Skip to content

LOCalizer HB300-05Recalls & reported problems

Health Beacons, Inc. · Tag Applicator S, 5 cm, Single · Class 2 · Implant

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00855436007143), received between Dec 15, 2025 and Jul 27, 2026 — classified as 2 injuries.

The most frequently reported specific device problem is “Failure to Read Input Signal” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Failure to Read Input Signal1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 27, 2026Failure to Read Input Signal

    It was reported that during a LOCalizer procedure on (b)(6) 2026, the "LOCalizer RFID tag would not emit a signal during surgery. Multiple LOCalizer readers and probes were used to troubleshoot." It was further reported that "Radiology confirmed the tag was working prior to placing it and they were able to get a signal of the tag once the specimen was out, and very close to it, but no signal inside the breast." Reporting indicates that in order to identify the tag location, "they had to bring in a C arm during surgery." No patient harm was reported; however, injury MDR is being submitted due to the additional intervention that was required.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50024650 ↗
  • InjuryDec 15, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING A LOCALIZER PROCEDURE ON AN UNKNOWN DATE IN (B)(6) 2025, "PATIENT WITH RIGHT BREAST CANCER UNDERGOING RF [RADIOFREQUENCY] TAG LOCALIZED PARTIAL MASTECTOMY AND SENTINEL NODE BIOPSY. RF TAG PLACED PRIOR TO SURGERY AND TAG SIGNAL CONFIRMED PRIOR TO INCISION. DURING PROCEDURE TAG COULD NOT BE LOCALIZED USING DETECTOR UNIT. SIGNAL WAS DETECTED WITHOUT TAG NUMBER. INTERMITTENTLY THE TAG SIGNAL WITH TAG NUMBER WAS IDENTIFIED. WE ULTIMATELY VISUALIZED THE TAG, AND THE PROBE DID NOT IDENTIFY A SIGNAL WHILE DIRECTLY ON THE TAG. THIS RESULTED IN THE CREATION OF A LARGER LUMPECTOMY SPECIMEN INCLUDING AN ADDITIONAL SUPERIOR MARGIN." NO FURTHER INFORMATION IS CURRENTLY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23809834 ↗

Other Marker, Radiographic, Implantable devices

All Marker, Radiographic, Implantable devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.