Remote Network Station RNS-6803-7460Recalls & reported problems

Nihon Kohden Digital Health Solutions, LLC · Remote Network Station with Dell OptiPlex 7460 23.8" All-In-One Computer · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00851725007184), received between Feb 23, 2021 and Aug 13, 2025 — classified as 1 death, 3 malfunctions.

The most frequently reported specific device problem is “Protective Measures Problem” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Protective Measures Problem3
  • Device Alarm System2
  • No Audible Alarm2
  • Application Program Problem1
  • Audible Prompt/Feedback Problem1
  • Communication or Transmission Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 13, 2025Device Alarm System; Protective Measures Problem; Device Alarm System; Protective Measures Problem

    THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE REMOTE NETWORK STATION (RNS) DID NOT ALARM FOR A PATIENT'S VENTRICULAR TACHYCARDIA. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22795597 ↗
  • MalfunctionMar 4, 2023Communication or Transmission Problem; Communication or Transmission Problem

    THE CUSTOMER IT/IS CONTACT REPORTED THAT ALL OF THEIR DEVICES ARE INTERMITTENTLY GOING IN AND OUT COMMUNICATION LOSS FOR ROUGHLY 30 SECONDS AT A TIME AND THEN COME BACK UP. THE CENTRAL NURSE'S STATION (CNS) / REMOTE NETWORK STATION (RNS) ARE ALL WIRED. THE BEDSIDE MONITORS ARE USING WIRELESS PACKS. THEY DO NOT KNOW IF ALL OF THE WIRELESS PACKS ARE DISCONNECTING ALL AT ONCE OR IF THE WHOLE NETWORK IS BECAUSE IT SHOWS COMM LOSS ON ALL TILES, SO IT IS HARD TO DETERMINE IF THE CNS IS GOING DOWN OR NOT AS WELL (IF THIS IS HAPPENING ON ONLY WIRELESS DEVICES OR ALL DEVICES). THE CUSTOMER SAID THAT THEY RECENTLY REPLACED ALL THE ACCESS POINTS (APS), AND THEY HAVE BEEN WORKING NORMALLY UNTIL NOW. NIHON KOHDEN COMPUTER INFORMATION TECHNOLOGY SYSTEMS SUPPORT (CITS) REMOTED INTO THE SERVERS. CITS CHECKED THE HOST 1000 SERVER AND NOTICED THAT THE SERVER WAS SET TO BROADCAST MODE, WHICH MEANT NO ISSUES THERE. CITS ALSO CHECKED THE DEVICES AND NOTICED THAT THESE WIRELESS DEVICES WERE NOT UTILIZING THE EG FOR WIRELESS COMMUNICATION. CITS TOOK A PACKET CAPTURE BUT DID NOT SEE ANY RELEVANT EVIDENCE OF LARGE DROPOUTS AND THE SERVERS ARE NOT INVOLVED WITH THIS ISSUE. TECH SUPPORT (TS) AND QA CONTACTED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16483285 ↗
  • DeathJul 18, 2022No Audible Alarm; Protective Measures Problem; Audible Prompt/Feedback Problem; No Audible Alarm; Protective Measures Problem; Audible Prompt/Feedback Problem

    THE CUSTOMER REPORTED THAT THIS REMOTE NURSE'S STATION (RNS) DID NOT ALARM; HOWEVER, THE CENTRAL NURSE'S STATION (CNS) DID ALARM. THIS CAUSED A DELAY IN TREATMENT AND THE PATIENT PASSED AWAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15048674 ↗
  • MalfunctionFeb 23, 2021Device Alarm System; No Audible Alarm; Application Program Problem; Protective Measures Problem; Device Alarm System; No Audible Alarm; Application Program Problem; Protective Measures Problem

    THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE REMOTE NETWORK SYSTEM (RNS) DID NOT HAVE AN AUDIBLE ALARM FOR THE SPO2 ALARM AND SAID THAT THE PATIENT SPO2 WENT DOWN TO 40. THE RNS WAS SHOWING THE ALARM VISUALLY, BUT IT WAS NOT MAKING AN AUDIBLE SOUND FOR IT. THEY WERE ABLE TO GET THE AUDIBLE ALARM TO WORK AGAIN BY REBOOTING THE RNS. THERE WAS NO PATIENT DEATH, BUT THEY DO NOT KNOW IF THERE WAS INJURY. THE NURSES TOLD HIM THAT THE PATIENT WAS CLOSE TO DEATH. THE BIOMED STATED THAT THE STAFF WAS ABLE TO PROVIDE AID TO THE PATIENT BECAUSE THERE WAS ANOTHER RNS WHICH WAS GIVING AN AUDIBLE ALARM AND THAT WAS HOW THE STAFF WAS ALERTED AND INFORMED. THE BEDSIDE MONITOR ON THE PATIENT WAS GIVING AN AUDIBLE ALARM AS WELL. THIS PATIENT WAS A COVID PATIENT, AND THE DOOR WAS CLOSED FOR THIS ROOM, SO THEY COULD NOT HEAR THE AUDIBLE ALARM COMING IT. NORMALLY FOR NON-COVID PATIENTS, THEY WOULD LEAVE THE DOOR OPEN. ONLY THIS RNS WAS NOT GIVING AN AUDIBLE ALARM. THE RNS IS A SECONDARY MONITORING SYSTEM. THE CENTRAL NURSE'S STATION (CNS) IS THE PRIMARY MONITORING SYSTEM. THIS INCIDENT OCCURRED ON (B)(6) 2021, AT AROUND 2 AM. ON BEDNAME: WHP3510-01. THE BIOMED COULD NOT PROVIDE THE MODEL AND SERIAL

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11369863 ↗

Other MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS) devices

All MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS) devices →