Aggreguide A-100Recalls & reported problems
Aggredyne, Inc. · Whole Blood Aggregometer · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Jul 23, 2019Completed
The firm identified a software design issue where the improper shut down of the instrument resulted in speeds.csv file contents of the software version 5.1 being erased. Due to this issue, the target RPM for the assay is 0 and the motor output required to achieve this RPM is also 0. Therefore, when the assay is run, the software applies 0 Volts to the motor and performs a 0 RPM assay. Because of this, there is no mixing of the blood with agonist and no moving of aggregates happens in front of the optical detectors. The resulting PAI is 0 and the result is displayed as Low.
Z-1639-2022 · Root cause: Software design · FDA record ↗
Other System, Automated Platelet Aggregation devices
- VerifyNow Platelet Reactivity Test 85225Accriva Diagnostics Holdings, Inc. · System, Automated Platelet Aggregation1 recall0 reports
- PFA-100® System 10444868Siemens Healthcare Diagnostics Products GmbH · SYSTEM, AUTOMATED PLATELET AGGREGATION29 reports
- TEG6s HEMOSTASIS SYSTEM 07-690-USHaemonetics Corporation · SYSTEM, AUTOMATED PLATELET AGGREGATION1 death1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page