Sniper SBC0629-STR-150Recalls & reported problems

Embolx, Inc. · The Sniper is a dual lumen, 150cm, straight tip catheter comprised of an inner guidewire/infusion lumen and an outer, annular lumen to inflate and deflate the balloon. The guidewire/infusion port may be used to inject contrast media or therapeutic agents following removal of the guidewire. The catheter has two radiopaque markers: one marker at the distal tip and another in the center of the balloon. · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00853359008032), received between May 19, 2020 and Nov 1, 2023 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Manufacturing, Packaging or Shipping Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Manufacturing, Packaging or Shipping Problem2
  • Insufficient Device Problem Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Unintended Radiation Exposure2
  • Intimal Dissection1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 1, 2023

    PATIENT UNDERGOING YTTRIUM-90 RADIOEMBOLIZATION OF LIVER METASTASIS FROM RECTAL CANCER. BALLOON OCCLUSION CATHETER DEPLOYED IN RIGHT HEPATIC ARTERY OSTENSIBLY TO DIRECT RADIOACTIVE SPHERES TO THE TARGET AREA WHILE PREVENTING NONTARGET EMBOLIZATION TO BOWEL AND THE REMAINDER OF THE LIVER NOT CONTAINING TUMOR. TWO CONSECUTIVE SNIPER BALLOON MICROCATHETERS FAILED, THE FIRST BEFORE ANY RADIOACTIVE BEADS HAD BEEN ADMINISTERED THROUGH IT AND THE SECOND DURING ADMINISTRATION OF AN ALIQUOT OF RADIOACTIVE BEADS. REMOVING THE SECOND MICROCATHETER EXPOSED THE PATIENT TO THE POSSIBILITY OF REFLUX OF THE PARTICLES INTO THE GASTRODUODENAL ARTERY AND NONTARGET EMBOLIZATION TO THE BOWEL AND REMAINDER OF THE LIVER. ULTIMATELY, A THIRD BALLOON OCCLUSION MICROCATHETER BY DIFFERENT MANUFACTURER WAS ABLE TO BE SUCCESSFULLY USED TO ACHIEVE TREATMENT WITHOUT A MISADMINISTRATION (DEFINED AS DELIVERING LESS THAN 75% OF THE DOSE IN THE WRITTEN DIRECTIVE) AND WITHOUT NONTARGET EMBOLIZATION OF THE RADIOEMBOLIZATION TO BOWEL. THE NATURE OF THE FAILURE HAS BEEN SEEN PREVIOUSLY WITH THE CATHETER - SPECIFICALLY, THE BALLOON INFLATION LUMEN COMMUNICATES WITH THE TRUE CATHETER LUMEN AND THE BALLOON DECOMPRESSES SPO

    Patient problems: Unintended Radiation Exposure

    MAUDE report 18055595 ↗
  • InjuryNov 1, 2023

    PATIENT UNDERGOING YTTRIUM-90 RADIOEMBOLIZATION OF LIVER METASTASIS FROM RECTAL CANCER. BALLOON OCCLUSION CATHETER DEPLOYED IN RIGHT HEPATIC ARTERY OSTENSIBLY TO DIRECT RADIOACTIVE SPHERES TO THE TARGET AREA WHILE PREVENTING NONTARGET EMBOLIZATION TO BOWEL AND THE REMAINDER OF THE LIVER NOT CONTAINING TUMOR. TWO CONSECUTIVE SNIPER BALLOON MICROCATHETERS FAILED, THE FIRST BEFORE ANY RADIOACTIVE BEADS HAD BEEN ADMINISTERED THROUGH IT AND THE SECOND DURING ADMINISTRATION OF AN ALIQUOT OF RADIOACTIVE BEADS. REMOVING THE SECOND MICROCATHETER EXPOSED THE PATIENT TO THE POSSIBILITY OF REFLUX OF THE PARTICLES INTO THE GASTRODUODENAL ARTERY AND NONTARGET EMBOLIZATION TO THE BOWEL AND REMAINDER OF THE LIVER. ULTIMATELY, A THIRD BALLOON OCCLUSION MICROCATHETER BY DIFFERENT MANUFACTURER WAS ABLE TO BE SUCCESSFULLY USED TO ACHIEVE TREATMENT WITHOUT A MISADMINISTRATION (DEFINED AS DELIVERING LESS THAN 75% OF THE DOSE IN THE WRITTEN DIRECTIVE) AND WITHOUT NONTARGET EMBOLIZATION OF THE RADIOEMBOLIZATION TO BOWEL. THE NATURE OF THE FAILURE HAS BEEN SEEN PREVIOUSLY WITH THE CATHETER - SPECIFICALLY, THE BALLOON INFLATION LUMEN COMMUNICATES WITH THE TRUE CATHETER LUMEN AND THE BALLOON DECOMPRESSES SPO

    Patient problems: Unintended Radiation Exposure

    MAUDE report 18055574 ↗
  • MalfunctionApr 16, 2022Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    IT WAS REPORTED PER HOSPITAL MEDWATCH REPORT #13450705 THAT THE HYDROPHILIC COATING ON THE CATHETER FAILED. THE CATHETER WAS REMOVED AND REPLACED WITH ANOTHER OF THE SAME WITH NO HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14131795 ↗
  • MalfunctionApr 16, 2022Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    IT WAS REPORTED PER HOSPITAL MEDWATCH REPORT # (B)(4): DEFECTIVE COATING ON THE CATHETER CAUSED A KINK IN THE CATHETER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14131789 ↗
  • MalfunctionMay 19, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT, DURING PROSTATE ARTERY EMBOLIZATION PROCEDURE, THE DEVICE BALLOON WAS INFLATED IN 2MM VESSEL, THEN EMBOLIZED WITH 40-80 EMBOZINE TO STASIS. THE BALLOON WOULD NOT DEFLATE AFTER SEVERAL ATTEMPTS. THE DR. DECIDED TO OVER INFLATE TO BURST. THE DR. THEN PULLED BACK THE DEVICE AND THERE WAS INTIMA DAMAGE TO WALL OF VESSEL. THE DR. THEN EMBOLIZED PROXIMAL TO INJURY WITH COILS TO STOP FLOW INTO INTIMA OF VESSEL WALL.

    Patient problems: Intimal Dissection

    MAUDE report 10070011 ↗

FDA clearance and classification

Cleared through 510(k) K180904 on Jun 8, 2018 (decision: Substantially Equivalent). Submitted by Embolx, Inc. as “Sniper Infusion Catheter with Balloon Occlusion”. The FDA review took 63 days. FDA record ↗

The FDA regulates product code MJN (catheter, intravascular occluding, temporary) under 21 CFR 870.4450 as a Class II device. 21 CFR 870.4450 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Sniper SBC0629-STR-150

Has the Sniper SBC0629-STR-150 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00853359008032), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Sniper SBC0629-STR-150?

The FDA's MAUDE database lists 5 adverse event reports for this device, received between May 19, 2020 and Nov 1, 2023. In those reports, the most common specific device problem is “Manufacturing, Packaging or Shipping Problem” (2); the most common patient problems are “Unintended Radiation Exposure” (2) and “Intimal Dissection” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Sniper SBC0629-STR-150?

None of the reports are classified as deaths, 2 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Sniper SBC0629-STR-150 FDA approved?

It was cleared through 510(k) K180904 on Jun 8, 2018. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Sniper SBC0629-STR-150, and is it still sold?

The FDA's device database lists Embolx, Inc. as the company responsible for it (brand Sniper, model SBC0629-STR-150). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Catheter, Intravascular Occluding, Temporary devices