GalaFORM 3D 100043-02Recalls & reported problems
Tepha, Inc. · GalaFORM 3D scaffold is a bioresorbable surgical mesh manufactured from poly-4-hydroxybutyrate (P4HB). P4HB is produced from a naturally occurring monomer and is processed into monofilament fibers and knitted into a surgical scaffold. The scaffold has a 3D curvature supported with a P4HB rim designed to promote better conformance with a patient's anatomy in locations in which a flat design does not easily conform. P4HB bioresorbs through a process of hydrolysis and hydrolytic enzymatic digestion. It has been developed to optimize resorption rate and prolong strength retention in order to provide support throughout the expected period of healing. Although the scaffold loses strength with time, its porous construction was designed to allow native tissue ingrowth and gradual transfer of load from the scaffold to the tissue. Pre-clinical implantation studies indicate that GalaFORM 3D scaffold retains approximately 70% of its strength at 12 weeks. Bioresorption of the scaffold material will be essentially complete within 18-24 months. · Class 2 · Implant
All models of GalaFORM 3D (5)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Tepha GalaFORM 3D surgical scaffold is a resorbable mesh used to support tissue during healing. Reports mostly describe poor wound healing and other breast-surgery complications, along with reports that the scaffold did not dissolve or was not compatible with the patient. There are 41 reports, with 40 injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 41
- Deaths reported
- 0
- Injuries reported
- 40
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 41 adverse event reports that list this exact device (DI 00855920006232), received between Sep 10, 2018 and Nov 1, 2022 — classified as 40 injuries, 1 malfunction.
The most frequently reported specific device problem is “Patient-Device Incompatibility” (27 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Patient-Device Incompatibility27
- Patient Device Interaction Problem4
- Device Appears to Trigger Rejection3
- Migration3
- Contamination2
- Adverse Event Without Identified Device or Use Problem1
- Expulsion1
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- Impaired Healing10
- Unspecified Infection4
- Unspecified Reproductive System or Breast Problem4
- Breast Discomfort/Pain3
- Hematoma3
- Erythema2
- Post Operative Wound Infection2
- Seroma2
Recent adverse event reports
The 25 most recent of 41 reports, in the reporter's own words as published by the FDA.
InjuryNov 1, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL (B)(4): ON (B)(6) 2018, PATIENT UNDERWENT BILATERAL AUGMENTATION MASTOPEXY USING A VERTICAL INCISION. 2 PIECES OF GALAFORM (GALAFLEX 3DR) WERE USED. CONCOMITANT PROCEDURE WAS CAPSULORRHAPY. ON 8/31/18, LEFT NECROSIS WAS REPORTED. PATIENT TREATED WITH MEDICATION (NITROPASTE). RESOLVED WITHOUT SEQUELAE ON UNKNOWN DATE. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE NECROSIS AS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Necrosis
MAUDE report 15711865 ↗InjuryNov 1, 2022Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
PER CLINICAL TRIAL (B)(4): ON (B)(6) 2018, PATIENT UNDERWENT BILATERAL SUPERIOR PEDICLE MASTOPEXY USING A VERTICAL INCISION. 2 PIECES OF GALAFORM (GALAFLEX 3DR) WERE USED. CONCOMITANT PROCEDURES INCLUDED EARLOBE REDUCTION, FAT GRAFTING TO FACE, AUTOAUGMENTATION LIPOSUCTION TO ARMS, THIGHS, BACK. ON (B)(6) 2018, DELAYED WOUND HEALING OF LEFT BREAST WAS REPORTED. THERE WAS A SMALL OPEN WOUND AT THE MIDPOINT OF THE VERTICAL INCISION OF THE LEFT BREAST. ON (B)(6) 2018 (SAME DAY), THE AREA WAS DEBRIDED - THERE WAS EXPOSED MESH AT THE BASE OF THE OPENING. NO EVIDENCE OF ANY WOUND INFECTION. PATIENT PROBABLY HAD A SEROMA WHICH DRAINED SIMULTANEOUSLY. PATIENT TREATED WITH POLYSPORIN AND DRESSING CHANGES. RESOLUTION DATE AND OUTCOME UNKNOWN. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE DELAYED WOUND HEALING AS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Impaired Healing; Seroma
MAUDE report 15711863 ↗InjuryAug 2, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL (B)(4): ON (B)(6) 2018, PATIENT UNDERWENT BILATERAL MASTOPEXY; SUPERIOR PEDICLE, USING A VERTICAL INCISION. 2 PIECES OF GALAFORM (GALAFLEX 3DR) WERE USED. CONCOMITANT PROCEDURES INCLUDED: BREAST IMPLANT REMOVAL, AUTOAUGMENTATION, LIPOSUCTION TO ABDOMEN, AXILLA, MID/LOWER BACK, HIPS. ON (B)(6) 2018, DELAYED WOUND HEALING; LEFT BREAST, WAS REPORTED. PATIENT HAD DEBRIDEMENT ON (B)(6) 2018 AND WET TO DRY DRESSING CHANGES. NO INDICATION OF INFECTION FROM CULTURES AND NO INDICATION OF MESH REMOVAL. RESOLVED WITH UNKNOWN SEQUELAE ON 11/13/18. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE DELAYED WOUND HEALING AS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Impaired Healing
MAUDE report 15144884 ↗InjuryAug 2, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL (B)(6): ON (B)(6) 2018, PATIENT UNDERWENT BILATERAL AUGMENTATION MASTOPEXY USING A FULL VERTICAL AND HORIZONTAL T INCISION. 2 PIECES OF GALAFORM (GALAFLEX 3DR) WERE USED. NO CONCOMITANT PROCEDURES WERE DONE. ON (B)(6) 2018, DELAYED WOUND HEALING; LEFT BREAST T REGION, WAS REPORTED. PATIENT HAD DRESSING CHANGES AND HYPERBARIC CHAMBER TREATMENT. RESOLVED WITHOUT SEQUELAE ON AN UNKNOWN DATE. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE REPORTED DELAYED WOUND HEALING IS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Impaired Healing
MAUDE report 15144883 ↗InjuryAug 2, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL (B)(4): ON (B)(6) 2019 PATIENT UNDERWENT BILATERAL AUGMENTATION MASTOPEXY USING A FULL VERTICAL AND HORIZONTAL T INCISION. 2 PIECES OF GALAFORM (GALAFLEX 3DR) WERE USED. NO CONCOMITANT PROCEDURES WERE DONE. ON (B)(6) 2019, DELAYED WOUND HEALING; RIGHT T REGION WAS REPORTED. PATIENT HAD WOUND DEBRIDEMENT AND DRESSING CHANGES. RESOLVED WITHOUT SEQUELAE ON (B)(6) 2019. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE DELAYED WOUND HEALING AS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Impaired Healing
MAUDE report 15144882 ↗InjuryJul 28, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
PER CLINICAL TRIAL (B)(4): ON (B)(6) 2019, PATIENT UNDERWENT BILATERAL AUGMENTATION MASTOPEXY USING A FULL VERTICAL AND HORIZONTAL T INCISION. 2 PIECES OF GALAFORM (GALAFLEX 3DR) WERE USED. NO CONCOMITANT PROCEDURES WERE DONE. ON (B)(6) 2020, AN AE OF: GALAFORM DID NOT DISSOLVE; LEFT LATERAL BREAST, WAS REPORTED. GALAFORM WAS EXCISED ON 6/18/20. RESOLVED WITHOUT SEQUELAE ON (B)(6) 2020. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE REPORTED AE IS RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Insufficient Information
MAUDE report 15123639 ↗InjuryJul 19, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL (B)(4): ON (B)(6) 2019, PATIENT UNDERWENT BILATERAL AUGMENTATION MASTOPEXY USING A VERTICAL INCISION. 2 PIECES OF GALAFORM (GALAFLEX 3DR) WERE USED. NO CONCOMITANT PROCEDURES WERE DONE. ON (B)(6) 2019, LEFT BREAST HEMATOMA WAS REPORTED. ON (B)(6) 2019, HEMATOMA WAS EVACUATED. RESOLVED WITHOUT SEQUELAE ON (B)(6) 2019. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE LEFT BREAST HEMATOMA IS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Hematoma
MAUDE report 15057490 ↗InjuryJul 12, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL CP-1060: ON (B)(6) 18, PATIENT UNDERWENT BILATERAL AUGMENTATION MASTOPEXY USING A FULL VERTICAL AND HORIZONTAL T INCISION. 2 PIECES OF GALAFORM (GALAFLEX 3DR) WERE USED. NO CONCOMITANT PROCEDURES WERE DONE. ON (B)(6) 18, RIGHT BREAST SEROMA AND INFECTION WAS REPORTED. PATIENT HAD CULTURE TAKEN AND WAS TREATED WITH MEDICATION (BACTRIM). RESOLVED WITHOUT SEQUELAE ON (B)(6) 18. ON (B)(6) 18, DELAYED WOUND HEALING; RIGHT T REGION, WAS REPORTED. PATIENT TREATED WITH DRESSING CHANGES. RESOLVED WITHOUT SEQUELAE ON (B)(6) 18. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE SEROMA, INFECTION AND DELAYED WOUND HEALING ARE LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Impaired Healing; Seroma; Unspecified Infection
MAUDE report 14984998 ↗InjuryJul 6, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL (B)(4): ON (B)(6) 2018, PATIENT UNDERWENT BILATERAL REVISION AUGMENTATION MASTOPEXY USING A FULL VERTICAL AND HORIZONTAL T INCISION. 1 PIECE OF GALAFORM (GALAFLEX 3DR) WAS USED. NO CONCOMITANT PROCEDURES WERE DONE. ON (B)(6) 2018, DOG EAR: MEDIAL FOLD OF BREASTS WAS REPORTED. PATIENT HAD DOG EAR REMOVAL PROCEDURE ON (B)(6) 2018. A PIECE OF MESH WAS REMOVED FROM LEFT BREAST. BILATERAL BREAST SCAR REVISION ALSO DONE. RESOLVED WITHOUT SEQUELAE ON (B)(6) 2018. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE REPORTED ADVERSE EVENT AS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Physical Asymmetry
MAUDE report 14938450 ↗InjuryJul 6, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL (B)(6): ON (B)(6) 2018, PATIENT UNDERWENT BILATERAL REVISION AUGMENTATION MASTOPEXY USING A FULL VERTICAL AND HORIZONTAL T INCISION. 1 PIECE OF GALAFORM (GALAFLEX 3DR) WAS USED. CONCOMITANT PROCEDURES INCLUDED BL SUPERIOR CAPSULOTOMY AND BL INFERIOR CAPSULORRHAPHY. ON (B)(6) 2018, RIGHT BREAST INCISION SITE COMPLICATION WAS REPORTED. SMALL ESCHAR ON INCISION. PATIENT WAS TREATED WITH MEDICATION (TOPICAL BETADINE). RESOLVED WITHOUT SEQUELAE ON (B)(6) 2018. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE REPORTED INCISION SITE COMPLICATION IS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Unspecified Tissue Injury
MAUDE report 14938446 ↗InjuryJun 21, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL (B)(6): ON (B)(6) 2017, PATIENT UNDERWENT BILATERAL MASTOPEXY; INFERIOR PEDICLE, USING A FULL VERTICAL AND HORIZONTAL T INCISION. 2 PIECES OF GALAFORM (GALAFLEX 3DR) WERE USED. CONCOMITANT PROCEDURES INCLUDED FAT GRAFTING OF LEFT BREAST AND LIPOSUCTION. ON (B)(6) 2017, RIGHT DELAYED WOUND HEALING WAS REPORTED. PATIENT TREATED WITH MEDICATION (BACTROBAN). RESOLVED WITHOUT SEQUELAE ON (B)(6) 2017. REVIEW BY MEDICAL AFFAIRS IDENTIFIES AS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Impaired Healing
MAUDE report 14757067 ↗InjuryJun 21, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL (B)(4): ON (B)(6) 2017 PATIENT UNDERWENT BILATERAL MASTOPEXY; INFERIOR PEDICLE, USING A FULL VERTICAL AND HORIZONTAL T INCISION. 2 PIECES OF GALAFORM (GALAFLEX 3DR) WERE USED. CONCOMITANT PROCEDURES INCLUDED FAT GRAFTING OF LEFT BREAST AND LIPOSUCTION. ON (B)(6) 2017 BREAST HEMATOMA WAS REPORTED. HEMATOMA EVACUATED ON (B)(6) 2017. RESOLVED WITHOUT SEQUELAE ON (B)(6) 2017. REVIEW BY MEDICAL AFFAIRS IDENTIFIES AS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Hematoma
MAUDE report 14756851 ↗MalfunctionJun 21, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL (B)(4): ON (B)(6) 2017, PATIENT UNDERWENT BILATERAL REVISION AUGMENTATION MASTOPEXY USING A FULL VERTICAL AND HORIZONTAL T INCISION. 2 PIECES OF GALAFORM (GALAFLEX 3DR) WERE USED. CONCOMITANT PROCEDURE INCLUDED LIPOSUCTION. ON (B)(6) 2017, HEMATOMA; BREAST WAS REPORTED. NO TREATMENT WAS GIVEN OR ACTION TAKEN. RESOLVED WITHOUT SEQUELAE ON (B)(6) 2017. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE REPORTED HEMATOMA IS LIKELY RELATED TO THE DEVICE AND NOT SERIOUS.
Patient problems: Hematoma
MAUDE report 14756183 ↗InjuryJun 21, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL (B)(6): ON (B)(6) 2017, PATIENT UNDERWENT BILATERAL MASTOPEXY; INFERIOR PEDICLE, USING A FULL VERTICAL AND HORIZONTAL T INCISION. 2 PIECES OF GALAFORM (GALAFLEX 3DR) WERE USED. CONCOMITANT PROCEDURES INCLUDED FAT GRAFTING TO BILATERAL BREASTS, ABDOMINOPLASTY, LIPOSUCTION, FAT GRAFTING TO BUTTOCKS. ON (B)(6) 2017, RIGHT BREAST DELAYED WOUND HEALING WAS REPORTED. PATIENT HAD DEBRIDEMENT AND APPLICATION OF WET TO DRY DRESSING CHANGES. ALSO TREATED WITH MEDICATION (AQUAPHOR). RESOLVED WITHOUT SEQUELAE ON (B)(6) 2018. REVIEW BY MEDICAL AFFAIRS IDENTIFIES AS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Impaired Healing
MAUDE report 14756176 ↗InjuryJun 21, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL (B)(4): ON (B)(6) 2017, PATIENT UNDERWENT BILATERAL BREAST REDUCTION; INFERIOR PEDICLE, USING A FULL VERTICAL AND HORIZONTAL T INCISION. 2 PIECES OF GALAFORM (GALAFLEX 3DR) WERE USED. CONCOMITANT PROCEDURES INCLUDED LIPOSUCTION AND FAT GRAFTING TO BUTTOCKS. ON (B)(6) 2017, INFECTION; BREAST WAS REPORTED. PATIENT TREATED WITH MEDICATION (BACTROBAN, KEFLEX, ULTRAM, BACTRIM). RESOLVED WITHOUT SEQUELAE ON (B)(6) 2017. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE REPORTED ADVERSE EVENT AS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Unspecified Infection
MAUDE report 14756160 ↗InjuryJun 21, 2022Migration; Migration
PER CLINICAL TRIAL CP-(B)(6): ON (B)(6) 2017, PATIENT UNDERWENT BILATERAL REVISION AUGMENTATION MASTOPEXY USING A FULL VERTICAL AND HORIZONTAL T INCISION. 2 PIECES OF GALAFORM (GALAFLEX 3DR) WERE USED. CONCOMITANT PROCEDURE INCLUDED LIPOSUCTION. ON (B)(6) 2017 MIGRATION, MALPOSITION OR CONTRACTION WAS REPORTED. PATIENT HAD PROCEDURE TO LOWER BREAST POCKET ON (B)(6) 2017. RESOLVED WITHOUT SEQUELAE ON (B)(6) 2017. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE REPORTED MIGRATION IS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Unspecified Reproductive System or Breast Problem
MAUDE report 14756137 ↗InjuryMay 7, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL CP-1060: ON (B)(6) 2018, PATIENT UNDERWENT REVISION BILATERAL MASTOPEXY/REDUCTION; INFERIOR PEDICLE, USING A FULL VERTICAL AND HORIZONTAL T INCISION. 2 PIECES OF GALAFORM (GALAFLEX 3DR) WERE USED. CONCOMITANT PROCEDURES INCLUDE FAT GRAFTING TO LEFT BREAST AND LIPOSUCTION. ON (B)(6) 2018, INFECTION OF LEFT BREAST WAS REPORTED. TREATED WITH MEDICATION (BACTRIM AND BACTROBAN). RESOLVED WITHOUT SEQUELAE ON (B)(6) 2018. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE ADVERSE EVENT AS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Unspecified Infection
MAUDE report 14319361 ↗InjuryApr 22, 2022Patient Device Interaction Problem; Patient Device Interaction Problem
PER CLINICAL TRIAL CP-(B)(4): ON (B)(6) 2019, PATIENT UNDERWENT BILATERAL AUGMENTATION MASTOPEXY USING A FULL VERTICAL AND HORIZONTAL T INCISION. 2 PIECES OF GALAFORM (GALAFLEX 3DR) WERE USED. CONCOMITANT PROCEDURES INCLUDED FAT GRAFTING TO BOTH BREASTS, ABDOMINAL SCAR REVISION AND REDUCTION LABIAPLASTY. ON (B)(6) 2019, BREAST BLISTER WAS REPORTED. SAE. ON (B)(6) 2019, PATIENT HAD BILATERAL DEBRIDEMENT AND BREAST IMPLANT/MESH REMOVAL. PATIENT WAS TREATED WITH WOUND VAC AND MEDICATION (BACTROBAN OINTMENT). RESOLVED WITH SEQUELAE ON (B)(6) 2019. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE BLISTER AS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Blister
MAUDE report 14190205 ↗InjuryApr 22, 2022Migration; Migration
PER CLINICAL TRIAL (B)(4): ON (B)(6) 2018, PATIENT UNDERWENT BILATERAL BREAST AUGMENTATION MASTOPEXY USING A SHORT SCAR INCISION. 2 PIECES OF GALAFORM (GALAFLEX 3DR) WERE USED. NO CONCOMITANT PROCEDURES WERE DONE. ON (B)(6) 2018, BILATERAL MIGRATION, MALPOSITION OR CONTRACTION WAS REPORTED. ON (B)(6) 2019, BILATERAL CAPSULORRHAPHIES WERE DONE. RESOLVED WITHOUT SEQUELAE ON (B)(6) 2019. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE REPORTED MIGRATION AS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Unspecified Reproductive System or Breast Problem
MAUDE report 14190081 ↗InjuryApr 22, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL (B)(4): ON (B)(6) 2018, PATIENT UNDERWENT BILATERAL AUGMENTATION MASTOPEXY USING A FULL VERTICAL AND HORIZONTAL T INCISION. 2 PIECES OF GALAFORM (GALAFLEX 3DR) WERE USED. NO CONCOMITANT PROCEDURES WERE DONE. ON (B)(6) 2018, LEFT BREAST DELAYED WOUND HEALING WAS REPORTED. PATIENT TREATED WITH MEDICATION (AQUAPHOR). RESOLVED WITHOUT SEQUELAE ON (B)(6) 2018. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE REPORTED DELAYED WOUND HEALING AS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Impaired Healing
MAUDE report 14189955 ↗InjuryApr 22, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL (B)(4): ON (B)(6) 2019, PATIENT UNDERWENT BILATERAL AUGMENTATION MASTOPEXY USING A FULL VERTICAL AND HORIZONTAL T INCISION. 2 PIECES OF GALAFORM (GALAFLEX 3DR) WERE USED. CONCOMITANT PROCEDURES INCLUDED BILATERAL NIPPLE REDUCTION, EARLOBE REPAIR, EXCISE FOREHEAD LESION. ON 2/26/19, BILATERAL CAPSULAR CONTRACTURE WAS REPORTED. ON (B)(6) 2019, PATIENT HAD OPEN BILATERAL CAPSULECTOMIES, REMOVAL OF BREAST IMPLANTS AND FAT GRAFTING TO BREASTS. RESOLVED WITHOUT SEQUELAE ON (B)(6) 2019. REVIEW BY MEDICAL AFFAIRS IDENTIFIES CAPSULAR CONTRACTURE AS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Capsular Contracture
MAUDE report 14189951 ↗InjuryApr 22, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL CP-1060: ON (B)(6) 2019, PATIENT UNDERWENT BILATERAL AUGMENTATION MASTOPEXY USING A FULL VERTICAL AND HORIZONTAL T INCISION. 2 PIECES OF GALAFORM (GALAFLEX 3DR) WERE USED. CONCOMITANT PROCEDURES INCLUDED BILATERAL NIPPLE REDUCTION, EARLOBE REPAIR, EXCISE FOREHEAD LESION. ON (B)(6) 2019, DRAINAGE BILATERAL BREASTS WAS REPORTED. PATIENT TREATED WITH MEDICATION (BACTROBAN, BACTRIM). RESOLVED WITHOUT SEQUELAE ON (B)(6)2019. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE WOUND DRAINAGE AS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Fluid Discharge
MAUDE report 14189946 ↗InjuryApr 20, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL (B)(6): ON (B)(6) 2018, PATIENT UNDERWENT BILATERAL AUGMENTATION MASTOPEXY USING A FULL VERTICAL AND HORIZONTAL T INCISION. 2 PIECES OF GALAFORM (GALAFLEX 3DR) WERE USED. NO CONCOMITANT PROCEDURES WERE DONE. ON (B)(6) 2019, RIGHT BREAST PTOSIS WAS REPORTED. NO ACTION WAS TAKEN. PATIENT MAY NEED SKIN EXCISION ON RIGHT BREAST. REVIEW OF MEDICAL AFFAIRS IDENTIFIED THAT THE REPORTED AE IS LIKELY RELATED TO THE DEVICE AND NOT SERIOUS.
Patient problems: Ptosis
MAUDE report 14159254 ↗InjuryApr 20, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL (B)(6): ON (B)(6) 2018, PATIENT UNDERWENT BILATERAL MASTOPEXY; INFERIOR PEDICLE USING A FULL VERTICAL AND HORIZONTAL T INCISION. TWO PIECES OF GALAFORM (GALAFLEX 3DR) WERE USED. CONCOMITANT PROCEDURES INCLUDED ABDOMINOPLASTY, LIPO SUCTION, FAT GRAFTING TO BUTTOCKS, FAT GRAFTING TO RIGHT BREAST. ON (B)(6) 2018, LEFT BREAST ERYTHEMA/REDNESS WAS REPORTED. PATIENT WAS TREATED WITH MEDICATION (BACTROBAN, BACTRIM). RESOLVED WITHOUT SEQUELAE ON (B)(6) 2018. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE ERYTHEMA/REDNESS AS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Erythema
MAUDE report 14159250 ↗InjuryApr 20, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
PER CLINICAL TRIAL (B)(4): ON (B)(6) 2019, PATIENT UNDERWENT BILATERAL AUGMENTATION MASTOPEXY USING A FULL VERTICAL AND HORIZONTAL T INCISION. 2 PIECES OF GALAFORM (GALAFLEX 3DR) WERE USED. CONCOMITANT PROCEDURE WAS ABDOMINOPLASTY. ON (B)(6) 2019, SWELLING OF RIGHT BREAST WAS REPORTED. ON (B)(6) 2019, PATIENT HAD BILATERAL REVISION MASTOPEXY. NO RESOLUTION DATE WAS GIVEN. REVIEW BY MEDICAL AFFAIRS IDENTIFIES THE ADVERSE EVENT AS LIKELY RELATED TO THE DEVICE AND SERIOUS.
Patient problems: Swelling/ Edema
MAUDE report 14159249 ↗
FDA clearance and classification
Cleared through 510(k) K162922 on Feb 23, 2017 (decision: Substantially Equivalent). Submitted by Tepha, Inc. as “GalaFORM 3D”. The FDA review took 128 days. FDA record ↗
The FDA regulates product code OOD (surgical film) under 21 CFR 878.3300 as a Class II device. 21 CFR 878.3300 ↗
The FDA lists this type of device as life-sustaining or life-supporting.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
FDA definition of this device type: To be implanted to reinforce soft tissue or bone where weakness exists. Intended to be used wherever temporary wound support is required, to reinforce soft tissues where weakness exists in the urological, gynecological, or gastrointestinal anatomy, or for the repair of hernia or other fascial defects that require the addition of a reinforcing or bridging material to obtain the desired surgical result. The absorbable protective film also may help minimize the potential for tissue attachment to the device in case of direct contact with the viscera.
Common questions about the GalaFORM 3D 100043-02
Has the GalaFORM 3D 100043-02 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00855920006232), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the GalaFORM 3D 100043-02?
The FDA's MAUDE database lists 41 adverse event reports for this device, received between Sep 10, 2018 and Nov 1, 2022. In those reports, the most common specific device problems are “Patient-Device Incompatibility” (27), “Patient Device Interaction Problem” (4) and “Device Appears to Trigger Rejection” (3); the most common patient problems are “Impaired Healing” (10), “Unspecified Infection” (4) and “Unspecified Reproductive System or Breast Problem” (4). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the GalaFORM 3D 100043-02?
None of the reports are classified as deaths, 40 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the GalaFORM 3D 100043-02 FDA approved?
It was cleared through 510(k) K162922 on Feb 23, 2017. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the GalaFORM 3D 100043-02 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the GalaFORM 3D 100043-02, and is it still sold?
The FDA's device database lists Tepha, Inc. as the company responsible for it (brand GalaFORM 3D, model 100043-02). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Surgical Film devices
- GalaFLEX Scaffold 100035-09Tepha, Inc. · Surgical Film2 deaths90 reports
- GalaFLEX Scaffold 100035-15Tepha, Inc. · Surgical Film65 reports
- Phasix 1190500Davol Inc. · surgical film39 reports
- GalaSHAPE 3D 100042-04Tepha, Inc. · Surgical Film37 reports
- GalaFORM 3D 100043-04Tepha, Inc. · Surgical Film29 reports
- Phasix 1190400Davol Inc. · surgical film26 reports
- Phasix 1190200Davol Inc. · surgical film22 reports
- Phasix 1190300Davol Inc. · surgical film2 deaths20 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page