FreeClimb 70 reperfusion catheter, 132cm with Tenzing 7 Gen 1Recalls & reported problems

Route 92 Medical · FreeClimb 70 reperfusion catheter, 132cm with Tenzing 7; Contains Tenzing 7 Delivery Catheter (00853799007541) and FreeClimb 70 reperfusion catheter, 132cm (00853799007183) · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00853799007497), received between Mar 20, 2024 and Mar 20, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 8, 2024Class IOpen, Classified

    Their is the potential for distal tip separation at the proximal marker band of the delivery catheters

    Z-1611-2024 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Insufficient Device Problem Information1

Patient problems

  • Vascular Dissection1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 20, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    PATIENT HAD A LEFT ICA TERMINUS OCCLUSION. A FIRST PASS WITH FREECLIMB 70 REPERFUSION SYSTEM AND SYNCHRO SELECT STANDARD LEFT SOME RESIDUAL CLOT IN THE MID PORTION OF THE M1 SEGMENT. DURING THE SECOND PASS THE DEVICES WERE ADVANCED INTO MID M1 SEGMENT. TENZING 7 AND SYNCHRO WIRE WERE REMOVED AND ASPIRATION WAS PERFORMED THROUGH FREECLIMB 70 FOR ROUGHLY 1-2 MINUTES. FREECLIMB 70 CATHETER WAS RETRACTED INTO THE ZOOM 88 CATHETER AT THE HORIZONTAL PETROUS SEGMENT OF THE LEFT ICA. A CONTRAST INJECTION THROUGH THE ZOOM 88 CATHETER IDENTIFIED EXTRAVASATION AT LEFT M1 SEGMENT. A DISSECTION OCCURRED.

    Patient problems: Vascular Dissection

    MAUDE report 18944700 ↗