On-X Aortic Heart Valve with Conform-X Sewing Ring and Extended Holder ONXACE-19Recalls & reported problems
On-X Life Technologies, Inc. · The On-X® Prosthetic Heart Valve is a bileaflet mechanical heart valve, which consists of an orifice housing and two leaflets. The orifice inflow area has a flared inlet designed to reduce flow turbulence, and the outflow rim consists of leaflet guards designed to protect the leaflets while in the closed position. The leaflets rotate around tabs located within the inner circumference of the orifice ring. In the closed position, each leaflet forms a nominal angle of 40º relative to the plane of the orifice. In the open position, the plane of each leaflet forms a nominal angle of 90deg relative to the plane of the orifice. The leaflets have a travel arc of 50deg to the closed position. The orifice, or housing, is composed of a graphite substrate coated with On-X® Carbon, a pure unalloyed form of pyrolytic carbon. The leaflets consist of On-X® Carbon deposited on a graphite substrate, which is impregnated with 10 weight% tungsten to provide radiopacity. The sewing cuff is constructed of polytetrafluoroethylene (PTFE) fabric mounted on the orifice using titanium retaining rings and 5-0 suture material. This form of sewing cuff attachment to the orifice allows for rotation of the sewing cuff in situ during implantation. Orientation reference marks are provided on the sewing cuff for valve orientation. The On-X® Prosthetic Heart Valve is available in 3 aortic and 2 mitral sewing cuff configurations. Aortic configurations are available in sizes 19, 21, 23, 25, and 27/29 mm. The standard mitral cuff is available in sizes 23, 25, 27/29 and 31/33, while the Conform-X® mitral cuff is available in size 25/33 only. Further detail for the On-X prosthetic heart valve is described in the Summary of Safety and Effectiveness documents for the On-X heart valves: http://www.accessdata.fda.gov/cdrh_docs/pdf/P000037b.pdf (aortic), and http://www.accessdata.fda.gov/cdrh_docs/pdf/P000037S001b.pdf (mitral). · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 1
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00851788001440), received between Nov 18, 2015 and Feb 21, 2025 — classified as 1 death, 4 injuries.
The most frequently reported specific device problem is “Perivalvular Leak” (2 reports).
4 other recalls cover devices cleared under the same FDA submission (P000037); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Perivalvular Leak2
- Insufficient Device Problem Information1
- Sticking1
Patient problems
- Aortic Insufficiency2
- Insufficient Information2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Congestive Heart Failure1
- Insufficiency, Valvular1
- Stroke/CVA1
- Thrombosis1
- Unspecified Infection1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryFeb 21, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE INITIAL REPORT AN AORTIC IMPLANT REGISTRATION CARD (IRC) WAS RECEIVED FOR A PATIENT WITH AN EXISTING AORTIC IMPLANT. ADDITIONAL INFORMATION PROVIDED THAT THE EXPLANT OCCURED DUE TO ¿IMPAIRED LEAFLET MOBILITY WITH CIRCUMFERENTIAL PANNUS BELOW THE MSAVR¿
Patient problems: Insufficient Information
MAUDE report 21435990 ↗InjuryOct 12, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE INITIAL REPORT THE ONXACE-19 DEHISCED BECAUSE OF FRAGILE PUSSED AORTIC ROOT. PATIENT WAS TRANSFERRED AND THINKS THE PATIENT HAD A HOMOGRAFT IMPLANTED. PER THE HOSPITAL REPRESENTATIVE, "VALVE PERFORMED AS WELL AS IT SHOULD." NO ADDITIONAL INFORMATION FORTHCOMING.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Unspecified Infection
MAUDE report 17919571 ↗DeathAug 1, 2022Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
ACCORDING TO THE INITIAL REPORT RECEIVED ON 7/7/2022, ARTIVION CUSTOMER SERVICE RECEIVED A CALL FROM [PATIENT] STATING SHE IS AT THE HOSPITAL AND WAS TOLD SHE HAS A FAULTY ON-X VALVE IMPLANTED ON (B)(6) 2020.
Patient problems: Insufficient Information; Stroke/CVA
MAUDE report 15138000 ↗InjuryNov 13, 2017Perivalvular Leak; Perivalvular Leak
ACCORDING TO THE IMPLANT REGISTRATION CARDS, FEMALE PATIENT IMPLANTED WITH ONXACE-19 (SN (B)(4)) ON (B)(6) 2013 AND REQUIRED INTERVENTION/EXPLANT ON (B)(6) 2017 AND SUBSEQUENT REPLACEMENT VIA ONXAE-23.INDICATION FOR INTERVENTION PER OP NOTES RECEIVED- "THE PATIENT IS A (B)(6)-YEAR-OLD LADY, WHO HAD COARCTATION REPAIR AS A CHILD AND THEN AORTIC VALVE REPLACEMENT AND ASD REPAIR IN 2014. SHE NOW RETURNS SYMPTOMATIC WITH SEVERE PARAVALVULAR LEAK OUTSIDE THE AORTIC VALVE PROSTHESIS AND 3+ POSTERIORLY DIRECTED MITRAL REGURGITATION."
Patient problems: Insufficiency, Valvular
MAUDE report 7025077 ↗InjuryApr 8, 2016Perivalvular Leak; Perivalvular Leak
THIS RE-OPERATION WAS OBSERVED WHEN REVIEWING IMPLANT REGISTRATION INFORMATION IN THE TRACKING SYSTEM DATABASE. ON-X INVESTIGATED AND OBTAINED THE OPERATIVE REPORT FOR THE RE-OPERATION. PER THE OP REPORT, THE PATIENT HAD ENDOCARDITIS (SEVERE AORTIC INSUFFICIENCY AND PARAVALVULAR LEAK, SECONDARY TO ENDOCARDITIS), 5 DAYS POST-OPERATIVELY. THE VALVE HAD DEHISCED (FROM SEWING RING-TO-TISSUE CONNECTION). ACUTE CONGESTIVE HEART FAILURE, AND ACQUIRED COAGULOPATHY. RE-DO AORTIC VALVE REPLACEMENT (AVR) WAS PERFORMED WITH AN ONXACE-23 , S/N (B)(4). POST-OP, PATIENT TRANSFERRED TO ICU IN STABLE CONDITION. TEE AT END OF CASE SHOWED EXCELLENT VALVE FUNCTION WITH MINIMAL GRADIENT. ENDOCARDITIS IS AN EXPECTED ADVERSE EVENT, DEFINED IN THE AATS/STS GUIDELINES AS VALVE-RELATED AND REPORTABLE, THEREFORE THIS IS BEING REPORTED. IT IS AMONG SEVERAL OTHER ADVERSE EVENTS POSSIBLE FOR HEART VALVE PATIENTS AS LISTED IN SECTION 5 OF THE IFU.
Patient problems: Aortic Insufficiency; Congestive Heart Failure
MAUDE report 5564710 ↗Nov 18, 2015Sticking
PATIENT ADMITTED FOR AVR (AORTIC VALVE REPLACEMENT). ROOT ENLARGEMENT PERFORMED IN ORDER TO FIT A 19MM AORTIC VALVE. VALVE IMPLANTED SUCCESSFULLY. INTRA-OPERATIVE TEE SHOWED VALVE LEAFLETS TO BE FUNCTIONING PROPERLY. UPON FOLLOW-UP AFTER SURGERY (35 DAYS POST-OP), ECHO SHOWED ONE LEAFLET NOT CLOSING WELL, CAUSING AORTIC INSUFFICIENCY (AI), BUT PATIENT TOLERATED IT FINE. NO SUBVALVULAR OBSTRUCTION CAN BE SEEN ON ECHO AND NO OBSERVED THROMBOSIS OR SUTURE IMPINGEMENT, THOUGH THIS CANNOT ALWAYS BE SEEN ON ECHO. A SMALL THROMBUS WAS SUSPECTED AND LOW DOSE LYTIC THERAPY ADMINISTERED ON (B)(6) 2015 AND AGAIN ON (B)(6) 2015. REPEAT TEE SHOWING OFFENDING LEAFLET NOW MOVING MUCH BETTER, AI ONE PLUS. TARGET INR 3.5 PLUS ASPIRIN GOING FORWARD. AS THE ISSUE WAS CLEARED UP WITH LYTIC THERAPY, VALVE THROMBOSIS IS THE SUSPECTED CONDITION. THIS IS BEING REPORTED BECAUSE IT REQUIRED MEDICAL INTERVENTION TO PREVENT PERMANENT DAMAGE, AND VALVE THROMBOSIS IS DEFINED TO BE VALVE-RELATED AND REPORTABLE PER AATS/STS GUIDELINES.
Patient problems: Aortic Insufficiency; Thrombosis
MAUDE report 5233125 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P000037. One submission often covers many models, so check the product and lot details in each record.
- Z-2133-2026 · Feb 1, 2026 · Open, ClassifiedOn-X Mitral Heart Valve with Conform-X Sewing Ring, Model: ONXMC-25/33
- Z-1073-2019 · Jan 17, 2019 · TerminatedOn-X Aortic Heart Valve with Anatomic Sewing Ring Product Usage: The On-X Prosthetic Heart Valve is indicated for the replacement of diseased, damaged, or malfunctioning native
- Z-2715-2017 · Jun 12, 2015 · TerminatedOn-X Prosthetic Heart Valve Product Usage: Indicated for the replacement of diseased, damaged, or malfunctioning native or prosthetic heart valves in the aortic and mitral positio
- Z-3108-2017 · Feb 12, 2014 · TerminatedOn-X Prosthetic Heart Valve with Gelweave Valsalva Graft, single use. Indicated for the replacement of diseased, damaged, or malfunctioning native or prosthetic heart valves in th
Other Heart-Valve, Mechanical devices
- On-X Mitral Heart Valve with Conform-X Sewing Ring ONXMC-25/33On-X Life Technologies, Inc. · Heart-Valve, Mechanical1 recall1 death34 reports
- SJM Regent™ AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL3 deaths70 reports
- SJM Regent™ AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL5 deaths66 reports
- SJM Regent™ 23AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL3 deaths44 reports
- CARBOMEDICS SUPRA-ANNULAR (TOP HAT) S5-021Corcym Srl · HEART-VALVE, MECHANICAL7 deaths37 reports
- Regent™ 21AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL3 deaths35 reports
- Regent™ 23AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL2 deaths33 reports
- Regent™ 19AGFN-756St. Jude Medical, Inc. · HEART-VALVE, MECHANICAL31 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page