Axogen Corporation · Axoguard HA+ Nerve Protector is provided sterile, for single use only, and in a variety of sizes to meet the surgeon’s needs. K223640: Axoguard HA+ Nerve Protector is a surgical implant that provides non-constricting protection for non-transected peripheral nerve injuries. Axoguard HA+ Nerve Protector is designed to be an interface between the nerve and the surrounding tissue. Axoguard HA+ Nerve Protector is comprised of an extracellular matrix (ECM) and is fully remodeled during the healing process. When hydrated, Axoguard HA+ Nerve Protector is easy to handle, soft, pliable, nonfriable, and porous. The lubricant coating on Axoguard HA+ Nerve Protector is composed of sodium hyaluronate and sodium alginate. When hydrated, the lubricant coating reduces friction between the nerve and the surrounding tissue. Axoguard HA+ Nerve Protector is flexible to accommodate movement of the joint and associated tendons and has sufficient mechanical strength to hold sutures. K231708: Axoguard HA+ Nerve Protector is a surgical implant that provides non-constricting protection for peripheral nerve injuries. Axoguard HA+ Nerve Protector is designed to aid in coaptation and protection of peripheral nerve injuries by serving as an interface between the nerve and the surrounding tissue and also provides tension relief when used as a coaptation aid. Axoguard HA+ Nerve Protector is comprised of an extracellular matrix (ECM) and is fully remodeled during the healing process. When hydrated, Axoguard HA+ Nerve Protector is easy to handle, soft, pliable, nonfriable, and porous. The lubricant coating on Axoguard HA+ Nerve Protector is composed of sodium hyaluronate and sodium alginate. When hydrated, the lubricant coating reduces friction between the nerve and the surrounding tissue. Axoguard HA+ Nerve Protector is flexible to accommodate movement of the joint and has sufficient mechanical strength to hold sutures. · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00850828007121), received between Nov 12, 2024 and Apr 3, 2026 — classified as 5 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Appropriate Device Problem Term/Code Not Available2
- Insufficient Device Problem Information1
Patient problems
- Bacterial Infection1
- Erythema1
- Hypersensitivity/Allergic reaction1
- Impaired Healing1
- Muscle Weakness/Atrophy1
- Numbness1
- Pain1
- Purulent Discharge1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryApr 3, 2026Adverse Event Without Identified Device or Use Problem
THIS IS AN INITIAL, SERIOUS, SPONTANEOUS REPORT RECEIVED FROM A SALES REPRESENTATIVE VIA SURGEON REGARDING A PATIENT WHO UNDERWENT TARSAL TUNNEL DECOMPRESSION IN WHICH AXOGUARD HA+ NERVE PROTECTOR WAS IMPLANTED AND EXPERIENCED PURULENT DISCHARGE. ON (B)(6) 2025, THE PATIENT UNDERWENT A TARSAL TUNNEL DECOMPRESSION WHERE AXOGUARD HA+ NERVE PROTECTOR WAS IMPLANTED. THE SURGEON LAID THE HA+ FLAT OVER THE NERVE AND DID NOT SUTURE IT IN PLACE. IN (B)(6) 2026, THE SURGEON WAS MADE AWARE OF CONCERNS FOR A POSSIBLE INFECTION WITH PURULENT DISCHARGE AT THE SURGICAL SITE. DURING EXPLANTATION, THE WRAP FELL APART INTO TWO PIECES AND THE SURGEON NOTED THAT IT ALMOST LOOKED LIKE HA+ WAS EXTRUDING, AS THE MATERIAL WAS LOCATED CLOSE TO THE INCISION SITE UPON OPENING. THE SURGEON CLEANED THE SITE AND PERFORMED CLOSURE IN THE CLINIC.
Patient problems: Purulent Discharge
MAUDE report 24780735 ↗InjuryFeb 15, 2026Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2025, FEMALE PATIENT UNDERWENT CARPAL TUNNEL REVISION SURGERY. AT THE ONE-WEEK POST-OP APPOINTMENT, PATIENT HAD REDNESS AT THE INCISION SITE. SURGEON PUT PATIENT ON A FEW DAYS OF DOXYCYCLINE. AT THE TWO-WEEK POST-OP APPOINTMENT, PATIENT HAD WORSENING REDNESS AND PAIN AT THE SITE. SURGEON TOOK PATIENT BACK TO THE OR, IRRIGATED THE SITE, AND NOTICED THE HA+ PROTECTOR LOOKED LIKE A PIECE OF FATTY COLLAGEN. THE HA+ PROTECTOR WAS EXPLANTED AND FELL APART IN TWO PIECES. THE SURGEON HAD CULTURES SENT WHICH WERE NEGATIVE FOR INFECTION, DID NOT SEND FOR PATHOLOGY. SINCE THE HA+ PROTECTOR WAS EXPLANTED, EVENTS HAVE RESOLVED. NO CAUSALITY WAS REPORTED, THOUGH THE SURGEON REMARKED THAT THE PATIENT MAY HAVE NOT TOLERATED THE DEVICE. AXOGEN ASSESSED CAUSALITY AS POSSIBLY RELATED. CULTURES OBTAINED DURING EVALUATION WERE NEGATIVE, AND NO PATHOLOGY WAS PERFORMED, LIMITING THE ABILITY TO DETERMINE THE UNDERLYING CAUSE OF THE REPORTED REDNESS AND PAIN.
Patient problems: Erythema; Pain
MAUDE report 24341672 ↗InjuryNov 25, 2025Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
ON (B)(6) 2025, PATIENT UNDERWENT A CARPAL TUNNEL REVISION WITH AXOGUARD HA+ (LOT UNKNOWN TO REP; FOUND WITH INTERNAL RECORDS AS LB1620172) IMPLANTATION. ON (B)(6) 2025, PATIENT UNDERWENT THIRD CARPAL TUNNEL REVISION WITH A POSSIBLE NEUROMA EXCISION. SURGEON DID NOT FIND ANY NERVE DAMAGE OR NEUROMA, INSTEAD THERE WAS ONLY HYPERSENSITIVITY AROUND THE SURGICAL SITE. DURING 27-OCT SURGERY, SURGEON PULLED OUT A PIECE OF FATTY TISSUE BUT DID NOT EXPLANT THE HA+. NO CULTURES/PATHOLOGY REPORTED. NO TREATMENT INFORMATION, SURGICAL TECHNIQUE, OR OUTCOMES REPORTED. SURGEON REPORTED CAUSALITY AS POSSIBLY RELATED. AXOGEN REPORTED CAUSALITY AS UNASSESSABLE. BASED ON THE LIMITED INFORMATION AND NO CUTLURES/PATHOLOGY REPORTS, IT IS IMPOSSIBLE TO MAKE A FIRM CONCLUSION. WE ARE UNABLE TO DETERMINE IF OTHER FACTORS, SUCH AS SURGICAL TECHNIQUE, COMORBIDITIES, OR POST-OPERATIVE CARE CONTRIBUTED TO THIS OUTCOME.
Patient problems: Hypersensitivity/Allergic reaction
MAUDE report 23646409 ↗InjuryJul 30, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
THE PATIENT REPORTED A COMPLETE LOSS OF SENSATION ON THE LEFT SIDE OF THEIR HAND THREE WEEKS FOLLOWING A PRC SURGERY, WHICH OCCURRED PRIOR TO THE IMPLANTATION OF THE AXOGUARD HA+. UPON FURTHER EXAMINATION, THE PATIENT WAS DIAGNOSED WITH MEDIAN NERVE ENTRAPMENT ASSOCIATED WITH CARPAL TUNNEL SYNDROME. THE AXOGUARD HA+ WAS SUBSEQUENTLY IMPLANTED DURING A CARPAL TUNNEL SURGERY APPROXIMATELY 2.5 MONTHS AGO, WITH THE SPECIFIC DATE RECORDED AS (B)(6) 2025. AFTER THE CARPAL TUNNEL AND AXOGUARD HA+ IMPLANTATION PROCEDURE, THE SURGICAL WOUND EXPERIENCED CHALLENGES IN HEALING EFFECTIVELY FOR A PERIOD OF ABOUT THREE WEEKS. ULTIMATELY, THE DEVICE WAS EXPLANTED DURING A REVISION SURGERY AT AN UNSPECIFIED DATE THEREAFTER. THE REVISION SURGERY LIKELY INVOLVED THE REMOVAL OF PART OR ALL OF THE AXOGUARD HA+ IMPLANT. DURING THIS PROCEDURE, THE SURGEON UTILIZED A TOURNIQUET ON THE AFFECTED ARM AND RE-CLOSED THE WOUND. SINCE THE REVISION, THE PATIENT HAS REPORTED ONGOING AND PROGRESSIVELY WORSENING NUMBNESS IN THE HAND, WHICH HAS RESULTED IN MUSCLE ATROPHY AND SIGNIFICANT SCARRING AROUND THE SURGICAL SITE. THE PATIENT HAS INDICATED THAT, DESPITE LOCAL WOUND CARE SUPPORT AND THE USE OF STERI-STRIPS, THE
Patient problems: Impaired Healing; Muscle Weakness/Atrophy; Numbness
MAUDE report 22663082 ↗InjuryNov 12, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
THE PATIENT HAD A REVISION CARPAL TUNNEL RELEASE SURGERY ON (B)(6) 2024, DURING WHICH AN AXOGUARD HA+ NERVE PROTECTOR WAS IMPLANTED. THE PATIENT WAS REPORTED DOING WELL DURING HER FOLLOW-UP VISIT, WITH STITCHES BEING REMOVED ON (B)(6) 2024. HOWEVER, SHORTLY AFTER, SHE EXPERIENCED DEHISCENCE OF THE PROXIMAL WOUND AND RETURNED TO THE CLINIC ON (B)(6) 2024 FOR EVALUATION. ON (B)(6) 2024, THE PATIENT UNDERWENT A DRAINAGE AND WASHOUT OF THE WOUND FOR A SUSPECTED INFECTION AND IT WAS CONFIRMED AT THAT TIME THAT THE PROTECTIVE COVERING REMAINED INTACT, HOWEVER THE AXOGUARD HA+ NERVE PROTECTOR WAS PROPHYLACTICALLY REMOVED TO ELIMINATE ANY POTENTIAL RISKS FROM A FOREIGN BODY IN AN INFECTED FIELD. A CULTURE TAKEN FROM THE FLUID IN THE WOUND IDENTIFIED ENTEROBACTER CLOACAE. CONSIDERING THE DEVICE WAS INTACT THROUGHOUT THE POSTOPERATIVE PERIOD, IT IS HIGHLY UNLIKELY THAT IT CONTRIBUTED TO THE WOUND COMPLICATION.
Patient problems: Bacterial Infection
MAUDE report 20666085 ↗
Other Cuff, Nerve devices
- Axoguard HA+ Nerve Protector AGHA36Axogen Corporation · Cuff, Nerve13 reports
- Axoguard HA+ Nerve Protector AGHA48Axogen Corporation · Cuff, Nerve6 reports
- VersaWrap VTP-2201Alafair Biosciences, Inc. · Cuff, Nerve2 reports
- NeuraWrap® NW1040Integra Lifesciences Corporation · CUFF, NERVE2 reports
- NeuraGen® PNG230Integra Lifesciences Corporation · CUFF, NERVE2 reports
- NeuraGen® PNG630Integra Lifesciences Corporation · CUFF, NERVE1 report
- NervAlign Nerve Cuff NA-NC3040Renerve Limited · Cuff, Nerve1 report
- NeuroMend CNW12050Collagen Matrix, Inc. · Cuff, Nerve1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page