Tourni-Cot - Small TCSRecalls & reported problems
Mar-Med Inc. · Exsanguinating digital tourniquets to create a bloodless field for procedures on fingers and toes. · Class 1
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00855364004030), received between Jan 2, 2026 and Jan 2, 2026 — classified as 1 injury.
The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).
1 other recall cover devices cleared under the same FDA submission (K853472); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Improper or Incorrect Procedure or Method1
Patient problems
- Ischemia1
- Necrosis1
- Tissue Breakdown1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJan 2, 2026Improper or Incorrect Procedure or Method
PATIENT UNDERWENT A SURGICAL PROCEDURE. THE PROCEDURE REQUIRED THE USE OF A DIGITAL TOURNIQUET, WHICH IS REMOVED AT THE END OF THE OPERATION. THE PATIENT UNDERWENT THE PROCEDURE ON THE IDENTIFIED FINGER AS PLANNED WITH THE TOURNIQUET BEING UTILISED AS PER STANDARD PRACTICE. DURING THE PROCEDURE, THE SAFETY TAG WAS REMOVED FROM THE TOURNIQUET TO AID VISIBILITY. THE PROCEDURE WAS COMPLETED AND THE PATIENT DISCHARGED HOME. UPON REVIEW IN THE OUTPATIENTS DEPARTMENT SIX DAYS LATER, THE DIGITAL TOURNIQUET WAS NOTED TO HAVE BEEN LEFT IN-SITU CAUSING THE FINGER TO BECOME SIGNIFICANTLY DAMAGED. THIS LED TO THE DIGIT BEING AMPUTATED THAT AFTERNOON.
Patient problems: Ischemia; Necrosis; Tissue Breakdown
MAUDE report 23952445 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K853472. One submission often covers many models, so check the product and lot details in each record.
- Z-0945-2021 · Dec 28, 2020 · TerminatedTourni-Cot - Universal, UDI 00855364004085, Model Number TCU, Non-pneumatic Tourniquet - Product Usage: The device is intended to be used in surgery on fingers or toes to provide
Other Tourniquet, Nonpneumatic devices
- BD Vacutainer® Stretch Latex Free Tourniquet 367203Becton, Dickinson And Company · TOURNIQUET, NONPNEUMATIC3 reports
- DeRoyal 30-196Deroyal Industries, Inc. · TOURNIQUET, NONPNEUMATIC1 report
- Tourni-Cot - Universal TCUMar-Med Inc. · Tourniquet, Nonpneumatic1 report
- SAM XT Extremity Tourniquet XT600-BK-ENSAM Medical Products · Tourniquet, Nonpneumatic1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.