NeuroPace® Remote Monitor Kit 5107Recalls & reported problems
Neuropace, Inc. · The NeuroPace® Remote Monitor package contains 1 NeuroPace® Remote Monitor, and 1 Carrying Case. · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00855547005427), received between Jan 16, 2026 and Jan 16, 2026 — classified as 1 injury.
The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Electrical /Electronic Property Problem1
- Fire1
Patient problems
- Burn(s)1
- Full thickness (Third Degree) Burn1
- Superficial (First Degree) Burn1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJan 16, 2026Electrical /Electronic Property Problem; Fire
INVESTIGATION RESULTS, REFER TO SECTION H10 ADDITIONAL INFORMATION: THE THERMAL EVENT RESULTED IN THE PATIENT SUSTAINING BURNS TO THEIR LEFT PALM, RIGHT FINGERTIPS, AND RIGHT ARM ABOVE THE ELBOW. WE NOTED A DISCREPANCY IN THE INFORMATION OBTAINED RELATING TO THE PATIENT. THE FDA REPORT SUBMITTED BY THE PHYSICIAN (MW5182362) REPORTED THE PATIENT SUSTAINED SUPERFICIAL (FIRST DEGREE) BURNS. NPCE WAS INFORMED BY THE MOTHER THAT THE PATIENT SUSTAINED THIRD-DEGREE BURNS TO LEFT PALM AND RIGHT FINGERTIPS AND A FIRST-DEGREE BURN TO RIGHT ARM ABOVE THE ELBOW. THE INITIAL MDR REPORTED BY NEUROPACE, INCLUDED THE INFORMATION AS REPORTED BY THE MOTHER (AS THIRD-DEGREE BURN TO LEFT PALM AND RIGHT FINGERTIPS AND FIRST-DEGREE BURN TO RIGHT ARM ABOVE THE ELBOW). ORIGINAL REPORT ON (B)(6) 2026, THE PATIENT'S MOTHER REPORTED THAT THE PATIENT'S REMOTE MONITOR IGNITED AT APPROXIMATELY 9:40 PM ON (B)(6) 2025. ACCORDING TO HER ACCOUNT, THE PATIENT WAS INTERROGATING HIS NEUROSTIMULATOR WITH THE TRM (SN (B)(6)). DURING THAT TIME THE MOTHER REPORTED THE TRM INITIALLY DISPLAYED ITS NORMAL SCREEN, BUT THEN STARTED GLOWING ORANGE, ACTIVELY BURNING, AND PRODUCING SMOKE. THE MOTHER REPORTED THAT THE PATIENT HAD
Patient problems: Burn(s); Full thickness (Third Degree) Burn; Superficial (First Degree) Burn
MAUDE report 24104387 ↗
Other Implanted brain stimulator for epilepsy devices
- RNS® Neurostimulator Kit RNS-320-KNeuropace, Inc. · Implanted brain stimulator for epilepsy2 deaths218 reports
- RNS® Neurostimulator Kit RNS-320-KNeuropace, Inc. · Implanted brain stimulator for epilepsy108 reports
- RNS® Neurostimulator Kit RNS-300M-KNeuropace, Inc. · Implanted brain stimulator for epilepsy105 reports
- NeuroPace® Depth Lead Kit DL-330-10-KNeuropace, Inc. · Implanted brain stimulator for epilepsy30 reports
- NeuroPace® Depth Lead Kit DL-344-10-KNeuropace, Inc. · Implanted brain stimulator for epilepsy26 reports
- NeuroPace® Depth Lead Kit DL-344-10-KNeuropace, Inc. · Implanted brain stimulator for epilepsy23 reports
- NeuroPace® Depth Lead Kit DL-330-10-KNeuropace, Inc. · Implanted brain stimulator for epilepsy22 reports
- NeuroPace® Depth Lead Kit DL-330-3.5-KNeuropace, Inc. · Implanted brain stimulator for epilepsy21 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page