LifeVest Wearable Defibrillator LifeVest 4000 Electrode Belt (REF 10A1025)Recalls & reported problems

Zoll Manufacturing Corporation · Class 3

All models of ZOLL LifeVest (10) · LifeVest Wearable Defibrillator (10)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Zoll LifeVest 4000 is a wearable defibrillator with an electrode belt. Its 4,299 adverse event reports mostly describe circuit failures, broken or structurally damaged parts and incorrect messages on the display, and some describe the belt or monitor becoming unusable. The reports include 15 deaths and 274 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
4,460
Deaths reported
14
Injuries reported
317
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4,460 adverse event reports that list this exact device (DI 00855778005135), received between Jun 14, 2017 and Sep 30, 2026 — classified as 14 deaths, 317 injuries, 4,129 malfunctions.

The most frequently reported specific device problem is “Circuit Failure” (1,846 reports).

6 other recalls cover devices cleared under the same FDA submission (P010030); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 10 reportsOct 2024: 108 reportsNov 2024: 74 reportsDec 2024: 74 reportsJan 2025: 125 reportsFeb 2025: 76 reportsMar 2025: 120 reportsApr 2025: 91 reportsMay 2025: 127 reportsJun 2025: 103 reportsJul 2025: 155 reportsAug 2025: 121 reportsSep 2025: 87 reportsOct 2025: 81 reportsNov 2025: 125 reportsDec 2025: 121 reportsJan 2026: 90 reportsFeb 2026: 115 reportsMar 2026: 113 reportsApr 2026: 125 reportsMay 2026: 109 reportsJun 2026: 98 reportsJul 2026: 118 reportsAug 2026: 125 reportsSep 2026: 180 reportsOct 2026: 0 reports202420252026180
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Circuit Failure1,846
  • Structural Problem495
  • Protective Measures Problem303
  • Device Displays Incorrect Message247
  • Break233
  • Incorrect Interpretation of Signal189
  • Display or Visual Feedback Problem171
  • Defective Component148

Patient problems

  • No Clinical Signs, Symptoms or Conditions2,610
  • No Known Impact Or Consequence To Patient878
  • Insufficient Information620
  • Arrhythmia331
  • Skin Inflammation/ Irritation10
  • Death8
  • No Information5
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available4

Recent adverse event reports

The 25 most recent of 4,460 reports, in the reporter's own words as published by the FDA.

  • InjurySep 30, 2026Over-Sensing

    A patient received 2 inappropriate shocks. Motion artifact contributed to the false detection. The response buttons were not pressed during the event. The patient was at the facility at time of treatment. the patient did continue use of lifevest. No injury was reported,

    Patient problems: Arrhythmia

    MAUDE report 50945465 ↗
  • InjurySep 30, 2026Incorrect Interpretation of Signal

    A patient received an inappropriate shock. SVT contributed to the false detection. The response buttons were not pressed during the event.The patient continues to use lifevest and was in the hospital at the time of treatment. No injury was reported

    Patient problems: Arrhythmia

    MAUDE report 50945117 ↗
  • InjurySep 30, 2026Over-Sensing

    A patient received two inappropriate shocks. Motion artifact contributed to the false detection. The response buttons were not pressed during the event.The patient was taken to the hospital for further evaluation. The patient did continue use of lifevest. No injury was reported.

    Patient problems: Arrhythmia

    MAUDE report 50945018 ↗
  • InjurySep 29, 2026Over-Sensing

    A patient received an inappropriate shock. Artifact contributed to the false detection. The response buttons were pressed after treatment.The patient continues wearing LifeVest. The patient was at the hospital at the time of treatment. No injury was reported.

    Patient problems: Arrhythmia

    MAUDE report 50935880 ↗
  • InjurySep 29, 2026Over-Sensing

    A patient received an inappropriate shock. Motion artifact contributed to the false detection. The response buttons were pressed during the event.The patient went to the hospital for further evaluation. The patient did continue use of lifevest. No injury was reported.

    Patient problems: Arrhythmia

    MAUDE report 50935702 ↗
  • InjurySep 29, 2026Incorrect Interpretation of Signal

    A patient received an inappropriate shock. SVT contributed to the false detection. The response buttons were not pressed during the event. The patient continues to wear lifevest. It is unknown if patient sought medical attention. No injury was reported.

    Patient problems: Arrhythmia

    MAUDE report 50935613 ↗
  • InjurySep 29, 2026Over-Sensing

    A patient received an inappropriate shock. Oversensing of low amplitude cardiac signal and motion artifact contributed to the false detection. The response buttons were pressed after the event.The patient continues wearing LifeVest. It is unknown if patient sought medical attention. No injury was reported.

    Patient problems: Arrhythmia

    MAUDE report 50935409 ↗
  • InjurySep 29, 2026Incorrect Interpretation of Signal

    A patient received an inappropriate shock. AF with RVR contributed to the false detection. The response buttons were not pressed during the event.The patient continues wearing LifeVest. The patient was at the hospital at the time of treatment. No injury was reported

    Patient problems: Arrhythmia

    MAUDE report 50934648 ↗
  • MalfunctionSep 29, 2026Structural Problem

    A US distributor reported that an electrode belt cannot run detect and treat testing.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50934059 ↗
  • MalfunctionSep 29, 2026Circuit Failure

    A US distributor reported that a patient was experiencing check therapy electrode messages.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50933968 ↗
  • MalfunctionSep 29, 2026Structural Problem

    A US distributor reported that an electrode belt does not communicate with a monitor.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50933884 ↗
  • InjurySep 29, 2026Over-Sensing

    A patient received an inappropriate shock. AF with RVR and motion artifact contributed to the false detection. The response buttons were not pressed during the event.The patient went to the hospital for further evaluation and continues to wear the LifeVest. No injury was reported

    Patient problems: Arrhythmia

    MAUDE report 50933828 ↗
  • MalfunctionSep 29, 2026Structural Problem

    A US distributor reported that an electrode belt connector latch does not secure with monitor.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50933807 ↗
  • InjurySep 29, 2026Incorrect Interpretation of Signal

    A patient received an inappropriate shock. AF with RVR contributed to the false detection. The response buttons were not pressed during the event. The patient did continue use of lifevest. It is unknown if patient sought medical attention. No injury was reported

    Patient problems: Arrhythmia

    MAUDE report 50932989 ↗
  • InjurySep 29, 2026Incorrect Interpretation of Signal

    A patient received an inappropriate shock. AF with RVR contributed to the false detection. The response buttons were pressed during the event.The patient continues wearing LifeVest. It is unknown if patient sought medical attention. No injury was reported.

    Patient problems: Arrhythmia

    MAUDE report 50932358 ↗
  • InjurySep 29, 2026Over-Sensing

    A patient received an inappropriate shock. Motion artifact contributed to the false detection. The response buttons were not pressed during the event.The patient continues wearing LifeVest. No injury was reported.

    Patient problems: Arrhythmia

    MAUDE report 50931943 ↗
  • InjurySep 29, 2026Incorrect Interpretation of Signal

    A patient received an inappropriate shock. Oversensing of low amplitude cardiac signal contributed to the false detection. The response buttons were pressed after the event. The patient did continue use of lifevest. It is unknown if patient sought medical attention. No injury was reported.

    Patient problems: Arrhythmia

    MAUDE report 50931575 ↗
  • MalfunctionSep 29, 2026Circuit Failure

    A US distributor reported that a patient's electrode belt could not run detect and treat testing.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50931273 ↗
  • MalfunctionSep 29, 2026Circuit Failure

    A US distributor reported that an electrode belt had non-functional ECGs.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50930993 ↗
  • MalfunctionSep 29, 2026Structural Problem

    A US distributor reported that an electrode belt had non-functional therapy electrodes.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50930915 ↗
  • MalfunctionSep 29, 2026Structural Problem

    A US distributor reported that an electrode belt had non-functional therapy electrodes.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50930904 ↗
  • InjurySep 29, 2026Over-Sensing

    A patient received an inappropriate shock. Motion artifact contributed to the false detection. The response buttons were not pressed during the event.The patient went to the hospital for further evaluation and continues to wear the LifeVest. No injury was reported

    Patient problems: Arrhythmia

    MAUDE report 50930901 ↗
  • MalfunctionSep 29, 2026Structural Problem

    A US distributor reported that an electrode belt had non-functional therapy electrodes.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50930885 ↗
  • InjurySep 29, 2026Incorrect Interpretation of Signal

    A patient received an inappropriate shock. Self-converting VT contributed to the false detection. The response buttons were not pressed during the event.The patient continues wearing the LifeVest. The patient was at the hospital at time of event. No injury was reported.

    Patient problems: Arrhythmia

    MAUDE report 50930798 ↗
  • InjurySep 29, 2026Incorrect Interpretation of Signal

    A patient received an inappropriate shock. SVT contributed to the false detection. The response buttons were pressed during the event.The patient continues to use lifevest. It is unknown if patient sought medical attention. No injury was reported.

    Patient problems: Arrhythmia

    MAUDE report 50930539 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P010030. One submission often covers many models, so check the product and lot details in each record.

  • Z-0353-2018 · Sep 12, 2017 · Terminated
    LifeVest Wearable Defibrillator Model 4000, Product Number 10A0988-A01. The LifeVest system is indicated for patients 18 years of age and older who are at risk for sudden cardiac
  • Z-0625-2011 · Sep 24, 2010 · Terminated
    Model 4000 Monitor component of the LifeVest Wearable Defibrillator
  • Z-0883-2009 · Oct 21, 2008 · Terminated
    Monitor portion of the LifeVest device - specifically the Defibrillator printed circuit assembly (PCA) within the monitor. The LifeVest is indicated for adult patients who are at
  • Z-0061-2008 · Sep 14, 2007 · Terminated
    Battery Pack portion of the LifeVest device. The LifeVest is a wearable cardioverter defibrillator worn by a patient at risk for sudden cardiac arrest (SCA).
  • Z-0936-2007 · Apr 27, 2007 · Terminated
    Electrode Belt, component of the LifeVest Wearable Defibrillator, Product number 10A0889-A01. Specifically the 3 Therapy Electrode Pads attached to the Electrode Belt.
  • Z-0578-03 · Jan 10, 2003 · Terminated
    Life Vest (WCD 3000) Electrode Belt. Wearable Cardioverter Defibrillator. Product label 20A0051-FI.