APenumbra, Inc. · Peripheral Mechanical Thrombectomy With Aspiration1 death46 reports
- Triever 24 22-101Inari Medical, Inc. · Peripheral Mechanical Thrombectomy With Aspiration20 deaths34 reports
- ClotTriever Catheter, 16 mm 40-102Inari Medical, Inc. · Peripheral Mechanical Thrombectomy With Aspiration11 deaths27 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page