Acessa ProVu Transducer with sensor 7700Recalls & reported problems
Acessa Health Inc. · Acessa ProVu Transducer with sensor, Reusable, connects to the Acessa ProVu Console · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00854763006188), received between Jul 3, 2019 and Jul 18, 2019 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Scratched Material” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Scratched Material2
Patient problems
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 18, 2019Scratched Material; Scratched Material
ON 6/20/2019, IT WAS REPORTED THAT ABOUT 75% THROUGH A CASE, THAT THE ACESSA PROVU TRANSDUCER WITH SENSOR (7700) WAS IDENTIFIED AS HAVING A SURFACE DEFECT ON THE SILICONE ACOUSTIC LENS OF THE TRANSDUCER. PRIOR TO THAT POINT IN THE CASE, NEITHER THE OPERATING PHYSICIAN NOR ANY MEMBER OF THE SURGICAL STAFF HAD OBSERVED ANY DAMAGE TO THE TRANSDUCER. UPON RECOGNITION OF THE DAMAGED SILICONE, THE CAVITY WAS SUCTIONED AND CLEARED TO ENSURE REMOVAL OF ANY POTENTIAL DEBRIS, WITH A STATEMENT OF CONFIDENCE FROM THE OPERATING PHYSICIAN STATING THAT THE CAVITY HAD BEEN APPROPRIATELY CLEARED. THE CASE WAS COMPLETED SUCCESSFULLY, THERE WAS NO REPORTED HARM, INJURY, ADVERSE OUTCOME, OR PATIENT COMPLICATIONS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8804606 ↗MalfunctionJul 3, 2019Scratched Material; Scratched Material
ON 6/7/2019, IT WAS REPORTED THAT DURING A CASE, THE ACESSA PROVU HANDPIECE (7300) MADE CONTACT WITH THE ACESSA PROVU TRANSDUCER WITH SENSOR'S (7700) ACOUSTIC LENS. THE OPERATING PHYSICIAN INADVERTENTLY MADE CONTACT BETWEEN THE DISTAL TIP OF THE HANDPIECE AND THE SILICONE LENS OF THE TRANSDUCER. THE CONTACT OF THE SHARP DISTAL MANDREL OF THE HANDPIECE WITH THE SILICONE SCANNING WINDOW RESULTED IN DAMAGE TO THE SILICONE LAYER OF THE LENS IN TWO SEPARATE LOCATIONS. UPON RECOGNITION OF THE DAMAGED SILICONE, THE CAVITY WAS SUCTIONED AND CLEARED TO ENSURE REMOVAL OF ANY POTENTIAL DEBRIS, WITH A STATEMENT OF CONFIDENCE FROM THE OPERATING PHYSICIAN STATING THAT THE CAVITY HAD BEEN APPROPRIATELY CLEARED. THE CASE WAS COMPLETED SUCCESSFULLY, THERE WAS NO REPORTED HARM, INJURY, ADVERSE OUTCOME, OR PATIENT COMPLICATIONS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8761571 ↗
Other System, Imaging, Pulsed Echo, Ultrasonic devices
- EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE OLYMPUS GF TYPE UCT180Olympus Medical Systems Corp. · System, imaging, pulsed echo, ultrasonic1 recall8 deaths5,133 reports
- Site~Rite 8 Ultrasound System 9770600Bard Access Systems, Inc. · System, imaging, pulsed echo, ultrasonic629 reports
- Site-Rite 9770501Bard Access Systems, Inc. · System, imaging, pulsed echo, ultrasonic546 reports
- Transducer X8-2t 989605455171Philips Ultrasound LLC · System, imaging, pulsed echo, ultrasonic8 recalls19 reports
- Transducer C10-3v 989605409562Philips Ultrasound LLC · System, imaging, pulsed echo, ultrasonic7 recalls0 reports
- Site~Rite 8 Ultrasound System 9770600YBard Access Systems, Inc. · System, imaging, pulsed echo, ultrasonic330 reports
- Transducer C9-2 989605391092Philips Ultrasound LLC · System, imaging, pulsed echo, ultrasonic5 recalls0 reports
- Transducer S5-1 989605412082Philips Ultrasound LLC · System, imaging, pulsed echo, ultrasonic5 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page