Skip to content

NeuroPace® Cortical Strip Lead Kit CL-325-10-K (UDI 00855547005311)Recalls & reported problems

Neuropace, Inc. · The NeuroPace® Cortical Strip Lead package contains 1 Cortical Strip Lead, 2 Lead Caps, 4 Suture Sleeves and 1 Tunneling Tool Straw. · Class 3 · Implant

All models of NeuroPace Cortical Strip Lead Kit (4)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00855547005311), received between Aug 30, 2021 and Feb 19, 2026 — classified as 5 injuries.

The most frequently reported specific device problem is “Difficult or Delayed Positioning” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem4
  • Difficult or Delayed Positioning1
  • Malposition of Device1

Patient problems

  • Intracranial Hemorrhage3
  • Hemorrhage/Blood Loss/Bleeding2
  • Cerebrospinal Fluid Leakage1
  • Partial Hearing Loss1
  • Speech Disorder1
  • Stroke/CVA1
  • Swelling/ Edema1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 19, 2026Adverse Event Without Identified Device or Use Problem

    THE PATIENT EXPERIENCED A SUBDURAL HEMORRHAGE WHEN THE TEMPORAL NON MESIAL CORTICAL LEAD (SECURED WITH A DOG BONE WITH LEAD PROTECTION) MIGRATED AND RUPTURED A BLOOD VESSEL. THE PATIENT WAS HOSPITALIZED FOR ONE WEEK FOR MONITORING. THE TEMPORAL NON MESIAL CORTICAL LEAD SECURED WITH A DOG BONE WAS EXPLANTED. FURTHER DETAILS WERE NOT MADE AVAILABLE TO NEUROPACE AFTER REPEATED ATTEMPTS TO CONTACT THE TREATING PHYSICIAN.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Intracranial Hemorrhage

    MAUDE report 24393945 ↗
  • InjuryDec 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ON (B)(6) 2025, THE PATIENT HAD A SYMPTOMATIC HEMORRHAGE AROUND THE STRIP ELECTRODE COVERING THE LEFT POSTERIOR SUPERIOR TEMPORAL GYRUS. IT APPEARS THE PATIENT HAD MRI PERFORMED ON (B)(6) 2025. THE RESULTS OF THE MRI WERE NOT PROVIDED. THE PATIENT WAS HOSPITALIZED FOR 6 DAYS FOR OBSERVATION. NO SURGICAL EVACUATION WAS NEEDED. NO OTHER TREATMENT WAS INDICATED. NO RESIDUAL NEUROLOGICAL SEQUELAE WERE OBSERVED. THE PHYSICIAN REPORTED THAT HE WAS NOT ENTIRELY CERTAIN THAT IT WAS AN RNS SYSTEM RELATED EVENT, BUT OTHER CAUSES OF HEMORRHAGE HAD BEEN RULED OUT.

    Patient problems: Intracranial Hemorrhage

    MAUDE report 23705977 ↗
  • InjuryNov 25, 2025Difficult or Delayed Positioning; Malposition of Device; Difficult or Delayed Positioning; Malposition of Device

    ON (B)(6) 2025 THE PATIENT UNDERWENT RESECTION AND IMPLANT. POSTOPERATIVELY, THE PATIENT EXHIBITED UNSPECIFIED SPEECH ISSUES. A CT SCAN ON (B)(6) 2025 REVEALED A SMALL HEMORRHAGE BENEATH A COILED LEFT FRONTAL STRIP LEAD WIRE. THE WIRE WAS COILED UNDER THE DURA, COMPRESSING A BLOOD VESSEL AND CAUSING CAPILLARY BACKUP, RESULTING IN THE HEMORRHAGE. ON (B)(6) 2025, THE PATIENT UNDERWENT A PROCEDURE TO REPOSITION THE WIRE. FOLLOWING SUCCESSFUL REPOSITIONING, VENOUS FLOW WAS RESTORED.". THE LEAD WAS SECURED WITH ADHESIVE, AND THE SKULL WAS CLOSED. AS OF (B)(6) 2025, THE PATIENT WAS RECOVERING WELL.

    Patient problems: Intracranial Hemorrhage; Speech Disorder

    MAUDE report 23647791 ↗
  • InjuryAug 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT STARTED TO NOTICE A BOGGY AREA OF FLUID ON HER SCALP AND THEN EXPERIENCED SOME HEARING LOSS. DETAILS ARE UNCLEAR, BECAUSE THE PATIENT WAS SEEN AT YALE AT THIS TIME. ULTIMATELY, A CT WAS COMPLETED. THE CT SHOWED A PSEUDOMENINGOCELE. AT YALE, THE PATIENT UNDERWENT REPAIR OF THE DURA AND DRAINAGE OF THE PSEUDOMENINGOCELE WITH LUMBAR DRAIN PLACEMENT. THE PATIENT NOTICED THE SAME BOGGY AREA AND HEARING LOSS IN THE SPRING OF 2024. A HEAD CT WAS COMPLETED. THE CT SHOWED A PSEUDOMENINGOCELE AROUND THE SAME LEAD. THE PATIENT UNDERWENT REMOVAL OF LEFT TEMPORAL STRIP LEADS X 3. NEW LEFT PULVINAR AND LEFT HIPPOCAMPAL LEADS WERE PLACED. ACCORDING TO THE PA, THE RNS STRIP LEAD ERODED AT THE DURA, THE CHRONIC DURAL OPENING CAUSED A PSEUDOMENINGOCELE TO FORM. THE FLUID DRAINED TOWARD THE EAR AND CAUSED A SENSORINEURAL HEARING LOSS DUE TO THE FLUID BUILDING UP IN THE INNER EAR. THE HEARING LOSS HAS RESOLVED NOW THAT THE FLUID HAS BEEN DRAINED. THE PATIENT IS BACK TO HER BASELINE AND DOING WELL.

    Patient problems: Cerebrospinal Fluid Leakage; Partial Hearing Loss; Swelling/ Edema

    MAUDE report 20066670 ↗
  • InjuryAug 30, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT UNDERWENT SUCCESSFUL PLACEMENT OF AN RNS SYSTEM DEPTH LEAD DURING THE INITIAL IMPLANT PROCEDURE. HOWEVER WHEN ATTEMPTING TO OPEN THE DURA FOR PLACEMENT OF A CORTICAL STRIP LEAD, THE PATIENT EXPERIENCED SIGNIFICANT BLEEDING. THE NEUROSURGEON ABANDONED THE PROCEDURE AND EXPLANTED THE DEPTH LEAD. A LARGER CRANIOTOMY WAS CREATED TO RELIEVE PRESSURE ON THE SKIN. POST-WOUND CLOSURE, THE PATIENT WAS SENT FOR A CT SCAN. THE NEUROSURGEON SUSPECTED A VENOUS INFARCT, CONSIDERING THE PATIENT HAS UNDERGONE PREVIOUS CRANIAL SURGERIES (NOT RELATED TO THE RNS SYSTEM) WHICH LED TO SCAR TISSUE AND DURAL ADHESIONS. AS OF (B)(6) 2021 THE PATIENT IS REPORTED TO BE VERY WEAK AND APHASIC.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Stroke/CVA; Thrombosis/Thrombus

    MAUDE report 12389049 ↗

Other Implanted brain stimulator for epilepsy devices

All Implanted brain stimulator for epilepsy devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page