Skip to content

Triever16 Curve 25-201Recalls & reported problems

Inari Medical, Inc. · Triever16 Curve (FlowTriever Retrieval/Aspiration System) · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00850291007437), received between Mar 5, 2024 and Mar 24, 2026 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Material Separation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation1
  • Use of Device Problem1

Patient problems

  • Cardiac Perforation1
  • Cardiac Tamponade1
  • Cardiogenic Shock1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 24, 2026Material Separation

    ON (B)(6) 2026, A PATIENT UNDERWENT PULMONARY EMBOLISM (PE) THROMBECTOMY USING STRYKER PERIPHERAL VASCULAR DEVICES. WHILE REMOVING THE TRIEVER16, A SECTION OF THE CATHETER THAT WAS ALREADY OUTSIDE OF THE PATIENT'S VASCULATURE BROKE IN HALF. THE DISTAL END WAS ABLE TO BE REMOVED FROM THE PATIENT WITHOUT ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24678050 ↗
  • InjuryMar 5, 2024Use of Device Problem; Use of Device Problem

    A FEMALE PATIENT WITH A BILATERAL PULMONARY EMBOLISM (PE) PRESENTED TO THE EMERGENCY DEPARTMENT AND WAS SCHEDULED FOR A THROMBECTOMY WITH THE INARI FLOWTRIEVER SYSTEM ON (B)(6) 2024. THE CLOT AGE WAS ESTIMATED AT 8 DAYS. THE INITIAL THROMBECTOMY WAS SUCCESSFUL IN REMOVING APPROXIMATELY 80% OF THE CLOT BURDEN, BUT THE PHYSICIAN WANTED TO RETRIEVE A SMALL CLOT REMAINING IN THE TRUNCUS ANTERIOR. THE PHYSICIAN ATTEMPTED TO PASS THE TRIEVER16 CURVE OVER A WHOLEY GUIDEWIRE TO THE RIGHT TRUNCUS ANTERIOR. HOWEVER, THE RIGHT VENTRICLE PERFORATED AND THE PATIENT ARRESTED; A CODE WAS INITIATED. THE PATIENT WAS UNRESPONSIVE FOR 30 MINUTES, HOWEVER, INTERVENTION MEASURES (EXTRACORPOREAL MEMBRANE OXYGENATION (ECMO) AND INTUBATION) WERE ULTIMATELY SUCCESSFUL IN ACHIEVING RETURN OF SPONTANEOUS CIRCULATION (ROSC) AND THE PATIENT'S VITALS WERE STABLE. PERICARDIOCENTESIS WAS PERFORMED WITH A LARGE VOLUME (300 CC) OF BLOOD REMOVED AND RETURNED VIA THE INARI FLOWSAVER AND THE INARI INTRI24 SHEATH. THE PATIENT WAS TRANSFERRED TO THE ICU IN STABLE CONDITION AND WILL BE EVALUATED FOR ANY FURTHER INTERVENTION.

    Patient problems: Cardiac Perforation; Cardiac Tamponade; Cardiogenic Shock

    MAUDE report 18834250 ↗

Other Peripheral Mechanical Thrombectomy With Aspiration devices

All Peripheral Mechanical Thrombectomy With Aspiration devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page