ThermoSuit System TSP-50 Pump/ControllerRecalls & reported problems
Life Recovery Systems Hd LLC · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00851936007003), received between Feb 16, 2026 and Feb 16, 2026 — classified as 1 malfunction.
The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (1 report).
1 other recall cover devices cleared under the same FDA submission (K061023); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Electrical /Electronic Property Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 16, 2026Electrical /Electronic Property Problem
CLINICIANS AT (B)(6) MEDICAL CENTER USED THE LIFE RECOVERY SYSTEMS THERMOSUIT SYSTEM TO COOL A PATIENT FOLLOWING CARDIAC ARREST IN ACCORDANCE WITH THEIR STANDARD CLINICAL PRACTICE. THE TSP-50 PUMP WAS ACTIVATED WITH THE THERMOSUIT BODY ENCLOSURE (TBE) APPLIED TO THE PATIENT. THE TOP SHEET WAS ATTACHED TO THE TBE OVER THE FRONT OF THE PATIENT, LEAVING THE FACE EXPOSED. THE TSP-50 RESERVOIR WAS FILLED WITH APPROXIMATELY 40 LITERS OF ICE WATER. A 400-SERIES THERMISTOR TEMPERATURE PROBE FROM THE DISPOSABLES KIT WAS INSERTED INTO THE PATIENT'S ESOPHAGUS AND CONNECTED TO THE TSP-50 FOR CONTINUOUS TEMPERATURE MONITORING. SHORTLY AFTER ACTIVATION OF THE WATER PUMPS, THE TSP-50 EXPERIENCED INTERMITTENT POWER LOSS AND STOPPED CIRCULATING WATER THROUGH THE ENCLOSURE, REPRESENTING A DEVICE MALFUNCTION AND INTERRUPTION OF COOLING THERAPY. THE ATTENDING PHYSICIAN CONTACTED LRS TECHNICAL SUPPORT. DURING THE CALL, TECHNICAL SUPPORT DISCUSSED ALTERNATIVE TEMPERATURE MONITORING CONFIGURATION USING THE 400-SERIES PROBE AND A NEARBY COMPATIBLE TEMPERATURE MONITOR. CLINICAL STAFF REMOVED THE TOP SHEET FROM THE ENCLOSURE AND MANUALLY TRANSFERRED WATER FROM THE RESERVOIR ONTO THE FRONT OF THE PATIENT USI
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24345778 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K061023. One submission often covers many models, so check the product and lot details in each record.
- Z-0456-2012 · Oct 21, 2011 · TerminatedThermoSuit Manufactured for Life Recovery Systems. HD, LLC, 170 Kinnelon Rd. Kinnelon, NJ 07405 USA Tel.: (973) 283-2800 FAx.: (973) 283-2910 Temperature reduction in patients wh
Other Hyperthermia Monitor devices
- Thermogard™ Vest, L/XLZoll Circulation, Inc. · Hyperthermia Monitor7 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page